- OBJECTIVE: Three-dimensional (3D) printed models are used in the medical field. This study aimed to evaluate the feasibility and safety of a 3D-printed guide plate for use in brain biopsy. - METHODS: Twelve patients with intracranial lesions were retrospectively reviewed to determine clinical outcomes and technical procedural operability. These patients - nderwent brain biopsy assisted with the 3D-printed guide plate. Postoperative computed tomography was performed to assess the accuracy and associated complications of this guide plate. - RESULTS: All patients received definite diagnoses assisted by this guide plate. The deviations of the entry and target points were 3.93 + 0.96 mm and 2.59 + 0.11 mm, respectively. The angle drift of the puncture path was 5.12 degrees + 0.14 degrees , and the deviation of the puncture depth was 2.35 + 1.13 mm. The operation time ranged from 38.5 minutes with local anesthesia to 76.2 minutes with general anesthesia. No patient experienced complications. - CONCLUSIONS: The 3D-printed guide plate was noninvasive and had acceptable accuracy and the flexibility of frameless systems. The economic and operative benefits of this device supported its status as a powerful tool for brain biopsy in medical facilities in economically disadvantaged areas or institutions without navigation systems.
引用
收藏
页码:e649 / e657
页数:9
相关论文
共 22 条
  • [1] Image-guided frameless stereotactic needle biopsy in awake patients without the use of rigid head fixation
    Amin, Devin V.
    Lozanne, Karl
    Parry, Phillip V.
    Engh, Johnathan A.
    Seelman, Kathleen
    Mintz, Arlan
    [J]. JOURNAL OF NEUROSURGERY, 2011, 114 (05) : 1414 - 1420
  • [2] Comparison of accuracy and precision between frame-based and frameless stereotactic navigation for deep brain stimulation electrode implantation
    Bjartmarz, Hjalmar
    Rehncrona, Stig
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2007, 85 (05) : 235 - 242
  • [3] The use of custom 3D printed stereotactic frames for laser interstitial thermal ablation: technical note
    Brandmeir, Nicholas J.
    McInerney, James
    Zacharia, Brad E.
    [J]. NEUROSURGICAL FOCUS, 2016, 41 (04)
  • [4] 3D-HA Scaffold Functionalized by Extracellular Matrix of Stem Cells Promotes Bone Repair
    Chi, Hui
    Chen, Guanghua
    He, Yixin
    Chen, Guanghao
    Tu, Hualei
    Liu, Xiaoqi
    Yan, Jinglong
    Wang, Xiaoyan
    [J]. INTERNATIONAL JOURNAL OF NANOMEDICINE, 2020, 15 : 5825 - 5838
  • [5] Clinical Accuracy of a Customized Stereotactic Platform for Deep Brain Stimulation after Accounting for Brain Shift
    D'Haese, Pierre-Francois
    Pallavaram, Srivatsan
    Konrad, Peter E.
    Neimat, Joseph
    Fitzpatrick, J. Michael
    Dawant, Benoit M.
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2010, 88 (02) : 81 - 87
  • [6] Frameless stereotactic brain biopsy: A prospective study on robot-assisted brain biopsies performed on 32 patients by using the RONNA G4 system
    Dlaka, Domagoj
    Svaco, Marko
    Chudy, Darko
    Jerbic, Bojan
    Sekoranja, Bojan
    Suligoj, Filip
    Vidakovic, Josip
    Romic, Dominik
    Raguz, Marina
    [J]. INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2021, 17 (03)
  • [7] Frameless stereotactic neurosurgery: Two steps towards the holy grail of surgical navigation
    Eljamel, MS
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1999, 72 (2-4) : 125 - 128
  • [8] Efficacy, Safety, and Duration of a Frameless Fiducial-Less Brain Biopsy versus Frame-based Stereotactic Biopsy: A Prospective Randomized Study
    Georgiopoulos, Miltiadis
    Ellul, John
    Chroni, Elisabeth
    Constantoyannis, Constantine
    [J]. JOURNAL OF NEUROLOGICAL SURGERY PART A-CENTRAL EUROPEAN NEUROSURGERY, 2018, 79 (01) : 31 - 38
  • [9] Blurring the boundaries between frame-based and frameless stereotaxy: feasibility study for brain biopsies performed with the use of a head-mounted robot
    Grimm, Florian
    Naros, Georgios
    Gutenberg, Angelika
    Keric, Naureen
    Giese, Alf
    Gharabaghi, Alireza
    [J]. JOURNAL OF NEUROSURGERY, 2015, 123 (03) : 737 - 742
  • [10] Kesserwan Mohamad Ali, 2021, Surg Neurol Int, V12, P52, DOI 10.25259/SNI_824_2020