摘要
The importance of efficient diabetes treatment and its' reliable automation is rising with the prevalence of this chronic condition worldwide. Robustness is one of the enablers of the safe automation of plasma glucose control, for it can ensure consistent behavior when the controlled dynamics are changing or partially unknown. Hence, this work focuses on assessing the capabilities of a robust nonlinear controller framework in a simulated environment. Linear Parameter -Varying modeling is combined with robust control techniques, supported by an Unscented Kalman filter. These controllers are then subjected to additional constraints in search of a practical trade-off between disturbance rejection and the severity of transient behavior. Simulations with unannounced meal intakes explore and compare the effect of these additional constraints and configurations using the virtual patients provided by a well-known in silico simulator. The simulation results indicate that this control method can ensure adequate blood glucose control and has the potential to support other control algorithms to realize a safe and reliable artificial pancreas.
引用
收藏
页码:27 / 46
页数:20
相关论文
共 57 条
[1]   Glycemic Monitoring in Critical Care using Nonlinear State Estimators [J].
Aguirre-Zapata, E. ;
Cardenas-Cartagena, J. ;
Garcia-Tirado, J. .
IFAC PAPERSONLINE, 2017, 50 (01) :4430-4435
[2]   Clinical Use of Real-Time Continuous Glucose Monitoring [J].
Battelino, Tadej ;
Bolinder, Jan .
CURRENT DIABETES REVIEWS, 2008, 4 (03) :218-222
[3]   PHYSIOLOGIC EVALUATION OF FACTORS CONTROLLING GLUCOSE-TOLERANCE IN MAN - MEASUREMENT OF INSULIN SENSITIVITY AND BETA-CELL GLUCOSE SENSITIVITY FROM THE RESPONSE TO INTRAVENOUS GLUCOSE [J].
BERGMAN, RN ;
PHILLIPS, LS ;
COBELLI, C .
JOURNAL OF CLINICAL INVESTIGATION, 1981, 68 (06) :1456-1467
[4]   Invalidation and low-order model set for artificial pancreas robust control design [J].
Bianchi, Fernando D. ;
Moscoso-Vasquez, Marcela ;
Colmegna, Patricio ;
Sanchez-Pena, Ricardo S. .
JOURNAL OF PROCESS CONTROL, 2019, 76 :133-140
[5]  
Blanc R, 2019, IEEE ENG MED BIO, P5465, DOI [10.1109/EMBC.2019.8857170, 10.1109/embc.2019.8857170]
[6]  
Boiroux Dimitri, 2015, IFAC - Papers Online, V48, P159, DOI 10.1016/j.ifacol.2015.10.132
[7]  
Boiroux D, 2016, 2016 EUROPEAN CONTROL CONFERENCE (ECC), P2115, DOI 10.1109/ECC.2016.7810604
[8]   Impact of obesity on the increasing incidence of type 1 diabetes [J].
Buzzetti, Raffaella ;
Zampetti, Simona ;
Pozzilli, Paolo .
DIABETES OBESITY & METABOLISM, 2020, 22 (07) :1009-1013
[9]  
Calupina D., 2018, 2018 IEEE 3 EC TECHN, P1
[10]   Glycemic control in the intensive care unit: A control systems perspective [J].
Chase, J. Geoffrey ;
Benyo, Balazs ;
Desaive, Thomas .
ANNUAL REVIEWS IN CONTROL, 2019, 48 :359-368