Experimental Evaluation of Analgesia With an Event-Based PID Control Strategy for Anesthesia

被引:0
作者
Schiavo, Michele [1 ]
Latronico, Nicola [2 ,3 ]
Paltenghi, Massimiliano [3 ]
Visioli, Antonio [1 ]
机构
[1] Univ Brescia, Dipartimentodi Ingn Meccan & Ind, I-25123 Brescia, Italy
[2] Univ Brescia, Dept Med & Surg Specialties, Radiol Sci & Publ Hlth, I-25123 Brescia, Italy
[3] Spedali Civili Brescia, Dept Anesthesiol Crit Care & Emergency, I-25123 Brescia, Italy
来源
IEEE CONTROL SYSTEMS LETTERS | 2024年 / 8卷
关键词
Indexes; Monitoring; Maintenance; Control systems; Anesthesia; MIMO communication; Anethesia control; PID control; event-based control; DRUG-DELIVERY SYSTEM; CLOSED-LOOP CONTROL; BISPECTRAL INDEX; PROPOFOL; REMIFENTANIL; NOCICEPTION; DESIGN; QCON; FEASIBILITY; STIMULATION;
D O I
10.1109/LCSYS.2024.3411518
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
The recent introduction of the Conox monitoring system in the clinical practice of total intravenous anesthesia (TIVA) has opened new research opportunities. Indeed, it provides the qCON and the qNOX indexes, which give a measure of hypnosis and analgesia, respectively. This letter presents the first experimental results of closed-loop anesthesia performed by using the Conox monitoring system. In particular, we exploited the qCON as the feedback variable for an event-based PID multiple-input single-output (MISO) controller and we recorded the qNOX to evaluate analgesia. Clinical experiments have been performed on four patients undergoing general anesthesia for elective plastic surgery. The controller demonstrated a satisfactory performance in maintaining qCON within the desired range for all patients. An optimal qNOX level has also been achieved in three out of four cases, while, in one patient, the qNOX level indicated a slight excess of analgesia. These findings suggest that the proposed MISO architecture is effective in providing adequate levels of hypnosis and analgesia and that, in future works, the qNOX information could be leveraged in the controller design to enhance analgesia control.
引用
收藏
页码:1427 / 1432
页数:6
相关论文
共 33 条
  • [1] Pharmacodynamic interaction between propofol and remifentanil regarding hypnosis, tolerance of laryngoscopy, bispectral index, and electroencephalographic approximate entropy
    Bouillon, TW
    Bruhn, J
    Radulescu, L
    Andresen, C
    Shafer, TJ
    Cohane, C
    Shafer, SL
    [J]. ANESTHESIOLOGY, 2004, 100 (06) : 1353 - 1372
  • [2] Comparison of the Conox (qCON) and Sedline (PSI) depth of anaesthesia indices to predict the hypnotic effect during desflurane general anaesthesia with ketamine
    Christenson, Catherine
    Martinez-Vazquez, Pablo
    Breidenstein, Max
    Farhang, Borzoo
    Mathews, Jackson
    Melia, Umberto
    Jensen, Erik Weber
    Mathews, Donald
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2021, 35 (06) : 1421 - 1428
  • [3] Preliminary Intraoperative Validation of the Nociception Level Index A Noninvasive Nociception Monitor
    Edry, Ruth
    Recea, Vasile
    Dikust, Yuri
    Sessler, Daniel I.
    [J]. ANESTHESIOLOGY, 2016, 125 (01) : 193 - 203
  • [4] Closed-Loop Control of Anesthesia: Survey on Actual Trends, Challenges and Perspectives
    Ghita, Mihaela
    Neckebroek, Martine
    Muresan, Cristina
    Copot, Dana
    [J]. IEEE ACCESS, 2020, 8 : 206264 - 206279
  • [5] Influence of nociceptive stimulation on analgesia nociception index (ANI) during propofol-remifentanil anaesthesia
    Gruenewald, M.
    Ilies, C.
    Herz, J.
    Schoenherr, T.
    Fudickar, A.
    Hoecker, J.
    Bein, B.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2013, 110 (06) : 1024 - 1030
  • [6] Evaluation of a novel closed-loop total intravenous anaesthesia drug delivery system: a randomized controlled trial
    Hemmerling, T. M.
    Arbeid, E.
    Wehbe, M.
    Cyr, S.
    Taddei, R.
    Zaouter, C.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2013, 110 (06) : 1031 - 1039
  • [7] An Open Source Patient Simulator for Design and Evaluation of Computer Based Multiple Drug Dosing Control for Anesthetic and Hemodynamic Variables
    Ionescu, Clara M.
    Neckebroek, Martine
    Ghita, Mihaela
    Copot, Dana
    [J]. IEEE ACCESS, 2021, 9 : 8680 - 8694
  • [8] Monitoring hypnotic effect and nociception with two EEG-derived indices, qCON and qNOX, during general anaesthesia
    Jensen, E. W.
    Valencia, J. F.
    Lopez, A.
    Anglada, T.
    Agusti, M.
    Ramos, Y.
    Serra, R.
    Jospin, M.
    Pineda, P.
    Gambus, P.
    [J]. ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2014, 58 (08) : 933 - 941
  • [9] Feasibility of Closed-loop Titration of Propofol and Remifentanil Guided by the Spectral M-Entropy Monitor
    Liu, Ngai
    Le Guen, Morgan
    Benabbes-Lambert, Fatima
    Chazot, Thierry
    Trillat, Bernard
    Sessler, Daniel I.
    Fischler, Marc
    [J]. ANESTHESIOLOGY, 2012, 116 (02) : 286 - 295
  • [10] Closed-Loop Coadministration of Propofol and Remifentanil Guided by Bispectral Index: A Randomized Multicenter Study
    Liu, Ngai
    Chazot, Thierry
    Hamada, Sophie
    Landais, Alain
    Boichut, Nathalie
    Dussaussoy, Corinne
    Trillat, Bernard
    Beydon, Laurent
    Samain, Emmanuel
    Sessler, Daniel I.
    Fischler, Marc
    [J]. ANESTHESIA AND ANALGESIA, 2011, 112 (03) : 546 - 557