Correlation Between Contrast-Detail Analysis and Clinical Image Quality Assessment of Intrapulmonary Lesions in Dual-Energy Subtraction Chest Radiography Using the TwoShot Method: A Phantom Study

被引:0
作者
Kuramoto, Taku [1 ]
Takarabe, Shinya [2 ]
Kanzaki, Yui [2 ]
Shibayama, Yusuke [2 ]
Yamasaki, Yuzo [3 ]
Kitamura, Yoshiyuki [3 ]
机构
[1] Kobe Tokiwa Univ, Fac Hlth Sci, Dept Radiol Technol, 2-6-2 Otanicho,Nagata Ku, Kobe 6530838, Japan
[2] Kyushu Univ Hosp, Dept Med Technol, Div Radiol, 3-1-1 Maidashi,Higashi Ku, Fukuoka 8128582, Japan
[3] Kyushu Univ, Grad Sch Med Sci, Dept Clin Radiol, 3-1-1,Maidashi,Higashi Ku, Fukuoka 8128582, Japan
基金
日本学术振兴会;
关键词
Dual-energy subtraction imaging; Contrast-detail phantom; Inverse image quality figure; Visual grading analysis; Image quality; PULMONARY NODULES; LUNG-CANCER; COMPUTED RADIOGRAPHY; OPTIMIZATION;
D O I
10.1016/j.acra.2023.10.016
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: Dual -energy subtraction (DES) imaging constitutes a technique that has demonstrated efficacy in enhancing the detectability of pulmonary nodules on chest radiographs. However, a simple and quantitative methodology for evaluating the clinical image quality of DES images is currently lacking. The objective of this study was to investigate the applicability of contrast -detail (C -D) phantom analysis to the visual clinical image quality evaluation of chest DES images. Materials and Methods: We employed a custom-made phantom incorporating the C -D phantom and a multipurpose anthropomorphic adult chest phantom. Two phantom sizes were utilized to simulate standard- and large -bodied adult patients for each phantom. The custom-made phantom images were scored automatically using dedicated software, yielding an inverse image quality figure (IQF inv ) value. The multipurpose anthropomorphic adult chest phantom was employed in a visual grading analysis (VGA) study that was conducted by two experienced radiologists and five radiological technologists. Each nodule placed in the chest phantom image was rated on a 4 -point Likert scale. Results: A statistically significant correlation was observed between the VGA scores of the seven observers and the obtained IQF inv values. Conclusion: The findings of this study suggest that DES image analysis of the C -D phantom possesses the potential to be utilized for the evaluation of clinical DES image quality based on chest lesion detectability.
引用
收藏
页码:2118 / 2127
页数:10
相关论文
共 33 条
  • [11] Peripheral lung cancer: Screening and detection with low-dose spiral CT versus radiography
    Kaneko, M
    Eguchi, K
    Ohmatsu, H
    Kakinuma, R
    Naruke, T
    Suemasu, K
    Moriyama, N
    [J]. RADIOLOGY, 1996, 201 (03) : 798 - 802
  • [12] Effect of high- and low-energy entrance surface dose allocation ratio for two-shot dual-energy subtraction imaging on low-contrast resolution
    Kanzaki, Y.
    Kuramoto, T.
    Takarabe, S.
    Shibayama, Y.
    Yoshikawa, H.
    Kato, T.
    [J]. RADIOGRAPHY, 2023, 29 (01) : 240 - 246
  • [13] Computerized detection of pulmonary nodules by single-exposure dual-energy computed radiography of the chest (part 1)
    Kido, S
    Nakamura, H
    Ito, W
    Shimura, K
    Kato, H
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2002, 44 (03) : 198 - 204
  • [14] Detection of simulated pulmonary nodules by single-exposure dual-energy computed radiography of the chest: effect of a computer-aided diagnosis system (Part 2)
    Kido, S
    Kuriyama, K
    Kuroda, C
    Nakamura, H
    Ito, W
    Shimura, K
    Kato, H
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2002, 44 (03) : 205 - 209
  • [15] CLINICAL-EVALUATION OF PULMONARY NODULES WITH SINGLE-EXPOSURE DUAL-ENERGY SUBTRACTION CHEST RADIOGRAPHY WITH AN ITERATIVE NOISE-REDUCTION ALGORITHM
    KIDO, S
    IKEZOE, J
    NAITO, H
    ARISAWA, J
    TAMURA, S
    KOZUKA, T
    ITO, W
    SHIMURA, K
    KATO, H
    [J]. RADIOLOGY, 1995, 194 (02) : 407 - 412
  • [16] MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA
    LANDIS, JR
    KOCH, GG
    [J]. BIOMETRICS, 1977, 33 (01) : 159 - 174
  • [17] Recent advances in chest radiography
    McAdams, H. Page
    Samei, Ehsan
    Dobbins, James, III
    Tourassi, Georgia D.
    Ravin, Carl E.
    [J]. RADIOLOGY, 2006, 241 (03) : 663 - 683
  • [18] Correlation of the clinical and physical image quality in chest radiography for average adults with a computed radiography imaging system
    Moore, C. S.
    Wood, T. J.
    Beavis, A. W.
    Saunderson, J. R.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2013, 86 (1027)
  • [19] Detection of small pulmonary nodules on chest radiographs: efficacy of dual-energy subtraction technique using flat-panel detector chest radiography
    Oda, S.
    Awai, K.
    Funama, Y.
    Utsunomiya, D.
    Yanaga, Y.
    Kawanaka, K.
    Nakaura, T.
    Hirai, T.
    Murakami, R.
    Nomori, H.
    Yamashita, Y.
    [J]. CLINICAL RADIOLOGY, 2010, 65 (08) : 609 - 615
  • [20] Characterization and validation of the thorax phantom Lungman for dose assessment in chest radiography optimization studies
    Perez, Sunay Rodriguez
    Marshall, Nicholas William
    Struelens, Lara
    Bosmans, Hilde
    [J]. JOURNAL OF MEDICAL IMAGING, 2018, 5 (01)