Comparison of baseline correction algorithms for in vivo 1H-MRS

被引:0
作者
Pasmino, Diego [1 ]
Slotboom, Johannes [2 ]
Schweisthal, Brigitte [2 ,3 ]
Guevara, Pamela [1 ]
Valenzuela, Waldo [2 ]
Pino, Esteban J. [1 ]
机构
[1] Univ Concepcion, Elect Engn Dept, Edmundo Larenas 219, Concepcion, Bio Bio, Chile
[2] Univ Hosp Inselspital, Support Ctr Adv Neuroimaging SCAN, Neuroradiol, Bern, Switzerland
[3] Politehn Univ Timisoara, Timisoara, Romania
基金
瑞士国家科学基金会;
关键词
baseline correction; CSI; in vivo H-1-MRI; MRS; MRSI; AUTOMATED SPECTRAL-ANALYSIS; PROTON MR SPECTROSCOPY; RESONANCE-SPECTROSCOPY; QUANTITATION; TIME; QUALITY; BRAIN; 1D;
D O I
10.1002/nbm.5203
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Proton MRS is used clinically to collect localized, quantitative metabolic data from living tissues. However, the presence of baselines in the spectra complicates accurate MRS data quantification. The occurrence of baselines is not specific to short-echo-time MRS data. In short-echo-time MRS, the baseline consists typically of a dominating macromolecular (MM) part, and can, depending on B-0 shimming, poor voxel placement, and/or localization sequences, also contain broad water and lipid resonance components, indicated by broad components (BCs). In long-echo-time MRS, the MM part is usually much smaller, but BCs may still be present. The sum of MM and BCs is denoted by the baseline. Many algorithms have been proposed over the years to tackle these artefacts. A first approach is to identify the baseline itself in a preprocessing step, and a second approach is to model the baseline in the quantification of the MRS data themselves. This paper gives an overview of baseline handling algorithms and also proposes a new algorithm for baseline correction. A subset of suitable baseline removal algorithms were tested on in vivo MRSI data (semi-LASER at T-E = 40 ms) and compared with the new algorithm. The baselines in all datasets were removed using the different methods and subsequently fitted using spectrIm-QMRS with a TDFDFit fitting model that contained only a metabolite basis set and lacked a baseline model. The same spectra were also fitted using a spectrIm-QMRS model that explicitly models the metabolites and the baseline of the spectrum. The quantification results of the latter quantification were regarded as ground truth. The fit quality number (FQN) was used to assess baseline removal effectiveness, and correlations between metabolite peak areas and ground truth models were also examined. The results show a competitive performance of our new proposed algorithm, underscoring its automatic approach and efficiency. Nevertheless, none of the tested baseline correction methods achieved FQNs as good as the ground truth model. All separately applied baseline correction methods introduce a bias in the observed metabolite peak areas. We conclude that all baseline correction methods tested, when applied as a separate preprocessing step, yield poorer FQNs and biased quantification results. While they may enhance visual display, they are not advisable for use before spectral fitting.
引用
收藏
页数:16
相关论文
共 52 条
  • [1] A new automatic baseline correction method based on iterative method
    Bao, Qingjia
    Feng, Jiwen
    Chen, Fang
    Mao, Wenping
    Liu, Zao
    Liu, Kewen
    Liu, Chaoyang
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2012, 218 : 35 - 43
  • [2] Barker P., 2009, CLIN MR SPECTROSCOPY, P1, DOI DOI 10.1017/CBO9780511770647.002
  • [3] Barker PB., 2009, CLIN MR SPECTROSCOPY, P34
  • [4] Bazgir O, 2018, IEEE SW SYMP IMAG, P97, DOI 10.1109/SSIAI.2018.8470319
  • [5] Belki D., 2010, SIGNAL PROCESSING MA, DOI [10.1201/9781439806456, DOI 10.1201/9781439806456]
  • [6] Brain Proton Magnetic Resonance Spectroscopy Introduction and Overview
    Bertholdo, Debora
    Watcharakorn, Arvemas
    Castillo, Mauricio
    [J]. NEUROIMAGING CLINICS OF NORTH AMERICA, 2013, 23 (03) : 359 - +
  • [7] 1H-MRS processing parameters affect metabolite quantification: The urgent need for uniform and transparent standardization
    Bhogal, Alex A.
    Schur, Remmelt R.
    Houtepen, Lotte C.
    van de Bank, Bart
    Boer, Vincent O.
    Marsman, Anouk
    Barker, Peter B.
    Scheenen, Tom W. J.
    Wijnen, Jannie P.
    Vinkers, Christiaan H.
    Klomp, Dennis W. J.
    [J]. NMR IN BIOMEDICINE, 2017, 30 (11)
  • [8] Influence of macromolecule baseline on 1H MR spectroscopic imaging reproducibility
    Birch, Rebecca
    Peet, Andrew C.
    Dehghani, Hamid
    Wilson, Martin
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2017, 77 (01) : 34 - 43
  • [9] Imaging in breast cancer: Magnetic resonance spectroscopy
    Bolan, PJ
    Nelson, MT
    Yee, D
    Garwood, M
    [J]. BREAST CANCER RESEARCH, 2005, 7 (04): : 149 - 152
  • [10] ProFit-1D-A 1D fitting software and open-source validation data sets
    Borbath, Tamas
    Murali-Manohar, Saipavitra
    Dorst, Johanna
    Wright, Andrew Martin
    Henning, Anke
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2021, 86 (06) : 2910 - 2929