Biomechanical testing of a computationally optimized far cortical locking plate versus traditional implants for distal femur fracture repair

被引:2
作者
Brzozowski, Pawel [1 ]
Inculet, Clayton [2 ]
Schemitsch, Emil H. [1 ,2 ]
Zdero, Radovan [1 ]
机构
[1] Victoria Hosp, Orthopaed Biomech Lab, Room A6-144,800 Commissioners Rd, London, ON N6A 5W9, Canada
[2] Western Univ, Div Orthopaed Surg, London, ON, Canada
关键词
Biomechanics; Optimization; Distal femur fracture; Far cortical locking; INTERFRAGMENTARY MOVEMENT; MECHANICAL-PROPERTIES; WORKING LENGTH; INITIAL PHASE; FIXATION; BONE; SHEAR; NAIL;
D O I
10.1016/j.clinbiomech.2024.106296
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: This study experimentally validated a computationally optimized screw number and screw distribution far cortical locking distal femur fracture plate and compared the results to traditional implants. Methods: 24 artificial femurs were osteotomized with a 10 mm fracture gap 60 mm proximal to the intercondylar notch. Three fixation constructs were used. (i) Standard locking plates secured with three far cortical locking screws inserted according to a previously optimized distribution in the femur shaft (n = 8). (ii) Standard locking plates secured with four standard locking screws inserted in alternating plate holes in the femur shaft (n = 8). (iii) Retrograde intramedullary nail secured proximally with one anterior-posterior screw and distally with two oblique screws (n = 8). Axial hip forces (700 and 2800 N) were applied while measuring axial interfragmentary motion, shear interfragmentary motion, and overall stiffness. Findings: Experimental far cortical locking plate results compared well to published computational findings. Far cortical locking femurs contained the highest axial motion within the potential ideal range of 0.2-1 mm and a sheer-to-axial motion ratio < 1.6 at toe-touch weight-bearing (700 N). At full weight-bearing (2800 N), Standard locking-plated femurs had the only axial motion within 0.2-1 mm but had an excess shear-to-axial motion ratio. Nail-implanted femurs underperformed at both forces. Interpretation: For toe-touch weight-bearing, the far cortical locking construct provided optimal biomechanics to allow moderate motion, which has been suggested to encourage early callus formation. Conversely, at full weight-bearing, the standard locking construct offered the biomechanical advantage on fracture motion.
引用
收藏
页数:8
相关论文
共 46 条
[1]  
[Anonymous], 2017, ASTM F382, V99, P1, DOI [10.1520/F0382-17, DOI 10.1520/F0382-17]
[2]  
Bottlang M, 2011, J ORTHOP TRAUMA, V25, pE60, DOI [10.1097/BOT.0b013e318207885b, 10.1097/01.bot.0000398506.96521.7f]
[3]   Far Cortical Locking Can Improve Healing of Fractures Stabilized with Locking Plates [J].
Bottlang, Michael ;
Lesser, Maren ;
Koerber, Julia ;
Doornink, Josef ;
von Rechenberg, Brigitte ;
Augat, Peter ;
Fitzpatrick, Daniel C. ;
Madey, Steven M. ;
Marsh, J. Lawrence .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (07) :1652-1660
[4]   Far Cortical Locking Can Reduce Stiffness of Locked Plating Constructs While Retaining Construct Strength [J].
Bottlang, Michael ;
Doornink, Josef ;
Fitzpatrick, Daniel C. ;
Madey, Steven M. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (08) :1985-1994
[5]   Weight-bearing recommendations after operative fracture treatment-fact or fiction? Gait results with and feasibility of a dynamic, continuous pedobarography insole [J].
Braun, Benedikt J. ;
Veith, Nils T. ;
Rollmann, Mika ;
Orth, Marcel ;
Fritz, Tobias ;
Herath, Steven C. ;
Holstein, Jorg H. ;
Pohlemann, Tim .
INTERNATIONAL ORTHOPAEDICS, 2017, 41 (08) :1507-1512
[6]  
Brinckmann P., 2016, Orthopedic Biomechanics, V2, DOI [10.1055/b-004-140238, DOI 10.1055/B-004-140238]
[7]   Parametric design of patient-specific fixation plates for distal femur fractures [J].
Chen, Xiaozhong .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2018, 232 (09) :901-911
[8]  
Claes LE, 1998, CLIN ORTHOP RELAT R, pS132
[9]   Far Cortical Locking Enables Flexible Fixation With Periarticular Locking Plates [J].
Doornink, Josef ;
Fitzpatrick, Daniel C. ;
Madey, Steven M. ;
Bottlang, Michael .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 :S29-S34
[10]   Auxiliary locking plate improves fracture stability and healing in intertrochanteric fractures fixated by intramedullary nail [J].
Eberle, Sebastian ;
Gabel, Johannes ;
Hungerer, Sven ;
Hoffmann, Stefanie ;
Paetzold, Robert ;
Augat, Peter ;
Buehren, Volker .