Disinfection of needleless connectors for catheters in one second using a hand-held UV device

被引:1
作者
Fourkas, Michael [1 ]
Takami, Eri [1 ]
Schears, Gregory J. [2 ]
Farr-Jones, Shauna [3 ]
Rasooly, Julia [1 ]
机构
[1] PuraCath Med Inc, 37600 Cent Court,Suite 210, Newark, CA 94560 USA
[2] Mayo Clin, Dept Anesthesiol & Perioperat Med, Rochester, MN USA
[3] Univ Calif San Francisco, Zuckerberg San Francisco Hosp, Dept Anesthesia & Perioperat Care, San Francisco, CA USA
关键词
Catheter disinfection; UV-disinfection; CLABSI prevention; Ultraviolet; Catheter line associated bloodstream; infection; INACTIVATION; LIGHT;
D O I
10.1016/j.ajic.2024.03.017
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: The standard of care for disinfecting needleless connectors (NCs) of central venous catheters includes alcohol-containing caps or up to a 15-second scrub with alcohol or chlorhexidine. Due to the clinical impact and high cost of treating Central line-associated bloodstream infections (CLABSI), reducing the incidence of CLABSI is a priority for public health and of the Centers for Disease Control. Alcohol-containing caps have been demonstrated to disinfect external NC surfaces, but not the internal surface. Ultraviolet light (UV-C) is a strategy for disinfection of NC internal and external surfaces. Methods: Four clinically relevant bacteria (Staphylococcus aureus, Staphylococcus epidermidis, Escherichia coli, Pseudomonas aeruginosa) and Candida albicans were inoculated on NCs. Disinfection efficacy was measured after exposure to one second of 285 nm UV-C light at 48 mW/cm2 in a proprietary handheld device and UV-C transparent NC or standard of care. Disinfection of internal and external surfaces of NC inoculated with S aureus using alcohol caps, and UV-C was also compared. Results: A 4-log reduction in colony forming units (CFUs) on the interior and exterior surfaces of the UVtransparent NC of clinically relevant pathogens was observed with UV-C light at this power for 1 second. Discussion: We demonstrated the efficacy of UV-C for the disinfection of NCs in one second using the UV-C device in benchtop studies. Conclusions: This device holds promise for reducing CLABSI, and clinical studies are planned. Published by Elsevier Inc. on behalf of Association for Professionals in Infection Control and Epidemiology, Inc. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
引用
收藏
页码:915 / 918
页数:4
相关论文
共 23 条
  • [2] [Anonymous], National and state healthcare-associated infections report
  • [3] [Anonymous], HOSP ACQUIRED CONDIT
  • [4] Effect of UV Light on Disinfection of Peritoneal Dialysis Catheter Connections
    Ashley, John
    Rasooly, Julia A.
    Tran, Ian
    Yost, Lawrence E.
    Chertow, Glenn M.
    [J]. PERITONEAL DIALYSIS INTERNATIONAL, 2017, 37 (01): : 109 - U164
  • [5] Bolton JR., 2011, ULTRAVIOLET DISINFEC, V1st
  • [6] UV INACTIVATION OF PATHOGENIC AND INDICATOR MICROORGANISMS
    CHANG, JCH
    OSSOFF, SF
    LOBE, DC
    DORFMAN, MH
    DUMAIS, CM
    QUALLS, RG
    JOHNSON, JD
    [J]. APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 1985, 49 (06) : 1361 - 1365
  • [7] The colonization rate of needleless connector and the impact of disinfection for 15 s on colonization: A prospective pre- and post-intervention study
    Devrim, Ilker
    Demiray, Nevbahar
    Oruc, Yeliz
    Sipahi, Kenan
    Caglar, Ilknur
    Sari, Ferhat
    Turgut, Nuriye
    Atakul, Gulhan
    Ozdamar, Nihal
    Dursun, Vecihe
    Sorguc, Yelda
    Bayram, Nuri
    Agin, Hasan
    [J]. JOURNAL OF VASCULAR ACCESS, 2019, 20 (06) : 604 - 607
  • [8] Gahlot R, 2014, Int J Crit Illness Injury Sci., V4, P160
  • [9] Visualization of Biofilm Formation in Candida albicans Using an Automated Microfluidic Device
    Gulati, Megha
    Ennis, Craig L.
    Rodriguez, Diana L.
    Nobile, Clarissa J.
    [J]. JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2017, (130):
  • [10] Health Care-Associated Infections, National HAI targets & metrics