Intraocular pressure measurement using ICare rebound tonometer in different positions of eye and different locations on cornea

被引:0
|
作者
Wongwanwatana, Sirada [1 ]
Treesit, Isaraporn [1 ]
Funarunart, Panrapee [1 ]
Iemsomboon, Wallop [1 ]
Choontanom, Raveewan [1 ]
机构
[1] Phramongkutklao Hosp, Dept Ophthalmol, Phramongkutklao Coll Med, 315 Ratchawithi Rd, Bangkok 10400, Thailand
关键词
IC200; iCare rebound tonometer; intraocular pressure; GOLDMANN APPLANATION TONOMETER; AGREEMENT;
D O I
10.1097/MD.0000000000034874
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Intraocular pressure (IOP) is one of the most crucial aspects for diagnosis and treatment plan among patients with glaucoma. Although the gold standard for IOP measurement is Goldmann applanation tonometer (GAT)([1]), it must be mounted to a slit lamp biomicroscope. However, rebound tonometer has become popular due to its ease of operation and portable design, does not require topical anesthesia, and results do not differ significantly from those of GAT([2]). The purpose of this cross-sectional study is to investigate the difference in IOP measurement with iCare IC200 in different angles of the eye and different corneal locations. All participants underwent IOP measurement by GAT twice. Then, IOP was measured with iCare by a single physician. IOP was measured in a straight manner in the upright patient position; then participants were asked to look at fixation targets, which located in four different points. IOP was measured in upgaze, downgaze, medial gaze, and lateral gaze. Then, IOP was measured at 2mm from limbus in superior, inferior, nasal, and temporal cornea. All methods were measured twice, and the mean was used for calculation. The physician who measured IOP by iCare was masked from GAT results. A total of 168 eyes were tested with a mean age of 62.15 +/- 12.34 years. Mean IOP measured by GAT and iCare at the central cornea was 15.53 +/- 5.57 and 14.78 +/- 6.14 mmHg, respectively. The standardized mean difference (SMD) between iCare and GAT was 0.13 (-0.09-0.34), which is insignificant. The average IOP was 0.6, 0.47, 0.91, and 0.44 mmHg lower than the primary position in upgaze, downgaze, medial gaze, and lateral gaze 15degrees angulated positions respectively (p<.01). IOPs at 2mm from limbus in the inferior, nasal, and temporal cornea were 0.5, 0.69, and 0.57 mmHg lower than IOP measured at the central cornea (p=<.01). IOP measurements with iCare in different angles of eye were statistically significantly lower than in the primary position. Similarly, IOPs at different locations on cornea were lower than at the central cornea. However, the difference in IOP measurements with iCare in different angles of the eye and different corneal locations was in the trivial range and might be clinically insignificant.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Relationship of Central Corneal Thickness and Intraocular Pressure by iCare Rebound Tonometer
    Rao, Aparna
    Kumar, Mukesh
    Prakash, B.
    Varshney, Gopal
    JOURNAL OF GLAUCOMA, 2014, 23 (06) : 380 - 384
  • [2] Evaluation of intraocular pressure estimates obtained using an iCare rebound tonometer
    Khanal, Safal
    Walton, Meredith
    Davey, Pinakin Gunvant
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2017, 100 (02) : 179 - 183
  • [3] Evaluation of the ICare Rebound Tonometer as a Home Intraocular Pressure Monitoring Device
    Asrani, Sanjay
    Chatterjee, Ashmita
    Wallace, David K.
    Santiago-Turla, Cecilia
    Stinnett, Sandra
    JOURNAL OF GLAUCOMA, 2011, 20 (02) : 74 - 79
  • [4] Intraocular Pressure Measurement Using Rebound Tonometer for Deviated Angles and Positions in Human Eyes
    Takenaka, Joji
    Mochizuki, Hideki
    Kunihara, Eriko
    Tanaka, Junko
    Kiuchi, Yoshiaki
    CURRENT EYE RESEARCH, 2012, 37 (02) : 109 - 114
  • [5] Accuracy of Intraocular Pressure Measurement With the Icare Tonometer in Children
    Chan, Wai Hong
    Lloyd, I. Chris
    Symes, Richard J.
    Ashworth, Jane L.
    Cosgrove, Edel
    Pilling, Rachel
    Biswas, Susmito
    ASIA-PACIFIC JOURNAL OF OPHTHALMOLOGY, 2015, 4 (06): : 357 - 359
  • [6] Comparability of Icare Pro Rebound Tonometer with Goldmann Applanation and Noncontact Tonometer in a Wide Range of Intraocular Pressure and Central Corneal Thickness
    Tamcelik, Nevbahar
    Atalay, Eray
    Cicik, Erdogan
    Ozkok, Ahmet
    OPHTHALMIC RESEARCH, 2015, 54 (01) : 18 - 25
  • [7] Rebound Tonometer: Ideal Tonometer for Measurement of Accurate Intraocular Pressure
    Suman, Suwarna
    Agrawal, Ajai
    Pal, Virendra K.
    Pratap, Vir B.
    JOURNAL OF GLAUCOMA, 2014, 23 (09) : 633 - 637
  • [8] Comparison of the Intraocular Pressure Measured Using the New Rebound Tonometer Icare ic100 and Icare TA01i or Goldmann Applanation Tonometer
    Nakakura, Shunsuke
    Mori, Etsuko
    Fujio, Yuki
    Fujisawa, Yasuko
    Matsuya, Kanae
    Kobayashi, Yui
    Tabuchi, Hitoshi
    Asaoka, Ryo
    Kiuchi, Yoshiaki
    JOURNAL OF GLAUCOMA, 2019, 28 (02) : 172 - 177
  • [9] Intraocular Pressure Values using Icare® Rebound Tonometer and Correlation with Postconceptional Age in Premature Infants
    Monteiro, Catarina
    Vivas, Maria
    Almeida, Julio
    Ramalho, Mario
    Mota, Mafalda
    Teixeira, Susana
    JOURNAL OF OPHTHALMIC & VISION RESEARCH, 2023, 18 (03) : 267 - 271
  • [10] Comparability of three intraocular pressure measurement: iCare pro rebound, non-contact and Goldmann applanation tonometry in different IOP group
    Min Chen
    Lina Zhang
    Jia Xu
    Xinyi Chen
    Yuxiang Gu
    Yuping Ren
    Kaijun Wang
    BMC Ophthalmology, 19