Calibration algorithms for continuous glucose monitoring systems based on interstitial fluid sensing

被引:2
作者
Sun, Tianyi [1 ]
Liu, Jentsai [2 ]
Chen, Ching Jung [3 ,4 ]
机构
[1] Univ Chinese Acad Sci, Sch Elect Elect & Commun Engn, Beijing, Peoples R China
[2] Univ Chinese Acad Sci, Coll Mat Sci & Optielect Technol, Res Ctr Mat Sci & Optielect Technol, Beijing, Peoples R China
[3] Univ Chinese Acad Sci, Res Ctr Mat Sci & Optielectron Technol, Sch Optoelect, Beijing, Peoples R China
[4] Univ Chinese Acad Sci, Res Ctr Mat Sci & Optielect Technol, Beijing, Peoples R China
关键词
Continuous glucose monitoring (CGM); Calibration; Interstitial fluid glucose (IG); Blood glucose (BG); IG-BG kinetic; Dynamic estimation; SENSORS; KINETICS; ACCURACY; SCIENCE; PLASMA; SKIN;
D O I
10.1016/j.bios.2024.116450
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Continuous glucose monitoring (CGM) is of great importance to the treatment and prevention of diabetes. As a proven commercial technology, electrochemical glucose sensor based on interstitial fluid (ISF) sensing has high sensitivity and wide detection range. Therefore, it has good promotion prospects in noninvasive or minimallyinvasive CGM system. However, since there are concentration differences and time lag between glucose in plasma and ISF, the accuracy of this type of sensors are still limited. Typical calibration algorithms rely on simple linear regression which do not account for the variability of the sensitivity of sensors. To enhance the accuracy and stability of CGM based on ISF, optimization of calibration algorithm for sensors is indispensable. While there have been considerable researches on improving calibration algorithms for CGM, they have still received less attention. This article reviews the problem of typical calibration and presents the outstanding calibration algorithms in recent years. Finally, combined with existing research and emerging sensing technologies, this paper makes an outlook on the future calibration algorithms for CGM sensors.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] A method for continuous glucose monitoring in interstitial fluid based on microdialysis
    Li, D. (dchli@tju.edu.cn), 1600, Tianjin University (46): : 482 - 486
  • [2] Estimating Plasma Glucose from Interstitial Glucose: The Issue of Calibration Algorithms in Commercial Continuous Glucose Monitoring Devices
    Rossetti, Paolo
    Bondia, Jorge
    Vehi, Josep
    Fanelli, Carmine G.
    SENSORS, 2010, 10 (12) : 10936 - 10952
  • [3] An Interstitial Fluid Transdermal Extraction System for Continuous Glucose Monitoring
    Yu, Haixia
    Li, Dachao
    Roberts, Robert C.
    Xu, Kexin
    Tien, Norman C.
    JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2012, 21 (04) : 917 - 925
  • [4] Kalman-Based Calibration Algorithm for AgaMatrix Continuous Glucose Monitoring System
    Singh, Abhinoy Kumar
    Rebec, Mihailo V.
    Haidar, Ahmad
    IEEE TRANSACTIONS ON CONTROL SYSTEMS TECHNOLOGY, 2021, 29 (03) : 1257 - 1267
  • [5] Glucose measurement in interstitial fluid by microdialysis for the calibration of minimally invasive blood glucose monitoring
    Li, Dachao
    Wang, Ridong
    Chong, Hao
    Liu, Yu
    Xu, Kexin
    ADVANCED BIOMEDICAL AND CLINICAL DIAGNOSTIC SYSTEMS XI, 2013, 8572
  • [6] MICRONEEDLE-BASED SYSTEM FOR MINIMALLY INVASIVE CONTINUOUS MONITORING OF GLUCOSE IN THE DERMAL INTERSTITIAL FLUID
    Ribet, Federico
    Stemme, Goran
    Roxhed, Niclas
    2018 IEEE MICRO ELECTRO MECHANICAL SYSTEMS (MEMS), 2018, : 408 - 411
  • [7] Interstitial fluid glucose time-lag correction for real-time continuous glucose monitoring
    Keenan, D. Barry
    Mastrototaro, John J.
    Weinzimer, Stuart A.
    Steil, Garry M.
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2013, 8 (01) : 81 - 89
  • [8] Impact of Retrospective Calibration Algorithms on Hypoglycemia Detection in Newborn Infants Using Continuous Glucose Monitoring
    Signal, Matthew
    Le Compte, Aaron
    Harris, Deborah L.
    Weston, Philip J.
    Harding, Jane E.
    Chase, J. Geoffrey
    DIABETES TECHNOLOGY & THERAPEUTICS, 2012, 14 (10) : 883 - 890
  • [9] Differences in glucose readings between the continuous glucose monitoring calibration free interstitial sensors versus capillary blood glucose monitoring by glucometer: An analysis of two cases
    Quah, Phaik Ling
    Chai, Sally Mun Hua
    Tan, Kok Hian
    METABOLISM OPEN, 2024, 22
  • [10] Consistency of Continuous Ambulatory Interstitial Glucose Monitoring Sensors
    Wu, Pei T.
    Segovia, David E.
    Lee, Cathy C.
    Nguyen, Kim-Lien
    BIOSENSORS-BASEL, 2018, 8 (02):