Development and optimisation of a near-infrared spectroscopic system for glucose quantification in aqueous and intralipid-based samples

被引:0
|
作者
Davison, Nicholas B. [1 ]
Gaffney, Christopher J. [1 ]
Kerns, Jemma G. [1 ]
Zhuang, Qiandong [2 ]
机构
[1] Univ Lancaster, Lancaster Med Sch, Lancaster LA1 4AT, England
[2] Univ Lancaster, Phys Dept, Lancaster LA1 4YW, England
来源
ENGINEERING RESEARCH EXPRESS | 2024年 / 6卷 / 02期
关键词
near infrared spectroscopy; glucose monitoring; non-invasive sensing; optical system design; OPTICAL COHERENCE TOMOGRAPHY; BLOOD-GLUCOSE; SCATTERING PROPERTIES; 1ST OVERTONE; HUMAN SKIN; TISSUE; LIGHT; COMBINATION; LIMITATIONS; REGRESSION;
D O I
10.1088/2631-8695/ad4aed
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A non-invasive glucose sensing device could revolutionise diabetes treatment. Near Infrared (NIR) spectroscopy is a promising technology for glucose sensing; however, the design and choice of components for NIR spectroscopy can greatly affect the sensing accuracy. We aimed to develop a NIR absorption spectroscopy system to determine liquid glucose concentrations in the physiological range, by evaluating a range of NIR photodetector components and light sources. Three detection assemblies were tested: (i) a dispersive spectrometer with photodiode array, (ii) a Czerny-Turner monochromator with InGaAs photodiode and (iii) a miniature Fourier Transform Infrared (FTIR) spectrometer. A halogen lamp and NIR globar were trialled as potential light sources. The components were systematically tested by comparing the coefficient of determination and standard error of prediction (SEP) for the same set of aqueous glucose samples through 10 mmol l-1 concentration steps. The Czerny-Turner monochromator with InGaAs photodiode, along with the globar, were identified as the optimal components for the system. A range of concentration steps (1-10 mmol l-1) were scanned to identify the physiologically relevant limit of detection, which was identified as 5 mmol/l for glucose in solution. Spectra were then collected from glucose samples in 10% intralipid suspension in the 10-20 mmol l-1 range and the equivalent concentrations in solution. The SEP was greater for the intralipid samples due to strong scattering. Scattering was dominant above 1300 nm, whilst absorption was dominant below 1300 nm. Although alternative approaches achieve better resolution, our system uses simple and readily-available components and presents a platform for a non-invasive NIR glucose sensing device.
引用
收藏
页数:16
相关论文
共 50 条
  • [11] Noninvasive blood glucose measurement system based on three wavelengths in near-infrared region
    Chen, Yaqin
    Bai, Ge
    Xiao, Jun
    Wang, Long
    Luo, Qingming
    FIFTH INTERNATIONAL CONFERENCE ON PHOTONICS AND IMAGING IN BIOLOGY AND MEDICINE, PTS 1 AND 2, 2007, 6534
  • [12] Near-infrared spectroscopic measurement of urea in dialysate samples collected during hemodialysis treatments
    Eddy, CV
    Flanigan, M
    Arnold, MA
    APPLIED SPECTROSCOPY, 2003, 57 (10) : 1230 - 1235
  • [13] Research on nonlinear quantification of Rebaudioside A crystallization process based on near-infrared sensor fusion
    Dong, Hailing
    Huang, Ruiqi
    Gao, Lele
    Yang, Yan
    Xu, Xiuhua
    Nie, Lei
    Li, Lian
    Dong, Qin
    Zhang, Hui
    Xu, Jinke
    Sun, Jing
    Zang, Hengchang
    JOURNAL OF PHARMACEUTICAL INNOVATION, 2023, 18 (02) : 735 - 746
  • [14] Development of Near-infrared Spectroscopy Signal Collecting System Based on Digital Demodulation
    Liu Zhi-wei
    Chen Man
    Yan Rui
    Wang Jing-na
    Dong Xiao-hu
    Du Jiao-jiao
    2017 INTERNATIONAL CONFERENCE ON COMPUTER SYSTEMS, ELECTRONICS AND CONTROL (ICCSEC), 2017, : 1026 - 1030
  • [15] Non-invasive Blood Glucose Measurement Scheme Based on Near-infrared Spectroscopy
    Wang Shulei
    Yuan Xueguang
    Zhang Yangan
    2017 CONFERENCE ON LASERS AND ELECTRO-OPTICS PACIFIC RIM (CLEO-PR), 2017,
  • [16] Study on the Effect of Sodium Chloride Salt on Near-Infrared Spectroscopy of Glucose Aqueous Solution
    Yu Xu-yao
    Bai Zhi-liang
    Liu Rong
    Yuan Jing
    Yu Hui
    Wang Hai-jun
    Xu Ke-xin
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36 (06) : 1706 - 1711
  • [17] Near-infrared spectroscopic measurement system of the constituent concentration based on acousto-optic tunable filter
    Li, QB
    Jiang, CZ
    Xu, KX
    OPTICAL DESIGN AND TESTING, 2002, 4927 : 714 - 721
  • [18] Noninvasive Blood Glucose Analysis Based on Near-Infrared Reflectance Spectroscopy
    Lu Xiao-feng
    Zhang Ting-lin
    Xiao Feng
    Li Guang
    Wang You
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36 (07) : 2312 - 2317
  • [19] NONINVASIVE BLOOD-GLUCOSE SENSORS BASED ON NEAR-INFRARED SPECTROSCOPY
    HEISE, HM
    MARBACH, R
    KOSCHINSKY, T
    GRIES, FA
    ARTIFICIAL ORGANS, 1994, 18 (06) : 439 - 447
  • [20] Non-invasive Glucose Monitoring System Utilizing Near-Infrared Technology
    Duc Trinh-Minh Dinh
    Viet Anh Truong
    An Nhu-Phu Tran
    Hieu Xuan Le
    Hien Thi-Thu Pham
    7TH INTERNATIONAL CONFERENCE ON THE DEVELOPMENT OF BIOMEDICAL ENGINEERING IN VIETNAM (BME7): TRANSLATIONAL HEALTH SCIENCE AND TECHNOLOGY FOR DEVELOPING COUNTRIES, 2020, 69 : 401 - 405