Technical note: Can photon-counting CT improve PET/CT's quantitative accuracy?

被引:1
|
作者
Hamill, James [1 ]
Birge, Noah [1 ]
O'Doherty, Jim [2 ,3 ]
Nye, Jonathon [4 ]
Elojeimy, Saeed [4 ]
机构
[1] Siemens Med Solut USA, 810 Innovat Dr Knoxville, Tennessee, IL 37923 USA
[2] Siemens Med Solut USA, CT R&D Collaborat, Malvern, PA USA
[3] Med Univ South Carolina, Dept Radiol & Radiol Sci, Charleston, SC USA
[4] Med Univ South Carolina, Dept Radiol & Radiol Sci, Dept Nucl Med, Charleston, SC USA
关键词
attenuation correction; PET; photon-counting CT; ATTENUATION CORRECTION; INTRAVENOUS CONTRAST;
D O I
10.1002/mp.17299
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundLinear attenuation coefficients (LACs) in positron emission tomography combined with computed tomography (PET/CT) are derived from CT scans that utilize energy-integrating detectors (EID-CT). These LACs are inaccurate when iodine contrast has been injected. Photon counting detector CT (PCD-CT) may be able to improve the accuracy.PurposeTo investigate whether PCD-CT can improve PET/CT quantitative accuracy.MethodsTwo experiments were performed: one with CT only and one that combined PET and CT. The first experiment used an electron density phantom whose inserts were imaged with EID-CT and PCD-CT. The inserts simulated normal human tissues, including bone and iodinated blood. In the case of PCD-CT, virtual-monoenergetic images at 190 keV were created. LACs were derived in each case and compared against known values. For inserts with iodine, more accurate LACs were expected with PCD-CT. The second experiment involved a custom PET phantom with various materials simulating human tissues (blood, iodinated blood, and bone) and 18F radioactivity. Data were first acquired with an EID-CT-based PET/CT system and then separately in a PCD-CT system without PET. PET images were reconstructed using LAC from EID-CT and PCD-CT. PET image values were compared against known activity values using recovery coefficients (RC).ResultsIn the first experiment, LAC based on EID-CT were in error by as much as 18%, whereas the corresponding PCD-CT based measurements were within 3%. In the second experiment, minimum, maximum, and mean RC were (96.1%, 115.4%, and 103.8%) for the EID-CT method, and (95.8%, 105.5%, and 101.0%) for the PCD-CT method. The consistency of PET images in body and head orientations was improved.ConclusionsPCD-CT can acquire the information needed for accurate LAC for PET reconstruction in a single spiral acquisition.
引用
收藏
页码:7206 / 7213
页数:8
相关论文
共 50 条
  • [31] A systematic assessment and optimization of photon-counting CT for lung density quantifications
    Sotoudeh-Paima, Saman
    Segars, W. Paul
    Ghosh, Dhrubajyoti
    Luo, Sheng
    Samei, Ehsan
    Abadi, Ehsan
    MEDICAL PHYSICS, 2024, 51 (04) : 2893 - 2904
  • [32] Dual-source photon-counting CT: impact of residual cross-scatter on quantitative spectral results
    Liu, Leening P.
    Salazar, Edgar
    Sahbaee, Pooyan
    Litt, Harold I.
    Noel, Peter B.
    MEDICAL IMAGING 2024: PHYSICS OF MEDICAL IMAGING, PT 1, 2024, 12925
  • [33] Optimization of energy thresholds in photon-counting CT via a virtual clinical trial
    Abadi, Ehsan
    Rajagopal, Jayasai
    Sauer, Thomas
    Jones, Elizabeth
    Segars, W. Paul
    Samei, Ehsan
    MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312
  • [34] Development and Clinical Applications of a Virtual Imaging Framework for Optimizing Photon-counting CT
    Abadi, Ehsan
    McCabe, Cindy
    Harrawood, Brian
    Sotoudeh-Paima, Saman
    Segars, W. Paul
    Samei, Ehsan
    MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [35] Learned Tensor Neural Network Texture Prior for Photon-Counting CT Reconstruction
    Shi, Yongyi
    Gao, Yongfeng
    Xu, Qiong
    Li, Yang
    Mou, Xuanqin
    Liang, Zhengrong
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2024, 43 (11) : 3830 - 3842
  • [36] Effect of PET-CT misalignment on the quantitative accuracy of cardiac 15O-water PET
    Nordstrom, Jonny
    Harms, Hendrik J.
    Kero, Tanja
    Ebrahimi, Maryam
    Sorensen, Jens
    Lubberink, Mark
    JOURNAL OF NUCLEAR CARDIOLOGY, 2022, 29 (03) : 1119 - 1128
  • [37] Potential dose reduction and image quality improvement in chest CT with a photon-counting CT compared to a new dual-source CT
    Greffier, Joel
    Dabli, Djamel
    Faby, Sebastian
    Pastor, Maxime
    de Oliveira, Fabien
    Croisille, Cedric
    Erath, Julien
    Beregi, Jean-Paul
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 127
  • [38] The effect of breathing irregularities on quantitative accuracy of respiratory gated PET/CT
    Teo, Boon-Keng
    Saboury, Babak
    Munbodh, Reshma
    Scheuermann, Joshua
    Torigian, Drew A.
    Zaidi, Habib
    Alavi, Abass
    MEDICAL PHYSICS, 2012, 39 (12) : 7390 - 7397
  • [39] Low-dose lung cancer screening with photon-counting CT: a feasibility study
    Symons, Rolf
    Cork, Tyler E.
    Sahbaee, Pooyan
    Fuld, Matthew K.
    Kappler, Steffen
    Folio, Les R.
    Bluemke, David A.
    Pourmorteza, Amir
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (01) : 202 - 213
  • [40] Photon-Counting Detector CT Angiography for Endoleak Detection After Endovascular Aortic Repair
    Turrion Gomollon, Ana Maria
    Mergen, Victor
    Sartoretti, Thomas
    Polacin, Malgorzata
    Nakhostin, Dominik
    Puippe, Gilbert
    Alkadhi, Hatem
    Euler, Andre
    INVESTIGATIVE RADIOLOGY, 2023, 58 (11) : 816 - 821