X-ray computed tomography for effective atomic number and density evaluation: A review

被引:3
作者
Lima, Mateus Gruener [1 ]
Jussiani, Eduardo Inocente [1 ]
机构
[1] Univ Estadual Londrina, Appl Nucl Phys Res Grp, Rodovia Celso Garcia Cid PR 445 Km 380, BR-86057970 Londrina, Parana, Brazil
关键词
Spectral CT; Dual-energy CT; Single-energy CT; Photon-counting detector; DUAL-ENERGY CT; MATERIAL DECOMPOSITION METHOD; URINARY STONE COMPOSITION; MICRO-CT; RENAL STONES; ELECTRON-DENSITY; HOUNSFIELD UNITS; LIVER-LESIONS; IMAGE QUALITY; URIC-ACID;
D O I
10.1016/j.radphyschem.2024.112130
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
X-ray computed tomography (CT) is widely used in various fields, including medical, industrial, and fundamental scientific applications. Its quick ability to provide three-dimensional quantification of structures makes it a valuable tool. In addition to patient diagnostics, CT methods are employed for the investigation of varied samples. This includes in-depth analysis using high-resolution micro- and nano-CT imaging. However, X-ray CT techniques not only perform morphological characterization but also allow for the direct evaluation of the effective atomic number (Zeff) and mass density (rho) of the subject under investigation. These techniques collectively fall within the class of Spectral CT methods, encompassing single-energy, dual-energy, differential phase contrast, and photon counting detector CT. Spectral CT leverages data acquired from different radiation interaction phenomena to map Zeff and rho values. This paper presents a comprehensive overview of these methodologies, elucidating the underlying theory behind each. Furthermore, we review published studies centered on the development of theoretical formulations and diverse applications of the techniques. They cover formalism development, calibration methods, and reconstruction algorithms. Additionally, we introduce the commonly utilized equipment along with case studies in medical, mineralogical, and material fields, as well as biology and archaeology. These case studies will be thoroughly evaluated and discussed within the context of this paper.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] High-resolution X-ray computed tomography in geosciences: A review of the current technology and applications
    Cnudde, V.
    Boone, M. N.
    EARTH-SCIENCE REVIEWS, 2013, 123 : 1 - 17
  • [32] Simultaneous fluorescence and Compton scattering computed tomography based on linear polarization X-ray
    Chi, Zhi-Jun
    Zhang, Hong-Ze
    Lin, Jin
    Zhang, Xuan-Qi
    Ding, Hao
    Tian, Qi-Li
    Zhang, Zhi
    Du, Ying-Chao
    Huang, Wen-Hui
    Tang, Chuan-Xiang
    NUCLEAR SCIENCE AND TECHNIQUES, 2024, 35 (10)
  • [33] High-kVp Assisted Metal Artifact Reduction for X-Ray Computed Tomography
    Xi, Yan
    Jin, Yannan
    De Man, Bruno
    Wang, Ge
    IEEE ACCESS, 2016, 4 : 4769 - 4776
  • [34] Technological developments of X-ray computed tomography over half a century: User's influence on protocol optimization
    Booij, Ronald
    Budde, Ricardo P. J.
    Dijkshoorn, Marcel L.
    van Straten, Marcel
    EUROPEAN JOURNAL OF RADIOLOGY, 2020, 131
  • [35] Modeling and characterization of X-ray yield in a polychromatic, lab-scale, X-ray computed tomography system
    Mertens, J. C. E.
    Chawla, Nikhilesh
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 783 : 110 - 116
  • [36] Not all scans are equal: X-ray tomography image quality evaluation
    du Plessis, A.
    Tshibalanganda, M.
    le Roux, S. G.
    MATERIALS TODAY COMMUNICATIONS, 2020, 22
  • [37] X-ray scatter correction method for dedicated breast computed tomography
    Sechopoulos, Ioannis
    MEDICAL PHYSICS, 2012, 39 (05) : 2896 - 2903
  • [38] Optimization of the energy for Breast monochromatic absorption X-ray Computed Tomography
    Delogu, Pasquale
    Di Trapani, Vittorio
    Brombal, Luca
    Mettivier, Giovanni
    Taibi, Angelo
    Oliva, Piernicola
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [39] Optimal Iodine Staining of Cardiac Tissue for X-Ray Computed Tomography
    Butters, Timothy D.
    Castro, Simon J.
    Lowe, Tristan
    Zhang, Yanmin
    Lei, Ming
    Withers, Philip J.
    Zhang, Henggui
    PLOS ONE, 2014, 9 (08):
  • [40] X-ray computed tomography: from medical imaging to dimensional metrology
    Villarraga-Gomez, Herminso
    Herazo, Ericka L.
    Smith, Stuart T.
    PRECISION ENGINEERING-JOURNAL OF THE INTERNATIONAL SOCIETIES FOR PRECISION ENGINEERING AND NANOTECHNOLOGY, 2019, 60 : 544 - 569