Simulation Dosimetry Studies for FLASH Radiation Therapy (RT) with Ultra-High Dose Rate (UHDR) Electron Beam

被引:1
|
作者
Gazis, Nick [1 ,2 ]
Bignami, Andrea [1 ,2 ]
Trachanas, Emmanouil [1 ,2 ]
Moniaki, Melina [3 ]
Gazis, Evangelos [1 ,3 ]
Bandekas, Dimitrios [4 ]
Vordos, Nikolaos [4 ]
机构
[1] Inst Accelerating Syst & Applicat IASA, Athens 10024, Greece
[2] European Spallat Source ESS ERIC, POB 176, SE-22100 Lund, Sweden
[3] Natl Tech Univ Athens NTUA, Phys Dept, Zografou Campus, Athens 15780, Greece
[4] Democritus Univ Thrace DUTH, Phys Dept, St Lucas 65404, Greece
关键词
FLASH-RT; dosimetry; UHDR; PDD; ASTRA/FLUKA simulations;
D O I
10.3390/qubs8020013
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
FLASH-radiotherapy (RT) presents great potential as an alternative to conventional radiotherapy methods in cancer treatment. In this paper, we focus on simulation studies for a linear particle accelerator injector design using the ASTRA code, which permits beam generation and particle tracking through electromagnetic fields. Space charge-dominated beams were selected with the aim of providing an optimized generated beam profile and accelerator lattice with minimized emittance. The main results of the electron beam and ultra-high dose rate (UHDR) simulation dosimetry studies are reported for the FLASH mode radiobiological treatment. Results for the percentage depth dose (PDD) at electron beam energies of 5, 7, 15, 25, 50, 100 MeV and 1.2 GeV for Poly-methyl-methacrylate (PMMA) and water phantom vs. the penetration depth are presented. Additionally, the PDD transverse profile was simulated for the above energies, delivering the beam to the phantom. The simulation dosimetry results provide an UHDR electron beam under the conditions of the FLASH-RT. The performance of the beam inside the phantom and the dose depth depends on the linear accelerator beam's energy and stability.
引用
收藏
页数:15
相关论文
共 43 条
  • [31] The PTB water calorimeter for determining the absolute absorbed dose to water in ultra-high pulse dose rate electron beams
    Bourgouin, Alexandra
    Hackel, Thomas
    Kapsch, Ralf-Peter
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (11)
  • [32] Real-time monitoring of ultra-high dose rate electron beams using bremsstrahlung photons
    Kim, Hyun
    Jeong, Dong Hyeok
    Kang, Sang Koo
    Lee, Manwoo
    Lim, Heuijin
    Lee, Sang Jin
    Jang, Kyoung Won
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2023, 55 (09) : 3417 - 3422
  • [33] Implementation of ultra-high dose-rate electron beam from 6-MeV C-band linear accelerator for preclinical study
    Lim, H.
    Jeong, D. H.
    Jang, K. W.
    Kang, S. K.
    Kim, H. C.
    Kim, S. H.
    Lee, D. E.
    Lee, K. H.
    Lee, S. J.
    Lee, M.
    JOURNAL OF INSTRUMENTATION, 2020, 15 (09):
  • [34] State-of-the-art silicon carbide diode dosimeters for ultra-high dose-per-pulse radiation at FLASH radiotherapy
    Fleta, Celeste
    Pellegrini, Giulio
    Godignon, Philippe
    Gomez Rodriguez, Faustino
    Paz-Martin, Jose
    Kranzer, Rafael
    Schueller, Andreas
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (09)
  • [35] Determination of the ion collection efficiency of the Razor Nano Chamber for ultra-high dose-rate electron beams
    Cavallone, Marco
    Jorge, Patrik Goncalves
    Moeckli, Raphael
    Bailat, Claude
    Flacco, Alessandro
    Prezado, Yolanda
    Delorme, Rachel
    MEDICAL PHYSICS, 2022, 49 (07) : 4731 - 4742
  • [36] Preliminary study of luminescence phenomena from various materials under ultra-high dose rate proton beam irradiation for dose management
    Yamada, Ryosaku
    Nishio, Teiji
    Kinkawa, Daiki
    Tanaka, Taketo
    Omura, Mizuki
    Tabata, Yoji
    Yoshimura, Hitoshi
    Kataoka, Jun
    SCIENTIFIC REPORTS, 2024, 14 (01):
  • [37] Validation of an ultrahigh dose rate pulsed electron beam monitoring system using a current transformer for FLASH preclinical studies
    Goncalves Jorge, Patrik
    Grilj, Veljko
    Bourhis, Jean
    Vozenin, Marie-Catherine
    Germond, Jean-Francois
    Bochud, Francois
    Bailat, Claude
    Moeckli, Raphael
    MEDICAL PHYSICS, 2022, 49 (03) : 1831 - 1838
  • [38] Selection criteria for high-dose-rate surface brachytherapy and electron beam therapy in cutaneous oncology
    Buzurovic, Ivan M.
    O'Farrell, Desmond A.
    Harris, Thomas C.
    Friesen, Scott
    King, Martin T.
    Cormack, Robert A.
    Devlin, Phillip M.
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2021, 13 (02) : 195 - 204
  • [39] Ultra high dose rate Synchrotron Microbeam Radiation Therapy. Preclinical evidence in view of a clinical transfer
    Eling, Laura
    Bouchet, Audrey
    Nemoz, Christian
    Djonov, Valentin
    Balosso, Jacques
    Laissue, Jean
    Brauer-Krisch, Elke
    Adam, Jean Francois
    Serduc, Raphael
    RADIOTHERAPY AND ONCOLOGY, 2019, 139 : 56 - 61
  • [40] Secondary radiation dose modeling in passive scattering and pencil beam scanning very high energy electron (VHEE) radiation therapy
    Deut, Umberto
    Ronga, Maria Grazia
    Bonfrate, Anthony
    De Marzi, Ludovic
    MEDICAL PHYSICS, 2023, 50 (07) : 4491 - 4504