Frequency-Difference Electrical Impedance Imaging of Breast Specimens

被引:0
作者
Ouypornkochagorn, Taweechai [1 ]
Ngamdi, Napatsawan [1 ]
Buathong, Piyathida [1 ]
Laor-Iam, Phanpasorn [1 ]
Phisaiphan, Aiyada [1 ]
Ouypornkochagorn, Sairoong [2 ]
Sriwilai, Jaruwan [3 ]
Trongwongsa, Therdkiat [3 ]
机构
[1] Srinakharinwirot Univ, Fac Engn, Bangkok 26120, Thailand
[2] Naresuan Univ, Fac Sci, Phitsanulok 65000, Thailand
[3] Srinakharinwirot Univ, Fac Med, Dept Pathol, Bangkok 26120, Thailand
关键词
Conductivity; Electrodes; Voltage measurement; Probes; Electrical impedance tomography; Pathology; Imaging; Breast; conductivity; electrode probe; frequency-difference electrical impedance tomography (fdEIT); pathological specimens; DIELECTRIC-PROPERTIES; CANCER; TISSUES; SYSTEM; MODELS; WOMEN;
D O I
10.1109/TIM.2024.3418087
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Imaging abnormality in pathological breast specimens can greatly help pathologists select sites for conducting pathological examination through a microscope. In this study, the frequency-difference electrical impedance tomography (fdEIT) technique was proposed for localizing the abnormality in eight breast specimens. Three 16-electrode probes were developed for this purpose which worked with a high-signal-to-noise ratio (SNR) multifrequency electrical impedance tomography (EIT) system. The frequencies of 10 and 50 kHz were selected. The conductivity of the carcinoma masses was found 2.7-3.8 times higher than that of adipose tissues and 2%-21% higher than that of fibroglandular tissues. The conductivity difference of the masses over the selected frequencies was 2.5-3.5 times higher than that of the surrounding tissues. The probes could sense to a depth of similar to 12 mm from the specimen surface. Five of seven specimens (71.4%) were achieved to visualize the masses. Extremely and heterogeneously dense specimens provided better mass distinction in the images than the scattered dense specimens. A mass with a diameter over 20 mm and near the measuring surface was preferable for detection. However, the mass distinction in the images also depends on the conductivity of the mass itself. Mass could be hard and soft, and the soft masses could have different conductivity. The high conductive mass tended to have a better mass distinction in the images.
引用
收藏
页数:11
相关论文
共 45 条
  • [1] Akhtari-Zavare Mehrnoosh, 2015, Asian Pac J Cancer Prev, V16, P5595
  • [2] Borsic A, 2002, THESIS OXFORD BROOKE
  • [3] ELECTRODE MODELS FOR ELECTRIC-CURRENT COMPUTED-TOMOGRAPHY
    CHENG, KS
    ISAACSON, D
    NEWELL, JC
    GISSER, DG
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (09) : 918 - 924
  • [4] A 3D electrical impedance tomography (EIT) system for breast cancer detection
    Cherepenin, V
    Karpov, A
    Korjenevsky, A
    Kornienko, V
    Mazaletskaya, A
    Mazourov, D
    Meister, D
    [J]. PHYSIOLOGICAL MEASUREMENT, 2001, 22 (01) : 9 - 18
  • [5] Three-dimensional EIT imaging of breast tissues: System design and clinical testing
    Cherepenin, VA
    Karpov, AY
    Korjenevsky, AV
    Kornienko, VN
    Kultiasov, YS
    Ochapkin, MB
    Trochanova, OV
    Meister, JD
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (06) : 662 - 667
  • [6] Classification of breast tissue by electrical impedance spectroscopy
    da Silva, JE
    de Sá, JPM
    Jossinet, J
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2000, 38 (01) : 26 - 30
  • [7] Bioimpedance spectroscopy can precisely discriminate human breast carcinoma from benign tumors
    Du, Zhenggui
    Wan, Hangyu
    Chen, Yu
    Pu, Yang
    Wang, Xiaodong
    [J]. MEDICINE, 2017, 96 (04)
  • [8] Factors limiting the application of electrical impedance tomography for identification of regional conductivity changes using scalp electrodes during epileptic seizures in humans
    Fabrizi, L.
    Sparkes, M.
    Horesh, L.
    Abascal, J. F. Perez-Juste
    McEwan, A.
    Bayford, R. H.
    Elwes, R.
    Binnie, C. D.
    Holder, D. S.
    [J]. PHYSIOLOGICAL MEASUREMENT, 2006, 27 (05) : S163 - S174
  • [9] Chest electrical impedance tomography examination, data analysis, terminology, clinical use and recommendations: consensus statement of the TRanslational EIT developmeNt stuDy group
    Frerichs, Inez
    Amato, Marcelo B. P.
    van Kaam, Anton H.
    Tingay, David G.
    Zhao, Zhanqi
    Grychtol, Bartlomiej
    Bodenstein, Marc
    Gagnon, Herve
    Bohm, Stephan H.
    Teschner, Eckhard
    Stenqvist, Ola
    Mauri, Tommaso
    Torsani, Vinicius
    Camporota, Luigi
    Schibler, Andreas
    Wolf, Gerhard K.
    Gommers, Diederik
    Leonhardt, Steffen
    Adler, Andy
    [J]. THORAX, 2017, 72 (01) : 83 - 93
  • [10] The dielectric properties of biological tissues .2. Measurements in the frequency range 10 Hz to 20 GHz
    Gabriel, S
    Lau, RW
    Gabriel, C
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) : 2251 - 2269