Reasons for Removal of Rigid Internal Fixation Devices in Craniofacial Surgery: A 20-year Update

被引:0
作者
Lynn, Jeremy V. [1 ]
Daniel, Melissa C. [1 ]
Jackson, Charlotte M. K. [1 ]
Ulma, Raquel M. [1 ]
Vercler, Christian J. [1 ]
Kasten, Steven J. [1 ]
Buchman, Steven R. [1 ]
机构
[1] Univ Michigan, Sect Plast Surg, 1500 E Med Ctr Dr, Ann Arbor, MI 48109 USA
关键词
hardware; open reduction and internal fixation; rigid internal fixation; MOTOR-VEHICLE COLLISIONS; MAXILLOFACIAL TRAUMA; ORTHOGNATHIC SURGERY; FACIAL FRACTURES; BONE PLATES; PATTERNS; COMPLICATIONS; MANAGEMENT; SYSTEMS; SAFETY;
D O I
10.1097/SCS.0000000000010032
中图分类号
R61 [外科手术学];
学科分类号
摘要
The hardware utilized for rigid internal fixation of the craniofacial skeleton has evolved over time. Thus, the reasons for the unplanned removal of hardware continue to change. The purpose of this study is to compare past (1989-1995) and present (2000-2020) patient cohorts to establish trends related to unplanned removal of craniofacial hardware. A retrospective review study was designed. Data from our institution's original publication describing the unplanned removal of craniofacial hardware (1989-1995) was obtained. Data related to patients who underwent unplanned removal of hardware from 2000 to 2020 was collected from the electronic medical record. A descriptive statistical analysis was performed to compare demographics, reasons for hardware placement, and reasons for unplanned hardware removal between cohorts. This study includes 55 patients treated from 1989 to 1995 and 184 patients treated from 2000 to 2020. The average age at hardware placement decreased from 32 years (1989-1995) to 28 years (2000-2020). The most common reason for hardware placement changed from motor vehicle accident (1989-1995) to congenital deformity (2000-2020). The length of time with hardware in situ increased from 13 months (1989-1995) to 25 months (2000-2020). The most common reason for hardware removal changed from prominent hardware (1989-1995) to hardware exposure (2000-2020). In summary, patients who underwent rigid internal fixation of the craniofacial skeleton from 2000 to 2020 retained their hardware 2 times longer than patients treated from 1989 to 1995. Factors potentially contributing to increased retention include improved surgical technique, decreased profile of hardware, and increased surgeon experience. Further studies are warranted to define preoperative risk factors for unplanned hardware removal.
引用
收藏
页码:1052 / 1056
页数:5
相关论文
共 48 条
  • [41] Simoni Payman, 2003, Arch Facial Plast Surg, V5, P113
  • [42] SMITH A E, 1954, J Oral Surg (Chic), V12, P170
  • [43] Safety device use affects the incidence patterns of facial trauma in motor vehicle collisions: An analysis of the national trauma database from 2000 to 2004
    Stacey, D. Heath
    Doyle, John F.
    Gutowski, Karol A.
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 2008, 121 (06) : 2057 - 2064
  • [44] Inpatients With Gunshot Wounds to the Face
    Sun, Gordon H.
    Patil, Yash J.
    Harmych, Brian M.
    Hom, David B.
    [J]. JOURNAL OF CRANIOFACIAL SURGERY, 2012, 23 (01) : E62 - E65
  • [45] THOMA K H, 1948, J Oral Surg (Chic), V6, P125
  • [46] The Management of Facial Fractures: The Legacy of William Milton Adams
    Wallace, Robert Doyle
    Puente-Espel, Jordi
    Konofaos, Petros
    [J]. JOURNAL OF CRANIOFACIAL SURGERY, 2020, 31 (03) : 630 - 631
  • [47] COMBINED REPORT OF PROBLEMS AND COMPLICATIONS IN 793 CRANIOFACIAL OPERATIONS
    WHITAKER, LA
    MUNRO, IR
    SALYER, KE
    JACKSON, IT
    ORTIZMONASTERIO, F
    MARCHAC, D
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 1979, 64 (02) : 198 - 203
  • [48] COMPLICATIONS ASSOCIATED WITH RIGID INTERNAL-FIXATION OF FACIAL BONE-FRACTURES
    ZACHARIADES, N
    PAPADEMETRIOU, I
    RALLIS, G