UNDERSTANDING THE BIOMECHANICAL RESPONSE OF PROGRESSIVE THREAD DENTAL IMPLANTS USING MULTI-SCALE FINITE ELEMENT ANALYSIS

被引:0
作者
Chakraborty, Arindam [1 ]
Sahare, Kalash Darshan [1 ,2 ]
Majumder, Santanu [1 ]
Chowdhury, Amit Roy [1 ]
机构
[1] Indian Inst Engn Sci & Technol, Dept Aerosp Engn & Appl Mech, Sibpur 711103, W Bengal, India
[2] Indian Inst Technol, Dept Aerosp Engn, Kanpur, Uttar Pradesh, India
关键词
micro finite element analysis; dental implant; thread depth; bone strain; implant stress; MICRO-CT; STRESS-DISTRIBUTION; CORTICAL BONE; ELASTIC PROPERTIES; DESIGN; FIXATION; QUALITY; FE;
D O I
暂无
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Implant osseointegration is an important factor dictating its long-term efficacy in situ. Along with various biological factors, it is greatly influenced by the mechanical stimulus at the peri-implant bone. The present study aims to understand the biomechanical response of progressive thread dental implants using multi-scale-based finite element analysis employing macro and micro models of bone. mu-CT images of a cadaveric human mandible of its premolar region were obtained, along with CT scan of the same region to generate computational models. Total of six dental implants were designed having regular and progressive thread depths. Three different stages of healing of the bone-implant assembly were simulated parametrically. The biomechanical environment at the peri-implant bone was analyzed considering the 'Mechanostat' hypothesis. The obtained results revealed that bone strain is significantly higher during the initial healing phase when the bone is weakest. During this phase, implant stress and its displacement in both buccolingual and coronoapical directions are also noticeably higher. Also, displacements of progressive thread implants were lower in all the healing phases as compared to the implants with constant thread depth. The observations of this mu FEA study highlights the clinical applicability of a progressive thread dental implant as it generates larger functional surface area, thus engages higher trabeculae and therefore is suitable for weaker bone conditions. Furthermore, by comparing the stress values at bone and implant between the two bone models, the CT-based model having inhomogeneous material was deemed suitable as an alternative to computationally expensive mu FEA.
引用
收藏
页码:31 / 44
页数:14
相关论文
共 50 条
[21]   BIOMECHANICAL EVALUATION OF MANDIBULAR OVERDENTURES SUPPORTED BY MINI-IMPLANTS: A FINITE ELEMENT ANALYSIS [J].
Ganesh, Thiagarajan ;
Scully, Jessica ;
Walker, Mary P. ;
Petrie, Cynthia S. .
JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2022, 22 (06)
[22]   Micro finite element analysis of dental implants under different loading conditions [J].
Marcian, Petr ;
Wolff, Jan ;
Horackova, Ladislava ;
Kaiser, Jozef ;
Zikmund, Tomas ;
Borak, Libor .
COMPUTERS IN BIOLOGY AND MEDICINE, 2018, 96 :157-165
[23]   Stress-based performance comparison of dental implants by finite element analysis [J].
Pasquale Franciosa ;
Massimo Martorelli .
Martorelli, M. (massimo.martorelli@unina.it), 1600, Springer-Verlag France (06) :123-129
[24]   Design Optimization of Dental Implants Using Finite Element Analysis for Injecting Bioactive Materials [J].
Lee, Kangsoo ;
Lee, Yong Keun .
KOREAN JOURNAL OF MATERIALS RESEARCH, 2012, 22 (06) :292-297
[25]   Selection of the implant thread pitch for optimal biomechanical properties: A three-dimensional finite element analysis [J].
Kong, Liang ;
Zhao, Yunzhuan ;
Hu, Kaijin ;
Li, Dehua ;
Zhou, Hongzhi ;
Wu, Ziyan ;
Liu, Baolin .
ADVANCES IN ENGINEERING SOFTWARE, 2009, 40 (07) :474-478
[26]   Progressive damage analysis of functionally graded dental implants due to pre-tightening loads using extended finite element method [J].
Mohammadi Anari, Ali ;
Selahi, Ehsan .
JOURNAL OF THE BRAZILIAN SOCIETY OF MECHANICAL SCIENCES AND ENGINEERING, 2022, 44 (07)
[27]   Effects of assessing the bone remodeling process in biomechanical finite element stability evaluations of dental implants [J].
Park, Soyeon ;
Park, Jieun ;
Kang, Inyeong ;
Lee, Hyeonjong ;
Noh, Gunwoo .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2022, 221
[28]   Multi-Scale Finite Element Models in Bridge Field Tests [J].
Tian, Jun ;
Wang, Yelu .
CICTP 2021: ADVANCED TRANSPORTATION, ENHANCED CONNECTION, 2021, :874-883
[29]   Validation of the Finite Element Model versus Biomechanical Assessments of Dental Implants and Total Knee Replacements [J].
Kang, Kwan-Su ;
Park, Kwang-Min ;
Ahn, Jin-Woo ;
Jo, Min-Young ;
Oh, Yu-Rim ;
Youn, Jin-Ho ;
Lee, Jeong-Woo ;
Je, Dong-Young ;
Jung, Tae-Gon .
BIOENGINEERING-BASEL, 2023, 10 (12)
[30]   Biomechanical analysis of peri-prosthetic bone response to hybrid threaded zirconia dental implants: An in silico model [J].
Mishra, Deepa ;
Basu, Bikramjit .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2024, 150