Signal Quality-Aware Frequency Demodulation-Based ECG-Derived Respiration Rate Estimation Method With Reduced False Alarms

被引:0
作者
Nalwaya, Aditya [1 ]
Manikandan, M. Sabarimalai [2 ]
Pachori, Ram Bilas [1 ]
机构
[1] Indian Inst Technol Indore, Dept Elect Engn, Indore 453552, India
[2] Indian Inst Technol Palakkad, Dept Elect Engn, Palakkad 678623, India
关键词
Electrocardiography; Estimation; Monitoring; Noise; Frequency estimation; Wearable devices; Temperature measurement; Sensor signal processing; electrocardiogram (ECG); healthcare monitoring; respiratory rate (RR); respiratory sinus rhythms; signal quality assessment (SQA); wearable device; ELECTROCARDIOGRAM;
D O I
10.1109/LSENS.2024.3449328
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this letter, we present an automated signal quality-aware frequency demodulation (FD)-based electrocardiogram (ECG)-derived respiration rate (FD-ECG-derived RR) estimation method with reduced false alarms under noisy ECG signals, which are unavoidable in resting and ambulatory health monitoring applications. The proposed FD-ECG-derived RR estimation method includes three major steps of signal quality checking to discard noisy ECG signals, respiratory-induced frequency variation (RIFV) waveform extraction using a frequency demodulation envelope detector by determining peaks of the derivative ECG waveform using a simple R-peak detector, and respiration rate estimation using the Fourier magnitude spectrum of the extracted RIFV waveform. On the standard Capnobase and BIDMC databases, the proposed FD-ECG-derived RR estimation method provides promising results with mean absolute error values of 5.01 and 5.37 breaths/min, respectively. The signal quality-aware RR estimation method used can reduce false alarm rate of 84.85$% by discarding noisy ECG signals with quality assessment accuracy of 85.25%. The proposed simplistic method having lightweight signal processing approaches makes it suitable for real-time health monitoring applications.
引用
收藏
页数:4
相关论文
共 20 条
  • [11] Electrocardiogram Derived Respiratory Rate Using a Wearable Armband
    Lazaro, Jesus
    Reljin, Natasa
    Bailon, Raquel
    Gil, Eduardo
    Noh, Yeonsik
    Laguna, Pablo
    Chon, Ki H.
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2021, 68 (03) : 1056 - 1065
  • [12] A novel method for detecting R-peaks in electrocardiogram (ECG) signal
    Manikandan, M. Sabarimalai
    Soman, K. P.
    [J]. BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2012, 7 (02) : 118 - 128
  • [13] Mondal Achinta, 2023, 2023 9th International Conference on Signal Processing and Communication (ICSC), P532, DOI 10.1109/ICSC60394.2023.10441007
  • [14] Fast and Resource Efficient Atrial Fibrillation Detection Framework for Long Term Health Monitoring Devices
    Phukan, Nabasmita
    Manikandan, M. Sabarimalai
    Pachori, Ram Bilas
    [J]. IEEE SENSORS LETTERS, 2024, 8 (04) : 1 - 4
  • [15] Toward a Robust Estimation of Respiratory Rate From Pulse Oximeters
    Pimentel, Marco A. F.
    Johnson, Alistair E. W.
    Charlton, Peter H.
    Birrenkott, Drew
    Watkinson, Peter J.
    Tarassenko, Lionel
    Clifton, David A.
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2017, 64 (08) : 1914 - 1923
  • [16] Unified Quality-Aware Compression and Pulse-Respiration Rates Estimation Framework for Reducing Energy Consumption and False Alarms of Wearable PPG Monitoring Devices
    Reddy, Gangireddy Narendra Kumar
    Manikandan, M. Sabarimalai
    Murty, N. V. L. Narasimha
    Cenkeramaddi, Linga Reddy
    [J]. IEEE ACCESS, 2023, 11 : 41708 - 41740
  • [17] Real-Time Signal Quality-Aware ECG Telemetry System for IoT-Based Health Care Monitoring
    Satija, Udit
    Ramkumar, Barathram
    Manikandan, M. Sabarimalai
    [J]. IEEE INTERNET OF THINGS JOURNAL, 2017, 4 (03): : 815 - 823
  • [18] Wearable devices for continuous monitoring of biosignals: Challenges and opportunities
    Stuart, Tucker
    Hanna, Jessica
    Gutruf, Philipp
    [J]. APL BIOENGINEERING, 2022, 6 (02):
  • [19] Generalized frequency modulation
    Torres, WP
    Oppenheim, AV
    Rosales, RR
    [J]. IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS I-FUNDAMENTAL THEORY AND APPLICATIONS, 2001, 48 (12): : 1405 - 1412
  • [20] Predicting Respiratory Rate from Electrocardiogram and Photoplethysmogram Using a Transformer-Based Model
    Zhao, Qi
    Liu, Fang
    Song, Yide
    Fan, Xiaoya
    Wang, Yu
    Yao, Yudong
    Mao, Qian
    Zhao, Zheng
    [J]. BIOENGINEERING-BASEL, 2023, 10 (09):