A sensitive bioanalytical ultra-high-performance liquid chromatography-tandem mass spectrometry method for the simultaneous quantitation of lactone and carboxylate forms of topotecan in plasma and vitreous

被引:5
作者
Mudrova, Barbora [1 ]
Hrabakova, Katerina [1 ]
Kozlik, Petr [1 ]
Hobzova, Radka [2 ]
Sirc, Jakub [2 ]
Bosakova, Zuzana [1 ]
机构
[1] Charles Univ Prague, Fac Sci, Dept Analyt Chem, Albertov 2030, Prague 2, Czech Republic
[2] Inst Macromol Chem, Prague, Czech Republic
关键词
anti-cancer drug topotecan; carboxylate form; lactone form; topotecan stability; UHPLC-MS/MS; HPLC METHOD; I INHIBITOR; PHARMACOKINETICS; CAMPTOTHECIN; PHARMACOLOGY; STABILITY; INJECTION; 104864-A; ASSAY;
D O I
10.1002/jssc.202400181
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Topotecan (TPT) is used in the treatment of retinoblastoma, the most common malignant intraocular tumor in children. TPT undergoes pH-dependent hydrolysis of the lactone ring to the ring-opened carboxylate form, with the lactone form showing antitumor activity. A selective, and highly sensitive ultra-high-performance liquid chromatography-tandem mass spectrometry method was developed for the determination of both forms of TPT in one mobile phase composition in plasma and vitreous humor matrices. The method showed an excellent linear range of 0.375-120 ng/mL for the lactone. For the carboxylate, the linear range was from 0.75 to 120 ng/mL. The matrix effect and the recovery for the lactone ranged from 98.5% to 106.0% in both matrices, for the carboxylate form, it ranged from 94.9% to 101.2%. The dynamics of the transition between TPT lactone and TPT carboxylate were evaluated at different pH environments. The stability of TPT forms was assessed in plasma and vitreous humor at 8 and 37 degrees C and a very fast conversion of lactone to carboxylate form occurred at 37 degrees C in both matrices. The method developed facilitates the investigation of TPT pharmacodynamics and the release kinetics in the development of the innovative local drug delivery systems.
引用
收藏
页数:13
相关论文
共 31 条
[1]   Quantification of topotecan by liquid chromatography-mass spectrometry (LC-MS). Application to intestinal transport using rat everted gut sacs [J].
Arellano, Cecile ;
Gandia, Peggy ;
Bettuing, Laure ;
Woodley, John ;
Chatelut, Etienne .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2010, 878 (7-8) :645-652
[2]  
Arun B, 2001, Expert Opin Pharmacother, V2, P491, DOI 10.1517/14656566.2.3.491
[3]   Topotecan exposure estimation in pediatric acute myeloid leukemia supported by LC-MS-based drug monitoring and pharmacokinetic analysis [J].
Baczek, Tomasz ;
Konieczna, Lucyna ;
Belka, Mariusz ;
Maciejka-Kapuscinska, Lucyna ;
Wisniewski, Jakub ;
Niedzwiecki, Maciej ;
Balcerska, Anna ;
Adamkiewicz-Drozynska, Elzbieta ;
Wachowiak, Jacek .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2012, 70 :330-336
[4]   Concurrent determination of topotecan and model permeability markers (atenolol, antipyrine, propranolol and furosemide) by reversed phase liquid chromatography: Utility in Caco-2 intestinal absorption studies [J].
Bansal, Tripta ;
Singh, Manoj ;
Mishra, Gautam ;
Talegaonkar, Sushama ;
Khar, Roop K. ;
Jaggi, Manu ;
Mukherjee, Rama .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 859 (02) :261-266
[5]   Pharmacokinetic analysis of topotecan after intra-vitreal injection. Implications for retinoblastoma treatment [J].
Buitrago, Emiliano ;
Hoecht, Christian ;
Chantada, Guillermo ;
Fandino, Adriana ;
Navo, Elliot ;
Abramson, David H. ;
Schaiquevich, Paula ;
Bramuglia, Guillermo F. .
EXPERIMENTAL EYE RESEARCH, 2010, 91 (01) :9-14
[6]  
Burke TG., 2006, Ann NY Acad Sci, V922, P36, DOI [10.1111/j.1749-6632.2000.tb07023.x, DOI 10.1111/J.1749-6632.2000.TB07023.X]
[7]   Hydrogel implants for transscleral drug delivery for retinoblastoma treatment [J].
Cocarta, Ana-Irina ;
Hobzova, Radka ;
Sirc, Jakub ;
Cerna, Tereza ;
Hrabeta, Jan ;
Svojgr, Karel ;
Pochop, Pavel ;
Kodetova, Martina ;
Jedelska, Jarmila ;
Bakowsky, Udo ;
Uhlik, Jiri .
MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2019, 103
[8]   Stability and compatibility of topotecan hydrochloride for injection with common infusion solutions and containers [J].
Craig, SB ;
Bhatt, UH ;
Patel, K .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1997, 16 (02) :199-205
[9]   A KINETIC AND MECHANISTIC STUDY OF THE HYDROLYSIS OF CAMPTOTHECIN AND SOME ANALOGS [J].
FASSBERG, J ;
STELLA, VJ .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1992, 81 (07) :676-684
[10]   Classification of Vitreous Seeds in Retinoblastoma Correlations with Patient, Tumor, and Treatment Characteristics [J].
Francis, Jasmine H. ;
Marr, Brian P. ;
Abramson, David H. .
OPHTHALMOLOGY, 2016, 123 (07) :1601-1605