Body Core Temperature Assessment in Emergency Care Departments

被引:1
作者
Daanen, Hein A. M. [1 ]
Hoitinga, Gercora [2 ]
Kruijt, David J. [2 ]
Koning, Kevin S. [1 ]
Verheijen, Pooh P. [1 ]
de Baas, Sanne I. M. [1 ]
Bergsma, Anouk R. [1 ]
Snethlage, Cathelijne E. [1 ]
Al -Bander, Iman [1 ]
Teunissen, Lennart P. J. [3 ]
机构
[1] Vrije Univ Amsterdam, Fac Behav & Movement Sci, Dept Human Movement Sci, Amsterdam Movement Sci, Boechorststr 7, NL-1081 BT Amsterdam, Netherlands
[2] Univ Amsterdam, Dept Emergency Med, Amsterdam UMC, Amsterdam, Netherlands
[3] Minist Def, Dept Training Med & Training Physiol, Royal Netherlands Army, Utrecht, Netherlands
关键词
thermometry; tympanic temperature; fever; arteria temporalis; thermometer; exercise; calibration; RELIABILITY; ACCURACY; VALIDITY; SKIN; THERMOMETERS;
D O I
10.1016/j.jemermed.2023.10.027
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: There is concern that the values provided by devices using infrared thermometry in emergency departments (EDs) do not reflect body core temperature accurately. Objectives: Evaluation of three thermometers commonly used in the ED. Methods: Two infrared ear thermometers and an infrared forehead thermometer were evaluated using 1) the Voltcraft IRS -350 calibration device, 2) comparing temperature values to a rectal end -exercise temperature (T-RECT) of 38.1 degrees C in 12 participants, and 3) comparing temperature values to rectal temperature in 133 ED patients. Results: Calibration across the human core temperature range revealed that the ear thermometers underestimated radiant temperature by 0.77 +/- 0.39 degrees C and 1.84 +/- 0.26 degrees C, respectively, whereas the forehead thermometer overestimated radiant temperature by 0.90 +/- 0.51 degrees C. After cycling exercise, all thermometers underestimated T-RECT (0.54 +/- 0.27 degrees C and 1.03 +/- 0.48 degrees C for the ear thermometers and 1.14 +/- 0.38 degrees C for the forehead thermometer). In the ED, the ear thermometers underestimated T-RECT by 0.31 +/- 0.37 degrees C and 0.46 +/- 0.50 degrees C, whereas the forehead thermometer exhibited a nonsignificant overestimation of 0.04 +/- 0.46 degrees C. If the threshold for fever in all systems had been set to 37.5 degrees C instead of 38.0 degrees C, the sensitivity and specificity of the systems for real fever (T-RECT >= 38 degrees C) are, respectively, 71% and 96% (ear thermometer 1), 57% and 97% (ear thermometer 2), and 86% and 90% (forehead thermometer). Conclusion: We conclude that the investigated thermometers are not reliable as devices to measure radiant temperature, cannot be used to assess body core temperature during exercise, but may be used as a screening device, with 37.5 degrees C as a threshold for fever in emergency care settings. (c) 2023 The Author(s). Published by Elsevier Inc. This is an open access article under the CC BY license ( http://creativecommons.org/licenses/by/4.0/ )
引用
收藏
页码:e277 / e283
页数:7
相关论文
共 29 条
  • [1] ASTM, 2021, E284713 ASTM
  • [2] ASTM, 2016, E196598 ASTM
  • [3] Validity, Reliability, and Inertia of Four Different Temperature Capsule Systems
    Bongers, Coen C. W. G.
    Daanen, Hein A. M.
    Bogerd, Cornelis P.
    Hopman, Maria T. E.
    Eijsvogels, Thijs M. H.
    [J]. MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2018, 50 (01) : 169 - 175
  • [4] The ingestible telemetric body core temperature sensor: a review of validity and exercise applications
    Byrne, Christopher
    Lim, Chin Leong
    [J]. BRITISH JOURNAL OF SPORTS MEDICINE, 2007, 41 (03) : 126 - 133
  • [5] Cutuli SL, 2021, CRIT CARE RESUSC, V23, P6, DOI 10.51893/2021.1.SR1
  • [6] Daanen H. A. M., 2006, Journal of Medical Engineering & Technology, V30, P224, DOI 10.1080/03091900600711613
  • [7] Heat flux systems for body core temperature assessment during exercise
    Daanen, Hein A. M.
    Kohlen, Veerle
    Teunissen, Lennart P. J.
    [J]. JOURNAL OF THERMAL BIOLOGY, 2023, 112
  • [8] Temperature measurement: comparison of non-invasive methods used in adult critical care
    Farnell, S
    Maxwell, L
    Tan, S
    Rhodes, A
    Philips, B
    [J]. JOURNAL OF CLINICAL NURSING, 2005, 14 (05) : 632 - 639
  • [9] The validity and reliability of intestinal temperature during intermittent running
    Gant, Nicholas
    Atkinson, Greg
    Williams, Clyde
    [J]. MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2006, 38 (11) : 1926 - 1931
  • [10] International Organization for Standardization (ISO), 2017, 13154 ISO TR