UTILIZING STENT WALL SHEAR STRESS TO ASSESS STENT AND FLOW DIVERTER PERFORMANCE FOR TREATING INTRACRANIAL ANEURYSMS

被引:0
|
作者
Suess, Taylor N. [1 ]
Gent, Stephen P. [1 ]
机构
[1] South Dakota State Univ, Dept Mech Engn, Brookings, SD 57007 USA
来源
PROCEEDINGS OF ASME 2022 INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION, IMECE2022, VOL 4 | 2022年
关键词
flow diverter; stent; computational fluid dynamics; CFD; wall shear stress; HEMODYNAMICS; RUPTURE; OCCLUSION; MODEL;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Analysis of flow diverting intracranial aneurysm repair devices has traditionally focused on reducing intrasaccular blood flow velocity and aneurysm wall shear stress (WSS) as the primary metrics for improved perceived device performance. However, the interpretation of this data has been debated, particularly with regards to the specific biological benefits of high or low aneurysm WSS. Therefore, this research proposes an additional parameter of WSS at the stent struts that could provide valuable insight regarding the device's potential to promote occlusion at the aneurysm neck by indicating potential locations of increased platelet activation and microparticle shedding. Fluid flow effects were evaluated for two flow diverters (Pipeline (TM) and FRED (TM)) and three stents (Enterprise, Atlas (TM), and LVIS (TM)) using computational fluid dynamics (CFD) models developed from two patient-derived CTA datasets with aneurysms in the middle cerebral artery (MCA) and basilar artery (BA), respectively. The device WSS data provides an additional metric for evaluating the ability of the device to constrain the blood flow within the main vessel, as well as determining the point at which the blood flow is released into the aneurysm, indicating the potential locations of the initiation of aneurysm occlusion. It is hypothesized that high device WSS close to the aneurysm neck creates a higher likelihood of aneurysm occlusion due to platelet activation and microparticle shedding, while high device WSS proximal on the device would indicate higher likelihood of undesirable daughter vessel occlusion. Conversely, low-to-moderate device WSS throughout the device length could be interpreted as less flow diversion and more consistent blood flow allowed into the aneurysm, resulting in lesser device performance.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Comparative evaluation of brain perfusion with ASL and DSC MRI following intracranial vascular flow diverter stent replacement
    Kaydu, Nesrin Erdogan
    Cindil, Emetullah
    Asfuroglu, Berrak Barutcu
    Onal, Baran
    Oner, Ali Yusuf
    JOURNAL OF CLINICAL ULTRASOUND, 2023, 51 (08) : 1390 - 1396
  • [42] Visualization of flow diverter stent wall apposition during intracranial aneurysm treatment using a virtually diluted cone beam CT technique (Vessel ASSIST)
    Ameer E. Hassan
    Jim Wise
    Elizabeth M. Burke
    Wondwossen G. Tekle
    Neuroradiology, 2021, 63 : 125 - 131
  • [43] Visualization of flow diverter stent wall apposition during intracranial aneurysm treatment using a virtually diluted cone beam CT technique (Vessel ASSIST)
    Hassan, Ameer E.
    Wise, Jim
    Burke, Elizabeth M.
    Tekle, Wondwossen G.
    NEURORADIOLOGY, 2021, 63 (01) : 125 - 131
  • [44] Healing of Aneurysm after Treatment Using Flow Diverter Stent : Histopathological Study in Experimental Canine Carotid Side Wall Aneurysm
    Lee, Jong Young
    Cho, Young Dae
    Kang, Hyun-Seung
    Han, Moon Hee
    JOURNAL OF KOREAN NEUROSURGICAL SOCIETY, 2020, 63 (01) : 34 - 44
  • [45] A comparison of 4D flow MRI-derived wall shear stress with computational fluid dynamics methods for intracranial aneurysms and carotid bifurcations - A review
    Szajer, Jeremy
    Ho-Shon, Kevin
    MAGNETIC RESONANCE IMAGING, 2018, 48 : 62 - 69
  • [46] The adverse effects of flow-diverter stent-like devices on the flow pattern of saccular intracranial aneurysm models: computational fluid dynamics study
    Tamer Hassan
    Yasser Mohamed Ahmed
    Amr Ali Hassan
    Acta Neurochirurgica, 2011, 153 : 1633 - 1640
  • [47] Treatment of intracranial aneurysms. Reconstruction of the parent artery with flow-diverting (Silk) stent
    Wagner, Aase
    Cortsen, Marie
    Hauerberg, John
    Romner, Bertil
    Wagner, Mathias Pedersen
    NEURORADIOLOGY, 2012, 54 (07) : 709 - 718
  • [48] Treatment of intracranial aneurysms. Reconstruction of the parent artery with flow-diverting (Silk) stent
    Aase Wagner
    Marie Cortsen
    John Hauerberg
    Bertil Romner
    Mathias Pedersen Wagner
    Neuroradiology, 2012, 54 : 709 - 718
  • [49] Comparison of the flow diverter and stent-assisted coiling in large and giant aneurysms: safety and efficacy based on a propensity score-matched analysis
    Yongxin Zhang
    Yu Zhou
    Pengfei Yang
    Jianmin Liu
    Yi Xu
    Bo Hong
    Wenyuan Zhao
    Qi Chen
    Qing-Hai Huang
    European Radiology, 2016, 26 : 2369 - 2377
  • [50] Phase-Contrast Magnetic Resonance Imaging Measurements in Intracranial Aneurysms In Vivo of Flow Patterns, Velocity Fields, and Wall Shear Stress: Comparison with Computational Fluid Dynamics
    Boussel, Loic
    Rayz, Vitaliy
    Martin, Alastair
    Acevedo-Bolton, Gabriel
    Lawton, Michael T.
    Higashida, Randall
    Smith, Wade S.
    Young, William L.
    Saloner, David