Redesigned Electrodes for Improved Intraoperative Nerve Conduction Studies during the Treatment of Peripheral Nerve Injuries

被引:0
作者
Riemann, Nathaniel [1 ]
Coursen, Jack [1 ]
Porras, Laura Elena [1 ]
Sabogal, Bryan [1 ]
Liang, Xin-Hua [1 ]
Guaraca, Christian [1 ]
Belzberg, Allan [2 ]
Ringkamp, Matthias [2 ]
Wu, Gang [2 ]
Zhu, Lily [1 ]
Weed, Samantha [1 ]
Miranda, Constanza [1 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21218 USA
[2] Johns Hopkins Univ, Sch Med, Dept Neurosurg, Baltimore, MD 21287 USA
关键词
peripheral nerve; nerve conduction study; electrode; intraoperative neuromonitoring; action potential; signal artifact; nerve trauma injury; neurosurgical treatment; design process; clinical need;
D O I
10.3390/healthcare12131269
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Traumatic peripheral nerve injuries (PNI), present with symptoms ranging from pain to loss of motor and sensory function. Difficulties in intraoperative visual assessment of nerve functional status necessitate intraoperative nerve conduction studies (INCSs) by neurosurgeons and neurologists to determine the presence of functioning axons in the zone of a PNI. This process, also referred to as nerve "inching", uses a set of stimulating and recording electrode hooks to lift the injured nerve from the surrounding surgical field and to determine whether an electrical stimulus can travel through the zone of injury. However, confounding electrical signal artifacts can arise from the current workflow and electrode design, particularly from the mandatory lifting of the nerve, complicating the definitive assessment of nerve function and neurosurgical treatment decision-making. The objective of this study is to describe the design process and verification testing of our group's newly designed stimulating and recording electrodes that do not require the lifting or displacement of the injured nerve during INCSs. Ergonomic in vivo analysis of the device within a porcine model demonstrated successful intraoperative manipulation of the device, while quantitative nerve action potential (NAP) signal analysis with an ex vivo simulated "inching" procedure on healthy non-human primate nerve tissue demonstrated excellent reproducible recorded NAP fidelity and the absence of NAP signal artifacts at all points of recording. Lastly, electrode pullout force testing determined maximum forces of 0.43 N, 1.57 N, and 3.61 N required to remove the device from 2 mm, 5 mm, and 1 cm nerve models, respectively, which are well within established thresholds for nerve safety. These results suggest that these new electrodes can safely and successfully perform accurate PNI assessment without the presence of artifacts, with the potential to improve the INCS standard of care while remaining compatible with currently used neurosurgical technology, infrastructure, and clinical workflows.
引用
收藏
页数:14
相关论文
共 28 条
[11]   New Backstrap Vagus Electrode for Continuous Intraoperative Neuromonitoring in Thyroid Surgery [J].
Lamade, Wolfram ;
Ulmer, Christoph ;
Rieber, Fabian ;
Friedrich, Colin ;
Koch, Klaus P. ;
Thon, Klaus-Peter .
SURGICAL INNOVATION, 2011, 18 (03) :206-213
[12]   Requirements - Where do requirements come from? [J].
Maiden, N ;
Gizikis, A .
IEEE SOFTWARE, 2001, 18 (05) :10-12
[13]   A scoping review of current and emerging techniques for evaluation of peripheral nerve health, degeneration, and regeneration: part 1, neurophysiology [J].
Mandeville, Ross ;
Sanchez, Benjamin ;
Johnston, Benjamin ;
Bazarek, Stanley ;
Thum, Jasmine A. ;
Birmingham, Austin ;
See, Reiner Henson B. ;
Leochico, Carl Froilan D. ;
Kumar, Viksit ;
Dowlatshahi, Arriyan S. ;
Brown, Justin ;
Stashuk, Daniel ;
Rutkove, Seward B. .
JOURNAL OF NEURAL ENGINEERING, 2023, 20 (04)
[14]   Effect of human grip strategy on force control in precision tasks [J].
McDonnell, MN ;
Ridding, MC ;
Flavel, SC ;
Miles, TS .
EXPERIMENTAL BRAIN RESEARCH, 2005, 161 (03) :368-373
[15]   Five Qualitative Research Concepts Grounded in Anthropological Methods for Teaching Design in Healthcare [J].
Miranda, Constanza ;
Goni, Julian ;
Labruto, Nicole .
HEALTHCARE, 2022, 10 (02)
[16]   Developing an Innovative Medical Training Simulation Device for Peripheral Venous Access: A User-Centered Design Approach [J].
Miranda, Constanza ;
Altermatt, Fernando ;
Villagran, Ignacio ;
Goni, Julian .
HEALTHCARE, 2020, 8 (04)
[17]   Functional and Mechanical Evaluation of Nerve Stretch Injury [J].
Rickett, Todd ;
Connell, Sean ;
Bastijanic, Jennifer ;
Hegde, Satya ;
Shi, Riyi .
JOURNAL OF MEDICAL SYSTEMS, 2011, 35 (05) :787-793
[18]   INTRAOPERATIVE NERVE ACTION POTENTIAL RECORDINGS: TECHNICAL CONSIDERATIONS, PROBLEMS, AND PITFALLS [J].
Robert, Everett G. ;
Happel, Leo T. ;
Kline, David G. .
NEUROSURGERY, 2009, 65 (04) :A97-A104
[19]  
Robinson LR, 2000, MUSCLE NERVE, V23, P863, DOI 10.1002/(SICI)1097-4598(200006)23:6<863::AID-MUS4>3.0.CO
[20]  
2-0