High-resolution single-particle imaging at 100-200 keV with the Gatan Alpine direct electron detector

被引:2
|
作者
Chan, Lieza M. [1 ]
Courteau, Brandon J. [1 ]
Maker, Allison [1 ]
Wu, Mengyu [3 ,5 ]
Basanta, Benjamin [3 ,6 ]
Mehmood, Hev [1 ]
Bulkley, David [2 ]
Joyce, David [4 ]
Lee, Brian C. [4 ]
Mick, Stephen [4 ]
Czarnik, Cory [4 ]
Gulati, Sahil [4 ]
Lander, Gabriel C. [3 ]
Verba, Kliment A. [1 ]
机构
[1] Univ Calif San Francisco, Dept Cellular & Mol Pharmacol, San Francisco, CA 94158 USA
[2] Univ Calif San Francisco, Dept Biochem & Biophys, San Francisco, CA 94158 USA
[3] Scripps Res, Dept Integrat Struct & Computat Biol, La Jolla, CA 92037 USA
[4] Gatan Inc, Pleasanton, CA USA
[5] Neomorph, San Diego, CA USA
[6] Arzeda, Seattle, WA USA
关键词
Single -particle cryo-EM; Direct detectors; Advances in microscope hardware; 100 keV imaging; Gatan Alpine; LKB1; complex; BEAM-INDUCED MOTION; CRYO-EM; RADIATION-DAMAGE; MICROSCOPY; VISUALIZATION; LIMITATIONS; VALIDATION; CONTRAST; TOOLS;
D O I
10.1016/j.jsb.2024.108108
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Developments in direct electron detector technology have played a pivotal role in enabling high-resolution structural studies by cryo-EM at 200 and 300 keV. Yet, theory and recent experiments indicate advantages to imaging at 100 keV, energies for which the current detectors have not been optimized. In this study, we evaluated the Gatan Alpine detector, designed for operation at 100 and 200 keV. Compared to the Gatan K3, Alpine demonstrated a significant DQE improvement at these energies, specifically a- 4-fold improvement at Nyquist at 100 keV. In single-particle cryo-EM experiments, Alpine datasets yielded better than 2 & Aring; resolution reconstructions of apoferritin at 120 and 200 keV on a ThermoFisher Scientific (TFS) Glacios microscope fitted with a non-standard SP-Twin lens. We also achieved a- 3.2 & Aring; resolution reconstruction of a 115 kDa asymmetric protein complex, proving Alpine 's effectiveness with complex biological samples. In-depth analysis revealed that Alpine reconstructions are comparable to K3 reconstructions at 200 keV, and remarkably, reconstruction from Alpine at 120 keV on a TFS Glacios surpassed all but the 300 keV data from a TFS Titan Krios with GIF/K3. Additionally, we show Alpine 's capability for high-resolution data acquisition and screening on lower-end systems by obtaining- 3 & Aring; resolution reconstructions of apoferritin and aldolase at 100 keV and detailed 2D averages of a 55 kDa sample using a side-entry cryo holder. Overall, we show that Gatan Alpine performs well with the standard 200 keV imaging systems and may potentially capture the benefits of lower accelerating voltages, bringing smaller sized particles within the scope of cryo-EM.
引用
收藏
页数:13
相关论文
共 50 条
  • [41] DIRECT MOLECULAR IMAGING OF LOW DIMENSIONAL SOLIDS BY HIGH-RESOLUTION ELECTRON-MICROSCOPY
    UYEDA, N
    KOBAYASHI, T
    ISHIZUKA, K
    FUJIYOSHI, Y
    INOKUCHI, H
    SAITO, G
    MOLECULAR CRYSTALS AND LIQUID CRYSTALS, 1985, 125 (1-4): : 103 - 112
  • [42] DIRECT IMAGING OF CDTE(001) SURFACE RECONSTRUCTIONS BY HIGH-RESOLUTION ELECTRON-MICROSCOPY
    LU, P
    SMITH, DJ
    SURFACE SCIENCE, 1991, 254 (1-3) : 119 - 124
  • [43] Detector Characterization of a High-Resolution Ring for PET Imaging of Mice Heads With Sub-200-ps TOF
    Valladares, Celia
    Barrio, John
    Cucarella, Neus
    Freire, Marta
    Vidal, Luis F.
    Benlloch, Jose M.
    Gonzalez, Antonio J.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2024, 8 (08) : 876 - 885
  • [44] A Fast and Effective Microfluidic Spraying-Plunging Method for High-Resolution Single-Particle Cryo-EM
    Feng, Xiangsong
    Fu, Ziao
    Kaledhonkar, Sandip
    Jia, Yuan
    Shah, Binita
    Jin, Amy
    Liu, Zheng
    Sun, Ming
    Chen, Bo
    Grassucci, Robert A.
    Ren, Yukun
    Jiang, Hongyuan
    Frank, Joachim
    Lin, Qiao
    STRUCTURE, 2017, 25 (04) : 663 - +
  • [45] High-resolution single-particle 3D analysis on GroEL prepared by cryo-negative staining
    De Carlo, Sacha
    Boisset, Nicolas
    Hoenger, Andreas
    MICRON, 2008, 39 (07) : 934 - 943
  • [46] High-resolution, single-particle digital autoradiography of actinide sources using microcapillary array collimators and the iQID camera
    Miller, Brian W.
    Bowen, James M.
    Morrison, Erin C.
    APPLIED RADIATION AND ISOTOPES, 2020, 166
  • [47] High-resolution structure of phosphoketolase from Bifidobacterium longum determined by cryo-EM single-particle analysis
    Nakata, Kunio
    Miyazaki, Naoyuki
    Yamaguchi, Hiroki
    Hirose, Mika
    Kashiwagi, Tatsuki
    Kutumbarao, Nidamarthi H., V
    Miyashita, Osamu
    Tama, Florence
    Miyano, Hiroshi
    Mizukoshi, Toshimi
    Iwasaki, Kenji
    JOURNAL OF STRUCTURAL BIOLOGY, 2022, 214 (02)
  • [48] Symmetry-adapted spherical harmonics method for high-resolution 3D single-particle reconstructions
    Liu, Hongrong
    Cheng, Lingpeng
    Zeng, Songjun
    Cai, Canying
    Zhou, Z. Hong
    Yang, Qibin
    JOURNAL OF STRUCTURAL BIOLOGY, 2008, 161 (01) : 64 - 73
  • [49] Ab initio high-resolution single-particle 3D reconstructions: The symmetry adapted functions way
    Estrozi, Leandro F.
    Navaza, Jorge
    JOURNAL OF STRUCTURAL BIOLOGY, 2010, 172 (03) : 253 - 260
  • [50] Is radiation damage the limiting factor in high-resolution single particle imaging with X-ray free-electron lasers?
    Ostlin, C.
    Timneanu, N.
    Caleman, C.
    Martin, A. V.