Intraoperative Laser Speckle Contrast Imaging to Assess Vessel Flow in Neurosurgery: A Pilot Study

被引:3
|
作者
Goldberg, Johannes [1 ,2 ]
Miller, David R. [3 ]
Dimanche, Alexis [2 ,4 ]
Kissling, Cedric [1 ,2 ]
Mueller, Timothy [1 ,2 ]
Mueller, Mandy D. [1 ,2 ]
Jesse, Christopher Marvin [1 ,2 ]
Murek, Michael [1 ,2 ]
Bervini, David [1 ,2 ]
Dunn, Andrew K. [4 ]
Raabe, Andreas [1 ,2 ]
机构
[1] Univ Bern, Bern Univ Hosp, Inselspital, Dept Neurosurg, Freiburgstr 16, CH-3010 Bern, Switzerland
[2] Univ Bern, Bern Univ Hosp, Inselspital, Stroke Res Ctr Bern, Freiburgstr 16, CH-3010 Bern, Switzerland
[3] Dynam Light Inc, Austin, TX USA
[4] Univ Texas Austin, Dept Biomed Engn, Austin, TX USA
关键词
Cerebral blood flow; Cerebrovascular neurosurgery; Laser speckle contrast imaging; Indocyanine green angiography; AGREEMENT;
D O I
10.1227/neu.0000000000002776
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND OBJECTIVES: Laser speckle contrast imaging (LSCI) has emerged as a promising tool for assessment of vessel flow during neurosurgery. We aimed to investigate the feasibility of visualizing vessel flow in the macrocirculation with a new fully microscope-integrated LSCI system and assess the validity and objectivity of findings compared with fluorescence angiography (FA). METHODS: This is a single-center prospective observational study enrolling adult patients requiring microsurgical treatment for brain vascular pathologies or brain tumors. Three independent raters, blinded toward findings of FA, reviewed regions of interest (ROIs) placed in exposed vessels and target structures. The primary end point was the validity of LSCI for assessment of vessel flow as measured by the agreement with FA. The secondary end point was objectivity, measured as the inter-rater agreement of LSCI findings. RESULTS: During 18 surgical procedures, 23 observations using FA and LSCI were captured simultaneously. Using LSCI, vessel flow was assessable in 62 (86.1%) and not assessable in 10 (13.9%) ROIs. The agreement between LSCI and FA was 86.1%, with an agreement coefficient of 0.85 (95% CI: 0.75-0.94). Disagreement between LSCI and FA was observed in the 10 ROIs that were not assessable. The agreement between ROIs that were assessable using LSCI and FA was 100%. The inter-rater agreement of LSCI findings was 87.9%, with an agreement coefficient of 0.86 (95% CI: 0.79-0.94). CONCLUSION: Fully microscope-integrated LSCI is feasible and has a high potential for clinical utility. Because of its characteristics, LSCI can be viewed as a full-field visual micro-Doppler that can be used as a complementary method to FA for assessing vessel flow during neurosurgery. Despite technical limitations related to the early development phase of the fully microscope-integrated system, we demonstrated reasonable validity and objectivity of findings compared with FA. Further research and refinement of the system may enhance its value in neurosurgical applications.
引用
收藏
页码:983 / 992
页数:10
相关论文
共 50 条
  • [1] Laser speckle contrast imaging of cerebral blood flow in humans during neurosurgery: a pilot clinical study
    Parthasarathy, Ashwin B.
    Weber, Erica L.
    Richards, Lisa M.
    Fox, Douglas J.
    Dunn, Andrew K.
    JOURNAL OF BIOMEDICAL OPTICS, 2010, 15 (06)
  • [2] Laser Speckle Contrast Imaging in Neurosurgery: A Systematic Review
    Konovalov, Anton
    Gadzhiagaev, Vadim
    Grebenev, Fyodor
    Stavtsev, Dmitry
    Piavchenko, Gennadii
    Gerasimenko, Alexander
    Telyshev, Dmitry
    Meglinski, Igor
    Eliava, Shalva
    WORLD NEUROSURGERY, 2023, 171 : 35 - 40
  • [3] Intraoperative laser speckle contrast imaging for monitoring cerebral blood flow: results from a 10-patient pilot study
    Richards, Lisa M.
    Weber, Erica L.
    Parthasarathy, Ashwin B.
    Kappeler, Kaelyn L.
    Fox, Douglas J.
    Dunn, Andrew K.
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS VIII, PTS 1 AND 2, 2012, 8207
  • [4] Laser Speckle Contrast Imaging for Intraoperative Monitoring of Cerebral Blood Flow
    Stavtsev D.D.
    Konovalov A.N.
    Blinova E.V.
    Piavchenko G.A.
    Golodnev G.E.
    Zalogin S.D.
    Gorina A.V.
    Lapin K.N.
    Vagner S.A.
    Grebenev F.V.
    Meglinski I.V.
    Gerasimenko A.Y.
    Telyshev D.V.
    Kuznetsov S.L.
    Bulletin of the Russian Academy of Sciences: Physics, 2022, 86 (Suppl 1): : S229 - S233
  • [5] Intraoperative laser speckle contrast imaging with retrospective motion correction for quantitative assessment of cerebral blood flow
    Richards, Lisa M.
    Towle, Erica L.
    Fox, Douglas J., Jr.
    Dunn, Andrew K.
    NEUROPHOTONICS, 2014, 1 (01)
  • [6] Comparison between laser speckle contrast imaging and laser Doppler imaging to assess skin blood flow in humans
    Millet, C.
    Roustit, M.
    Blaise, S.
    Cracowski, J. L.
    MICROVASCULAR RESEARCH, 2011, 82 (02) : 147 - 151
  • [7] A study on an improved laser speckle contrast blood flow imaging methodology
    Han, Guang
    Lu, Jie
    Dong, Xinming
    Li, De
    Yuan, Jixin
    Yang, Qinglong
    Wang, Huiquan
    Chen, Ruijuan
    Wu, Yifan
    Wang, Jinhai
    Min, Xiaolin
    INFRARED PHYSICS & TECHNOLOGY, 2024, 142
  • [8] Spatiotemporal laser speckle contrast analysis for blood flow imaging with maximized speckle contrast
    Qiu, Jianjun
    Li, Pengcheng
    Luo, Weihua
    Wang, Jia
    Zhang, Hongyan
    Luo, Qingming
    JOURNAL OF BIOMEDICAL OPTICS, 2010, 15 (01)
  • [9] Continuous blood flow visualization with laser speckle contrast imaging during neurovascular surgery
    Miller, David R.
    Ashour, Ramsey
    Sullender, Colin T.
    Dunn, Andrew K.
    NEUROPHOTONICS, 2022, 9 (02)
  • [10] Laser Speckle Contrast Imaging of flow in a microfluidic device
    Parthasarathy, Ashwin B.
    Shin, Weon Gyu
    Zhang, X. J.
    Dunn, A. K.
    BIOMEDICAL APPLICATIONS OF LIGHT SCATTERING, 2007, 6446