Performance evaluation of deep learning image reconstruction algorithm for dual-energy spectral CT imaging: A phantom study

被引:3
作者
Li, Haoyan [1 ]
Li, Zhentao [2 ]
Gao, Shuaiyi [1 ]
Hu, Jiaqi [1 ]
Yang, Zhihao [1 ]
Peng, Yun [1 ]
Sun, Jihang [1 ]
机构
[1] Capital Med Univ, Beijing Childrens Hosp, Natl Ctr Childrens Hlth, Dept Radiol, 56 Nanlishi Rd, Beijing 100045, Peoples R China
[2] Peking Univ, Dept Radiol, Peoples Hosp, Beijing, Peoples R China
关键词
Multidetector computed tomography; image enhancement; image reconstruction; deep learning; TOMOGRAPHY PHYSICAL PRINCIPLES; FILTERED BACK-PROJECTION; NOISE POWER SPECTRUM; ITERATIVE RECONSTRUCTION; CHEST CT; DOSE REDUCTION; LIVER-LESIONS; CONTRAST; QUALITY; DETECTABILITY;
D O I
10.3233/XST-230333
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
OBJECTIVES: To evaluate the performance of deep learning image reconstruction (DLIR) algorithm in dual-energy spectral CT (DEsCT) as a function of radiation dose and image energy level, in comparison with filtered-back-projection (FBP) and adaptive statistical iterative reconstruction-V (ASIR-V) algorithms. METHODS: An ACR464 phantom was scanned with DEsCT at four dose levels (3.5 mGy, 5 mGy, 7.5 mGy, and 10 mGy). Virtual monochromatic images were reconstructed at five energy levels (40 keV, 50 keV, 68 keV, 74 keV, and 140 keV) using FBP, 50% and 100% ASIR-V, DLIR at low (DLIR-L), medium (DLIR-M), and high (DLIR-H) settings. The noise power spectrum (NPS), task-based transfer function (TTF) and detectability index (d') were computed and compared among reconstructions. RESULTS: NPS area and noise increased as keV decreased, with DLIR having slower increase than FBP and ASIR-V, and DLIR-H having the lowest values. DLIR had the best 40 keV/140 keV noise ratio at various energy levels, DLIR showed higher TTF (50%) than ASIR-V for all materials, especially for the soft tissue-like polystyrene insert, and DLIR-M and DLIR-H provided higher d' than DLIR-L, ASIR-V and FBP in all dose and energy levels. As keV increases, d' increased for acrylic insert, and d' of the 50 keV DLIR-M and DLIR-H images at 3.5 mGy (7.39 and 8.79, respectively) were higher than that (7.20) of the 50 keV ASIR-V50% images at 10 mGy. CONCLUSIONS: DLIR provides better noise containment for low keV images in DEsCT and higher TTF(50%) for the polystyrene insert over ASIR-V. DLIR-H has the lowest image noise and highest detectability in all dose and energy levels. DEsCT 50 keV images with DLIR-M and DLIR-H show potential for 65% dose reduction over ASIR-V50% with higher d'.
引用
收藏
页码:513 / 528
页数:16
相关论文
共 36 条
[1]   Deep learning reconstruction versus iterative reconstruction for cardiac CT angiography in a stroke imaging protocol: reduced radiation dose and improved image quality [J].
Bernard, Angelique ;
Comby, Pierre-Olivier ;
Lemogne, Brivael ;
Haioun, Karim ;
Ricolfi, Frederic ;
Chevallier, Olivier ;
Loffroy, Romaric .
QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2021, 11 (01) :392-401
[2]   Application of the noise power spectrum in modern diagnostic MDCT: Part II. Noise power spectra and signal to noise [J].
Boedeker, K. L. ;
McNitt-Gray, M. F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) :4047-4061
[3]   Pros and Cons of Dual-Energy CT Systems: "One Does Not Fit All" [J].
Borges, Ana P. P. ;
Antunes, Celia ;
Curvo-Semedo, Luis .
TOMOGRAPHY, 2023, 9 (01) :195-216
[4]   Improving Image Quality and Reducing Radiation Dose for Pediatric CT by Using Deep Learning Reconstruction [J].
Brady, Samuel L. ;
Trout, Andrew T. ;
Somasundaram, Elanchezhian ;
Anton, Christopher G. ;
Li, Yinan ;
Dillman, Jonathan R. .
RADIOLOGY, 2021, 298 (01) :180-188
[5]   A Deep Learning Image Reconstruction Algorithm for Improving Image Quality and Hepatic Lesion Detectability in Abdominal Dual-Energy Computed Tomography: Preliminary Results [J].
Chu, Bingqian ;
Gan, Lu ;
Shen, Yi ;
Song, Jian ;
Liu, Ling ;
Li, Jianying ;
Liu, Bin .
JOURNAL OF DIGITAL IMAGING, 2023, 36 (06) :2347-2355
[6]   Practical considerations for noise power spectra estimation for clinical CT scanners [J].
Dolly, Steven ;
Chen, Hsin-Chen ;
Anastasio, Mark ;
Mutic, Sasa ;
Li, Hua .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (03) :392-407
[7]   Dual-Energy Computed Tomography: Physical Principles, Approaches to Scanning, Usage, and Implementation: Part 2 [J].
Forghani, Reza ;
De Man, Bruno ;
Gupta, Rajiv .
NEUROIMAGING CLINICS OF NORTH AMERICA, 2017, 27 (03) :385-+
[8]  
Forghani R, 2017, NEUROIMAG CLIN N AM, V27, P371, DOI 10.1016/j.nic.2017.03.002
[9]   Preserving image texture while reducing radiation dose with a deep learning image reconstruction algorithm in chest CT: A phantom study [J].
Franck, Caro ;
Zhang, Guozhi ;
Deak, Paul ;
Zanca, Federica .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 81 :86-93
[10]   A comprehensive review of dual-energy and multi-spectral computed tomography [J].
Garnett, Richard .
CLINICAL IMAGING, 2020, 67 :160-169