Technical note: Can photon-counting CT improve PET/CT's quantitative accuracy?

被引:1
|
作者
Hamill, James [1 ]
Birge, Noah [1 ]
O'Doherty, Jim [2 ,3 ]
Nye, Jonathon [4 ]
Elojeimy, Saeed [4 ]
机构
[1] Siemens Med Solut USA, 810 Innovat Dr Knoxville, Tennessee, IL 37923 USA
[2] Siemens Med Solut USA, CT R&D Collaborat, Malvern, PA USA
[3] Med Univ South Carolina, Dept Radiol & Radiol Sci, Charleston, SC USA
[4] Med Univ South Carolina, Dept Radiol & Radiol Sci, Dept Nucl Med, Charleston, SC USA
关键词
attenuation correction; PET; photon-counting CT; ATTENUATION CORRECTION; INTRAVENOUS CONTRAST;
D O I
10.1002/mp.17299
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundLinear attenuation coefficients (LACs) in positron emission tomography combined with computed tomography (PET/CT) are derived from CT scans that utilize energy-integrating detectors (EID-CT). These LACs are inaccurate when iodine contrast has been injected. Photon counting detector CT (PCD-CT) may be able to improve the accuracy.PurposeTo investigate whether PCD-CT can improve PET/CT quantitative accuracy.MethodsTwo experiments were performed: one with CT only and one that combined PET and CT. The first experiment used an electron density phantom whose inserts were imaged with EID-CT and PCD-CT. The inserts simulated normal human tissues, including bone and iodinated blood. In the case of PCD-CT, virtual-monoenergetic images at 190 keV were created. LACs were derived in each case and compared against known values. For inserts with iodine, more accurate LACs were expected with PCD-CT. The second experiment involved a custom PET phantom with various materials simulating human tissues (blood, iodinated blood, and bone) and 18F radioactivity. Data were first acquired with an EID-CT-based PET/CT system and then separately in a PCD-CT system without PET. PET images were reconstructed using LAC from EID-CT and PCD-CT. PET image values were compared against known activity values using recovery coefficients (RC).ResultsIn the first experiment, LAC based on EID-CT were in error by as much as 18%, whereas the corresponding PCD-CT based measurements were within 3%. In the second experiment, minimum, maximum, and mean RC were (96.1%, 115.4%, and 103.8%) for the EID-CT method, and (95.8%, 105.5%, and 101.0%) for the PCD-CT method. The consistency of PET images in body and head orientations was improved.ConclusionsPCD-CT can acquire the information needed for accurate LAC for PET reconstruction in a single spiral acquisition.
引用
收藏
页码:7206 / 7213
页数:8
相关论文
共 50 条
  • [1] Can Photon-Counting CT Improve Estimation Accuracy of Morphological Radiomics Features? A Simulation Study for Assessing the Quantitative Benefits from Improved Spatial Resolution in Deep Silicon-Based Photon-Counting CT
    Sharma, Shobhit
    Pal, Debashish
    Abadi, Ehsan
    Sauer, Thomas
    Segars, Paul
    Hsieh, Jiang
    Samei, Ehsan
    ACADEMIC RADIOLOGY, 2023, 30 (06) : 1153 - 1163
  • [2] Technical Basics and Clinical Benefits of Photon-Counting CT
    Flohr, Thomas
    Schmidt, Bernhard
    INVESTIGATIVE RADIOLOGY, 2023, 58 (07) : 441 - 450
  • [3] Photon-counting CT review
    Flohr, Thomas
    Petersilka, Martin
    Henning, Andre
    Ulzheimer, Stefan
    Ferda, Jiri
    Schmidt, Bernhard
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 79 : 126 - 136
  • [4] Photon-counting CT systems: A technical review of current clinical possibilities
    Greffier, Joel
    Viry, Anais
    Robert, Antoine
    Khorsi, Mouad
    Si-Mohamed, Salim
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2025, 106 (02) : 53 - 59
  • [5] Photon-counting CT: technical features and clinical impact on abdominal imaging
    Onishi, Hiromitsu
    Tsuboyama, Takahiro
    Nakamoto, Atsushi
    Ota, Takashi
    Fukui, Hideyuki
    Tatsumi, Mitsuaki
    Honda, Toru
    Kiso, Kengo
    Matsumoto, Shohei
    Kaketaka, Koki
    Enchi, Yukihiro
    Kawabata, Shuichi
    Nakasone, Shinya
    Tomiyama, Noriyuki
    ABDOMINAL RADIOLOGY, 2024, 49 (12) : 4383 - 4399
  • [6] Abdominal applications of photon-counting CT
    Khanungwanitkul, Khanin
    Sliwicka, Olga
    Schwartz, Fides Regina
    BRITISH JOURNAL OF RADIOLOGY, 2024,
  • [7] Spectral CT imaging: Technical principles of dual-energy CT and multi-energy photon-counting CT
    Greffier, Joel
    Villani, Nicolas
    Defez, Didier
    Dabli, Djamel
    Si-Mohamed, Salim
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2023, 104 (04) : 167 - 177
  • [8] Advanced Neuroimaging With Photon-Counting Detector CT
    Abel, Frederik
    Schubert, Tilman
    Winklhofer, Sebastian
    INVESTIGATIVE RADIOLOGY, 2023, 58 (07) : 472 - 481
  • [9] Pediatric Applications of Photon-Counting Detector CT
    Cao, Joseph
    Bache, Steve
    Schwartz, Fides R.
    Frush, Donald
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2023, 220 (04) : 580 - 589
  • [10] Chances and challenges of photon-counting CT in musculoskeletal imaging
    Mourad, Charbel
    Gallego Manzano, Lucia
    Viry, Anais
    Booij, Ronald
    Oei, Edwin H. G.
    Becce, Fabio
    Omoumi, Patrick
    SKELETAL RADIOLOGY, 2024, 53 (09) : 1889 - 1902