Performance of cone-beam computed tomography imaging during megavoltage beam irradiation under phase-gated conditions

被引:0
作者
Iramina, Hiraku [1 ]
Nakamura, Mitsuhiro [2 ]
Sasaki, Makoto [3 ]
Mizowaki, Takashi [1 ]
机构
[1] Kyoto Univ, Grad Sch Med, Dept Radiat Oncol & Image Appl Therapy, 54 Kawahara cho,Sakyo ku, Kyoto 6068507, Japan
[2] Kyoto Univ, Grad Sch Med, Dept Adv Med Phys, 53 Kawahara cho,Sakyo ku, Kyoto 6068507, Japan
[3] Kyoto Univ Hosp, Div Clin Radiol Serv, 54 Kawahara cho,Sakyo ku, Kyoto 6068507, Japan
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2024年 / 123卷
关键词
CBCT acquired during beam delivery; Lung SBRT; Gating; Dose reduction; STEREOTACTIC BODY RADIOTHERAPY; CELL LUNG-CANCER; MOTION MANAGEMENT; TUMOR TRACKING; EXPERIENCE; POSITION;
D O I
10.1016/j.ejmp.2024.103409
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Target positions should be acquired during beam delivery for accurate lung stereotactic body radiotherapy. We aimed to perform kilovoltage (kV) imaging during beam irradiation (intra-irradiation imaging) under phase-gated conditions and evaluate its performance. Methods: Catphan 504 and QUASAR respiratory motion phantoms were used to evaluate image quality and target detectability, respectively. TrueBeam STx linac and the Developer Mode was used. The imaging parameters were 125 kVp and 1.2 mAs/projection. Flattened megavoltage (MV) X-ray beam energies 6, 10 and 15 MV and unflattened beam energies 6 and 10 MV were used with field sizes of 5 x 5 and 15 x 15 cm(2) and various frame rates for intra-irradiation imaging. In addition, using a QUASAR phantom, intra-irradiation imaging was performed during intensity-modulated plan delivery. The root-mean-square error (RMSE) of the CT-number for the inserted rods, image noise, visual assessment, and contrast-to-noise ratio (CNR) were evaluated. Results: The RMSEs of intra-irradiation cone-beam computed tomography (CBCT) images under gated conditions were 50-230 Hounsfield Unit (HU) (static < 30 HU). The noise of the intra-irradiation CBCT images under gated conditions was 15-35 HU, whereas that of the standard CBCT images was 8.8-27.2 HU. Lower frame rates exhibited large RMSEs and noise; however, the iterative reconstruction algorithm (IR) was effective at improving these values. Approximately 7 fps with the IR showed an equivalent CNR of 15 fps without the IR. The target was visible on all the gated intra-irradiation CBCT images. Conclusion: Several image quality improvements are required; however, intra-irradiated CBCT images showed good visual target detection.
引用
收藏
页数:11
相关论文
共 29 条
[1]   Outcome in a Prospective Phase II Trial of Medically Inoperable Stage I Non-Small-Cell Lung Cancer Patients Treated With Stereotactic Body Radiotherapy [J].
Baumann, Pia ;
Nyman, Jan ;
Hoyer, Morten ;
Wennberg, Berit ;
Gagliardi, Giovanna ;
Lax, Ingmar ;
Drugge, Ninni ;
Ekberg, Lars ;
Friesland, Signe ;
Johansson, Karl-Axel ;
Lund, Jo-Asmund ;
Morhed, Elisabeth ;
Nilsson, Kristina ;
Levin, Nina ;
Paludan, Merete ;
Sederholm, Christer ;
Traberg, Anders ;
Wittgren, Lena ;
Lewensohn, Rolf .
JOURNAL OF CLINICAL ONCOLOGY, 2009, 27 (20) :3290-3296
[2]   Motion management strategies and technical issues associated with stereotactic body radiotherapy of thoracic and upper abdominal tumors: A review from NRG oncology [J].
Brandner, Edward D. ;
Chetty, Indrin J. ;
Giaddui, Tawfik G. ;
Xiao, Ying ;
Huq, M. Saiful .
MEDICAL PHYSICS, 2017, 44 (06) :2595-2612
[3]   IGRT and motion management during lung SBRT delivery [J].
Caillet, Vincent ;
Booth, Jeremy T. ;
Keall, Paul .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 44 :113-122
[4]   First study of on-treatment volumetric imaging during respiratory gated VMAT [J].
Choi, Kihwan ;
Xing, Lei ;
Koong, Albert ;
Lia, Ruijiang .
MEDICAL PHYSICS, 2013, 40 (04)
[5]   Original MR-guided beam gating: Residual motion, gating efficiency and dose reconstruction for stereotactic treatments of the liver and lung [J].
Ehrbar, Stefanie ;
Kaeser, Sarah Braga ;
Chamberlain, Madalyne ;
Krayenbuehl, Jerome ;
Wilke, Lotte ;
Mayinger, Michael ;
Schueler, Helena Garcia ;
Guckenberger, Matthias ;
Andratschke, Nicolaus ;
Tanadini-Lang, Stephanie .
RADIOTHERAPY AND ONCOLOGY, 2022, 174 :101-108
[6]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[7]   Does Motion Management Technique for Lung SBRT Influence Local Control? A Single Institutional Experience Comparing Abdominal Compression to Breath-Hold Technique [J].
Gandhidasan, Senthilkumar ;
Woody, Neil M. ;
Stephans, Kevin L. ;
Videtic, Gregory M. M. .
PRACTICAL RADIATION ONCOLOGY, 2021, 11 (02) :E180-E185
[8]   Verifying tumor position during stereotactic body radiation therapy delivery using (limited-arc) cone beani computed tomography imaging [J].
Hazelaar, Cohen ;
Dahele, Max ;
Scheib, Stefan ;
Slotman, Ben J. ;
Verbakel, Wilko F. A. R. .
RADIOTHERAPY AND ONCOLOGY, 2017, 123 (03) :355-362
[9]   Localization of deformable tumors from short-arc projections using Bayesian estimation [J].
Hoegele, W. ;
Zygmanski, P. ;
Dobler, B. ;
Kroiss, M. ;
Koelbl, O. ;
Loeschel, R. .
MEDICAL PHYSICS, 2012, 39 (12) :7205-7214
[10]   VOLUMETRIC-MODULATED ARC THERAPY FOR STEREOTACTIC BODY RADIOTHERAPY OF LUNG TUMORS: A COMPARISON WITH INTENSITY-MODULATED RADIOTHERAPY TECHNIQUES [J].
Holt, Andrea ;
van Vliet-Vroegindeweij, Corine ;
Mans, Anton ;
Belderbos, Jose S. ;
Damen, Eugene M. F. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (05) :1560-1567