Sonography in the diagnosis of peri-implant bone defects: An in vitro study on native human mandibles

被引:0
作者
Bykhovsky, Igor [1 ]
Hildner, Alexander [2 ]
Kripfgans, Oliver D. [3 ]
Mengel, Reiner [1 ]
机构
[1] Philipps Univ Marburg Lahn, Sch Dent Med, Dept Prosthet Dent, Georg Voigt Str 3, D-35039 Marburg, Germany
[2] Alcedis GmbH, Giessen, Germany
[3] Univ Michigan, Dept Radiol, Ann Arbor, MI USA
关键词
dental implant; diagnostics; human cadaver study; peri-implant bone defects; sonography; DIGITAL VOLUME TOMOGRAPHY; ACCURACY; HEALTH; PIG;
D O I
10.1111/clr.14302
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim The aim of this study on native human cadavers was to compare clinical, sonographic, and radiological measurements of fenestrations, dehiscences, and 3-wall bone defects on implants. Materials and Methods The examination was carried out on five human mandibles. After the insertion of 27 implants, dehiscences (n = 14), fenestrations (n = 7) and 3-wall bone defects (n = 6) were prepared in a standardized manner. The direct measurement of the bone defects was carried out with a periodontal probe and the radiological examination was carried out using digital volume tomography (DVT). The ultrasound examination (US) was performed using a clinical 24-MHz US imaging probe. Means and standard deviations of the direct, US, and DVT measurements were calculated. Measurements were statistically compared using the Pearson correlation coefficient and Bland-Altman analysis. Results Bone defects were on average 3.22 +/- 1.58 mm per direct measurement, 2.90 +/- 1.47 mm using US, and 2.99 +/- 1.52 mm per DVT assessment. Pairwise correlations of these measurements were R = .94 (p < .0001) between direct and US, R = .95 (p < .0001) between DVT and US, and R = .96 (p < .0001) between direct and DVT. The mean differences of the measurements (and 95% CI) between direct and US was 0.41 (-0.47 to 1.29), US and DVT 0.33 (-0.30 to 0.97), and direct and DVT 0.28 (-0.50 to 1.07). Conclusion All peri-implant bone defects could be identified and sonographically measured. US measurements showed a strong correlation with direct and DVT measurements. The sonographic measurement accuracy was highest for dehiscences, followed by fenestrations and 3-wall bone defects.
引用
收藏
页码:1128 / 1137
页数:10
相关论文
共 29 条
  • [1] Ultrasonographic Tissue Perfusion in Peri-implant Health and Disease
    Barootchi, S.
    Tavelli, L.
    Majzoub, J.
    Chan, H. L.
    Wang, H. L.
    Kripfgans, O. D.
    [J]. JOURNAL OF DENTAL RESEARCH, 2022, 101 (03) : 278 - 285
  • [2] Overview of Ultrasound in Dentistry for Advancing Research Methodology and Patient Care Quality with Emphasis on Periodontal/Peri-implant Applications
    Betancourt, Amanda Rodriguez
    Samal, Ankita
    Chan, Hsun-Liang
    Kripfgans, Oliver D.
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2023, 33 (03): : 336 - 386
  • [3] Updates on ultrasound research in implant dentistry: a systematic review of potential clinical indications
    Bhaskar, Vaishnavi
    Chan, Hsun-Liang
    MacEachern, Mark
    Kripfgans, Oliver D.
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2018, 47 (06)
  • [4] Ultrasonography for diagnosis of peri-implant diseases and conditions: a detailed scanning protocol and case demonstration
    Chan, Hsun-Liang
    Kripfgans, Oliver D.
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2020, 49 (07)
  • [5] Ultrasonography for noninvasive and real-time evaluation of peri-implant tissue dimensions
    Chan, Hsun-Liang
    Sinjab, Khaled
    Li, Junying
    Chen, Zhaozhao
    Wang, Hom-Lay
    Kripfgans, Oliver D.
    [J]. JOURNAL OF CLINICAL PERIODONTOLOGY, 2018, 45 (08) : 986 - 995
  • [6] Non-ionizing real-time ultrasonography in implant and oral surgery: A feasibility study
    Chan, Hsun-Liang
    Wang, Hom-Lay
    Fowlkes, Jeffery Brian
    Giannobile, William V.
    Kripfgans, Oliver D.
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2017, 28 (03) : 341 - 347
  • [7] Non-invasive evaluation of facial crestal bone with ultrasonography
    Chan, Hsun-Liang
    Sinjab, Khaled
    Chung, Ming-Pang
    Chiang, Yi-Chen
    Wang, Hom-Lay
    Giannobile, William V.
    Kripfgans, Oliver D.
    [J]. PLOS ONE, 2017, 12 (02):
  • [8] Peri-implant health and disease. A systematic review of current epidemiology
    Derks, Jan
    Tomasi, Cristiano
    [J]. JOURNAL OF CLINICAL PERIODONTOLOGY, 2015, 42 : S158 - S171
  • [9] The accuracy of probing, ultrasound and cone-beam CT scans for determining the buccal bone plate dimensions around oral implants - A systematic review
    Dhondt, Rutger
    Quirynen, Marc
    Tarce, Mihai
    Teughels, Wim
    Temmerman, Andy
    Jacobs, Reinhilde
    [J]. JOURNAL OF PERIODONTAL RESEARCH, 2022, 57 (04) : 754 - 767
  • [10] Ultrasonic Assessment of Mucosal Thickness around Implants: Validity, Reproducibility, and Stability of Connective Tissue Grafts at the Buccal Aspect
    Eghbali, Aryan
    De Bruyn, Hugo
    Cosyn, Jan
    Kerckaert, Ingrid
    Tom Van Hoof