PND-Net: Physics-Inspired Non-Local Dual-Domain Network for Metal Artifact Reduction

被引:1
作者
Xia, Jinqiu [1 ]
Zhou, Yiwen [1 ]
Deng, Wenxin [1 ]
Kang, Jing [1 ]
Wu, Wangjiang [1 ]
Qi, Mengke [1 ]
Zhou, Linghong [1 ]
Ma, Jianhui [2 ]
Xu, Yuan [1 ]
机构
[1] Southern Med Univ, Sch Biomed Engn, Guangzhou 510515, Peoples R China
[2] Southern Med Univ, Nanfang Hosp, Dept Radiat Oncol, Guangzhou 510515, Peoples R China
关键词
Metals; Computed tomography; Implants; Image restoration; Photonics; Phantoms; Image reconstruction; Metal artifact reduction; non-local dual-domain network; unpaired learning; BEAM HARDENING CORRECTION; CT;
D O I
10.1109/TMI.2024.3354925
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Metal artifacts caused by the presence of metallic implants tremendously degrade the quality of reconstructed computed tomography (CT) images and therefore affect the clinical diagnosis or reduce the accuracy of organ delineation and dose calculation in radiotherapy. Although various deep learning methods have been proposed for metal artifact reduction (MAR), most of them aim to restore the corrupted sinogram within the metal trace, which removes beam hardening artifacts but ignores other components of metal artifacts. In this paper, based on the physical property of metal artifacts which is verified via Monte Carlo (MC) simulation, we propose a novel physics-inspired non-local dual-domain network (PND-Net) for MAR in CT imaging. Specifically, we design a novel non-local sinogram decomposition network (NSD-Net) to acquire the weighted artifact component and develop an image restoration network (IR-Net) to reduce the residual and secondary artifacts in the image domain. To facilitate the generalization and robustness of our method on clinical CT images, we employ a trainable fusion network (F-Net) in the artifact synthesis path to achieve unpaired learning. Furthermore, we design an internal consistency loss to ensure the data fidelity of anatomical structures in the image domain and introduce the linear interpolation sinogram as prior knowledge to guide sinogram decomposition. NSD-Net, IR-Net, and F-Net are jointly trained so that they can benefit from one another. Extensive experiments on simulation and clinical data demonstrate that our method outperforms state-of-the-art MAR methods.
引用
收藏
页码:2125 / 2136
页数:12
相关论文
共 45 条
  • [1] Metal artifact reduction in CT using tissue-class modeling and adaptive prefiltering
    Bal, Matthieu
    Spies, Lothar
    [J]. MEDICAL PHYSICS, 2006, 33 (08) : 2852 - 2859
  • [2] Directional View Interpolation for Compensation of Sparse Angular Sampling in Cone-Beam CT
    Bertram, Matthias
    Wiegert, Jens
    Schaefer, Dirk
    Aach, Til
    Rose, Georg
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2009, 28 (07) : 1011 - 1022
  • [3] Reduction of metal streak artifacts in x-ray computed tomography using a transmission maximum a posteriori algorithm
    De Man, B
    Nuyts, J
    Dupont, P
    Marchal, G
    Suetens, P
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (03) : 977 - 981
  • [4] Ghani W. C., IEEE Trans. Comput. Imag., V6, P181
  • [5] Metal artifacts in computed tomography for radiation therapy planning: dosimetric effects and impact of metal artifact reduction
    Giantsoudi, Drosoula
    De Man, Bruno
    Verburg, Joost
    Trofimov, Alexei
    Jin, Yannan
    Wang, Ge
    Gjesteby, Lars
    Paganetti, Harald
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (08) : R49 - R80
  • [6] Metal Artifact Reduction in CT: Where Are We After Four Decades?
    Gjesteby, Lars
    De Man, Bruno
    Jin, Yannan
    Paganetti, Harald
    Verburg, Joost
    Giantsoudi, Drosoula
    Wang, Ge
    [J]. IEEE ACCESS, 2016, 4 : 5826 - 5849
  • [7] Goodfellow IJ, 2014, ADV NEUR IN, V27, P2672
  • [8] Gu JW, 2006, J X-RAY SCI TECHNOL, V14, P11
  • [9] Gu JW, 2006, J X-RAY SCI TECHNOL, V14, P73
  • [10] An iterative approach to the beam hardening correction in cone beam CT
    Hsieh, J
    Molthen, RC
    Dawson, CA
    Johnson, RH
    [J]. MEDICAL PHYSICS, 2000, 27 (01) : 23 - 29