Correlation Between Contrast-Detail Analysis and Clinical Image Quality Assessment of Intrapulmonary Lesions in Dual-Energy Subtraction Chest Radiography Using the TwoShot Method: A Phantom Study

被引:0
作者
Kuramoto, Taku [1 ]
Takarabe, Shinya [2 ]
Kanzaki, Yui [2 ]
Shibayama, Yusuke [2 ]
Yamasaki, Yuzo [3 ]
Kitamura, Yoshiyuki [3 ]
机构
[1] Kobe Tokiwa Univ, Fac Hlth Sci, Dept Radiol Technol, 2-6-2 Otanicho,Nagata Ku, Kobe 6530838, Japan
[2] Kyushu Univ Hosp, Dept Med Technol, Div Radiol, 3-1-1 Maidashi,Higashi Ku, Fukuoka 8128582, Japan
[3] Kyushu Univ, Grad Sch Med Sci, Dept Clin Radiol, 3-1-1,Maidashi,Higashi Ku, Fukuoka 8128582, Japan
基金
日本学术振兴会;
关键词
Dual-energy subtraction imaging; Contrast-detail phantom; Inverse image quality figure; Visual grading analysis; Image quality; PULMONARY NODULES; LUNG-CANCER; COMPUTED RADIOGRAPHY; OPTIMIZATION;
D O I
10.1016/j.acra.2023.10.016
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: Dual -energy subtraction (DES) imaging constitutes a technique that has demonstrated efficacy in enhancing the detectability of pulmonary nodules on chest radiographs. However, a simple and quantitative methodology for evaluating the clinical image quality of DES images is currently lacking. The objective of this study was to investigate the applicability of contrast -detail (C -D) phantom analysis to the visual clinical image quality evaluation of chest DES images. Materials and Methods: We employed a custom-made phantom incorporating the C -D phantom and a multipurpose anthropomorphic adult chest phantom. Two phantom sizes were utilized to simulate standard- and large -bodied adult patients for each phantom. The custom-made phantom images were scored automatically using dedicated software, yielding an inverse image quality figure (IQF inv ) value. The multipurpose anthropomorphic adult chest phantom was employed in a visual grading analysis (VGA) study that was conducted by two experienced radiologists and five radiological technologists. Each nodule placed in the chest phantom image was rated on a 4 -point Likert scale. Results: A statistically significant correlation was observed between the VGA scores of the seven observers and the obtained IQF inv values. Conclusion: The findings of this study suggest that DES image analysis of the C -D phantom possesses the potential to be utilized for the evaluation of clinical DES image quality based on chest lesion detectability.
引用
收藏
页码:2118 / 2127
页数:10
相关论文
共 33 条
  • [1] An investigation into the validity of utilising the CDRAD 2.0 phantom for optimisation studies in digital radiography
    Al-Murshedi, Sadeq
    Hogg, Peter
    England, Andrew
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2018, 91 (1089)
  • [2] [Anonymous], 1996, Commission European Guidelines on Quality Criteria for Diagnostic Radiographic Images: Eur 16260 EN
  • [3] A comparison of digital radiography systems in terms of effective detective quantum efficiency
    Bertolini, Marco
    Nitrosi, Andrea
    Rivetti, Stefano
    Lanconelli, Nico
    Pattacini, Pierpaolo
    Ginocchi, Vladimiro
    Iori, Mauro
    [J]. MEDICAL PHYSICS, 2012, 39 (05) : 2617 - 2627
  • [4] Validation of a candidate instrument to assess image quality in digital mammography using ROC analysis
    Boita, Joana
    van Engen, Ruben E.
    Mackenzie, Alistair
    Tingberg, Anders
    Bosmans, Hilde
    Bolejko, Anetta
    Zackrisson, Sophia
    Wallis, Matthew G.
    Ikeda, Debra M.
    van Ongeval, Chantal
    Pijnappel, Ruud
    Broeders, Mireille
    Sechopoulos, Ioannis
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2021, 139
  • [5] On site evaluation of three flat panel detectors for digital radiography
    Borasi, G
    Nitrosi, A
    Ferrari, P
    Tassoni, D
    [J]. MEDICAL PHYSICS, 2003, 30 (07) : 1719 - 1731
  • [6] Correlation of Contrast-Detail Analysis and Clinical Image Quality Assessment in Chest Radiography with a Human Cadaver Study
    De Crop, An
    Bacher, Klaus
    Van Hoof, Tom
    Smeets, Peter V.
    Smet, Barbara S.
    Vergauwen, Merel
    Kiendys, Urszula
    Duyck, Philippe
    Verstraete, Koenraad
    D'Herde, Katharina
    Thierens, Hubert
    [J]. RADIOLOGY, 2012, 262 (01) : 298 - 304
  • [7] Optimization of dual-energy subtraction chest radiography by use of a direct-conversion flat-panel detector system
    Fukao M.
    Kawamoto K.
    Matsuzawa H.
    Honda O.
    Iwaki T.
    Doi T.
    [J]. Radiological Physics and Technology, 2015, 8 (1) : 46 - 52
  • [8] Detection of lung cancer using single-exposure dual-energy subtraction chest radiography
    Ide K.
    Mogami H.
    Murakami T.
    Yasuhara Y.
    Miyagawa M.
    Mochizuki T.
    [J]. Radiation Medicine, 2007, 25 (5): : 195 - 201
  • [9] International Atomic Energy Agency, 1996, IAEA Safety Series, V115
  • [10] ONE-SHOT DUAL-ENERGY SUBTRACTION IMAGING
    ISHIGAKI, T
    SAKUMA, S
    HORIKAWA, Y
    IKEDA, M
    YAMAGUCHI, H
    [J]. RADIOLOGY, 1986, 161 (01) : 271 - 273