Novel multileaf collimator designs with tongues and grooves in leaf end and Monte Carlo simulations

被引:2
作者
Cui, Weijie [1 ]
Tian, Yuan [1 ]
Dai, Jianrong [1 ]
机构
[1] Chinese Acad Med Sci & Peking Union Med Coll, Canc Hosp, Natl Canc Ctr, Dept Radiat Oncol, Beijing 100021, Peoples R China
关键词
Multileaf collimator; Leaf end; Leaf gap; Tongue and groove; SYSTEM; IMRT; GAP;
D O I
10.1016/j.meddos.2024.01.008
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
In this study, we proposed 2 new multileaf collimator leaf designs to eliminate leaf gaps for closed leaf pairs so that radiation leakage can be avoided. In the new designs, multi tongues and grooves were added to the conventional multileaf collimators leaf ends. Thus, when a pair of leaves closed, tongues of a leaf can enter grooves of its opposing leaf. Consequently, there would be no radiation leakage through closed leaves. One design was named finger-shaped MLC, and another design with doubled leaf end thickness was named hand-shaped MLC. Monte Carlo simulations were performed to simulate dosimetric characteristics of the new MLC designs and comparison to conventional MLCs was performed. The simulations show that for the closed field, the new designs reduce leakage dramatically. And for the open field, the finger-shaped MLC has a larger penumbra width than conventional MLC, while the penumbra for the hand-shaped MLC is comparable to that of conventional MLC. With the application of new MLC designs, it is expected to eliminate leaf gaps for MLC usage and protect normal tissues better. (c) 2024 Published by Elsevier Inc. on behalf of American Association of Medical Dosimetrists.
引用
收藏
页码:254 / 262
页数:9
相关论文
共 17 条
[11]   Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy [J].
LoSasso, T ;
Chui, CS ;
Ling, CC .
MEDICAL PHYSICS, 1998, 25 (10) :1919-1927
[12]   On the use of the MLC dosimetric leaf gap as a quality control tool for accurate dynamic IMRT delivery [J].
Mei, Xiangyang ;
Nygren, Ian ;
Villarreal-Barajas, J. Eduardo .
MEDICAL PHYSICS, 2011, 38 (04) :2246-2255
[13]   Machine QA for the Elekta Unity system: A Report from the Elekta MR-linac consortium [J].
Roberts, David A. ;
Sandin, Carlos ;
Vesanen, Panu T. ;
Lee, Hannah ;
Hanson, Ian M. ;
Nill, Simeon ;
Perik, Thijs ;
Lim, Seng Boh ;
Vedam, Sastry ;
Yang, Jinzhong ;
Woodings, Simon W. ;
Wolthaus, Jochem W. H. ;
Keller, Brian ;
Budgell, Geoff ;
Chen, Xinfeng ;
Li, X. Allen .
MEDICAL PHYSICS, 2021, 48 (05) :E67-E85
[14]   Commissioning of a 1.5T Elekta Unity MR-linac: A single institution experience [J].
Snyder, Jeffrey E. ;
St-Aubin, Joel ;
Yaddanapudi, Sridhar ;
Boczkowski, Amanda ;
Dunkerley, David A. P. ;
Graves, Stephen A. ;
Hyer, Daniel E. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (07) :160-172
[15]   Longitudinal assessment of quality assurance measurements in a 1.5T MR-linac: Part I-Linear accelerator [J].
Subashi, Ergys ;
Lim, Seng Boh ;
Gonzalez, Xesus ;
Tyagi, Neelam .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (10) :190-201
[16]   A six-bank multi-leaf system for high precision shaping of large fields [J].
Topolnjak, R ;
van der Heide, UA ;
Raaymakers, BW ;
Kotte, ANTJ ;
Welleweerd, J ;
Lagendijk, JJW .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (12) :2645-2656
[17]   Rounded leaf end modeling in Pinnacle VMAT treatment planning for fixed jaw linacs [J].
Young, Lori A. ;
Yang, Fei ;
Cao, Ning ;
Meyer, Juergen .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (06) :149-162