A rare case of taper junction corrosion in semi-constrained total knee arthroplasty

被引:0
作者
Ekweariri, Nnadozie [1 ]
White, Ryan [1 ]
Brown, Nicholas [1 ]
Schmitt, Daniel [1 ]
机构
[1] Loyola Univ, Med Ctr, Dept Orthopaed Surg, 2160 South 1st Ave, Maywood, IL 60153 USA
关键词
Total knee arthroplasty; Revision; Metallosis; Trunnionosis; IMPLANT; METALLOSIS; FAILURE; CREVICE; STEMS;
D O I
10.1016/j.knee.2024.02.010
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Metallosis is a known yet rare late complication of unicompartmental and total knee arthroplasty (TKA), usually secondary to either metal -backed patellar component failure, mobile -bearing polyethylene dislocation, or catastrophic polyethylene failure and wear through. The majority of literature surrounding metallosis has been published in relation to total hip arthroplasty (THA) metal on metal bearing wear or mechanically assisted crevice corrosion. This case report describes the development of metallosis in a 77 -year -old male patient with advanced (Kellgren-Lawrence Grade 4) osteoarthritis with associated valgus deformity, who underwent index TKA with a semiconstrained revision knee system due to intraoperative medial collateral ligament laxity. The taper junction between the titanium alloy stem and cobalt chromium femoral component was the source of diffuse intra-articular metallosis. (c) 2024 Elsevier B.V. All rights reserved.
引用
收藏
页码:46 / 51
页数:6
相关论文
共 24 条
  • [1] Mechanically Assisted Taper Corrosion in Modular TKA
    Arnholt, Christina M.
    MacDonald, Daniel W.
    Tohfafarosh, Mariya
    Gilbert, Jeremy L.
    Rimnac, Clare M.
    Kurtz, Steven M.
    Klein, Gregg
    Mont, Michael A.
    Parvizi, Javad
    Cates, Harold E.
    Lee, Gwo-Chin
    Malkani, Arthur
    Kraay, Mattheuw
    [J]. JOURNAL OF ARTHROPLASTY, 2014, 29 (09) : 205 - 208
  • [2] Arredondo J, 1997, Am J Orthop (Belle Mead NJ), V26, P598
  • [3] Trunnion corrosion WHAT SURGEONS NEED TO KNOW IN 2018
    Berstock, J. R.
    Whitehouse, M. R.
    Duncan, C. P.
    [J]. BONE & JOINT JOURNAL, 2018, 100B (01) : 44 - 49
  • [4] FRETTING CORROSION ACCELERATES CREVICE CORROSION OF MODULAR HIP TAPERS
    BROWN, SA
    FLEMMING, CAC
    KAWALEC, JS
    PLACKO, HE
    VASSAUX, C
    MERRITT, K
    PAYER, JH
    KRAAY, MJ
    [J]. JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) : 19 - 26
  • [5] Trunnion Failure in Revision Total Knee Arthroplasty
    Chihab, Samir
    Grommersch, Bryan M.
    Erens, Greg A.
    Guild III, George N.
    [J]. ARTHROPLASTY TODAY, 2021, 8 : 258 - 261
  • [6] Adverse Local Tissue Reaction Arising from Corrosion at the Femoral Neck-Body Junction in a Dual-Taper Stem with a Cobalt-Chromium Modular Neck
    Cooper, H. John
    Urban, Robert M.
    Wixson, Richard L.
    Meneghini, R. Michael
    Jacobs, Joshua J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (10) : 865 - 872
  • [7] Corrosion at the Head-Neck Taper as a Cause for Adverse Local Tissue Reactions After Total Hip Arthroplasty
    Cooper, H. John
    Della Valle, Craig J.
    Berger, Richard A.
    Tetreault, Matthew
    Paprosky, Wayne G.
    Sporer, Scott M.
    Jacobs, Joshua J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2012, 94A (18) : 1655 - 1661
  • [8] Klontz Karl C, 2014, J Orthop Case Rep, V4, P21, DOI 10.13107/jocr.2250-0685.142
  • [9] Koonen Laurens, 2020, Arthroplast Today, V6, P262, DOI 10.1016/j.artd.2020.02.015
  • [10] Proximal Component Modularity in THA-At What Cost?: An Implant Retrieval Study
    Kop, A. M.
    Keogh, C.
    Swarts, E.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (07) : 1885 - 1894