Safety and efficacy of custom foldable silicone artificial iris implantation: prospective compassionate-use case series

被引:19
作者
Bonnet, Clemence [1 ,2 ]
Miller, Kevin M. [1 ]
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, Stein Eye Inst, Dept Ophthalmol, Los Angeles, CA USA
[2] Paris Descartes Univ, Cochin Hosp, Opthalmole Paris, Paris, France
关键词
D O I
10.1097/j.jcrs.0000000000000172
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the preliminary safety and efficacy of custom silicone artificial iris implantation. Setting: Stein Eye Institute. Design: Prospective consecutive case series. Methods: Medical records of patients implanted with the artificial iris were reviewed and followed for 1 year. Safety measures included corrected distance visual acuity (CDVA), intraocular pressure (IOP), endothelial cell count (ECC), surgical complications, secondary interventions, and adverse events (AEs). Efficacy measures included CDVA with glare, subjective daytime and nighttime glare graded from 0 to 10 (very significant), and subjective cosmetic appearance graded from 0 to 10 (very satisfied). Results: Twenty eyes (19 patients) were implanted. Safety data were mixed. CDVA improved in 13 eyes and worsened in 6. ECC decreased from 1918 +/- 870 to 1405 +/- 705 cells/mm(2) (P = .02). Eight eyes experienced postoperative complications. There were 4 IOP elevations, 2 corneal decompensations, 1 case of cystoid macular edema, and 1 device dislocation. Four eyes underwent secondary surgical interventions including 2 AEs (1 glaucoma surgery and 1 device dislocation). Efficacy outcomes were excellent. CDVA with glare improved from 1.5 +/- 0.6 to 0.7 +/- 0.8 logarithm of the minimum angle of resolution (P <.01). Mean subjective daytime glare decreased from 8.9 +/- 1.8 to 2.7 +/- 2.6 (P <.01) and nighttime glare decreased from 7.9 +/- 1.8 to 2.5 +/- 2.7 (P <.01). Mean cosmesis improved from 2.2 +/- 1.6 to 8.8 +/- 2.1 (P <.01). Conclusions: Custom artificial iris implantation was moderately risky but very effective at reducing light and glare sensitivity and in improving ocular cosmesis.
引用
收藏
页码:893 / 901
页数:9
相关论文
共 19 条
  • [1] Small-incision insertion of artificial iris prostheses
    Ayliffe, Will
    Groth, Sylvia L.
    Sponsel, William E.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (02) : 362 - 367
  • [2] Incidence and risk factors for traumatic intraocular pressure elevation and traumatic glaucoma after open-globe injury
    Bojikian, K. D.
    Stein, A. L.
    Slabaugh, M. A.
    Chen, P. P.
    [J]. EYE, 2015, 29 (12) : 1579 - 1584
  • [3] Simultaneous correction of post-traumatic aphakia and aniridia with the use of artificial iris and IOL implantation
    Forlini, Cesare
    Forlini, Matteo
    Rejdak, Robert
    Prokopiuk, Agata
    Levkina, Oxana
    Bratu, Adriana
    Rossini, Paolo
    Cagampang, Perfecto R., III
    Cavallini, Gian Maria
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2013, 251 (03) : 667 - 675
  • [4] Jiang S, 2019, Retin Cases Brief Rep
  • [5] Implantation of an Artificial Iris in 11 Patients
    Kniestedt, C.
    Eberhard, R.
    Fleischhauer, J.
    [J]. KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2016, 233 (04) : 365 - 368
  • [6] Artificial iris devices: Benefits, limitations, and management of complications
    Koch, Konrad R.
    Heindl, Ludwig M.
    Cursiefen, Claus
    Koch, Hans-Reinhard
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (03) : 376 - 382
  • [7] Mayer CS, 2015, OPHTHALMOLOGE, V112, P865, DOI 10.1007/s00347-015-0123-6
  • [8] Artificial iris implantation in various iris defects and lens conditions
    Mayer, Christian
    Tandogan, Tamer
    Hoffmann, Andrea E.
    Khoramnia, Ramin
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (06) : 724 - 731
  • [9] Challenges and Complication Management in Novel Artificial Iris Implantation
    Mayer, Christian S.
    Laubichler, Andrea E.
    Khoramnia, Ramin
    Tandogan, Tamer
    Prahs, Philipp
    Zapp, Daniel
    Reznicek, Lukas
    [J]. JOURNAL OF OPHTHALMOLOGY, 2018, 2018
  • [10] Residual Iris Retraction Syndrome After Artificial Iris Implantation
    Mayer, Christian S.
    Laubichler, Andrea E.
    Masyk, Michael
    Prahs, Philipp
    Zapp, Daniel
    Khoramnia, Ramin
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2019, 199 : 159 - 166