Proximal cementation of a collarless polished tapered hip stem: biomechanical analysis using a validated finite element model

被引:0
|
作者
Ling, Carol Sze Yee [1 ]
Izmin, Aiman [2 ]
Todo, Mitsugu [2 ,3 ]
Merican, Azhar M. [1 ]
Chong, Desmond Y. R. [1 ,4 ]
机构
[1] Univ Malaya, Fac Med, Dept Orthopaed Surg, Kuala Lumpur, Malaysia
[2] Kyushu Univ, Interdisciplinary Grad Sch Engn Sci, Fukuoka, Japan
[3] Kyushu Univ, Res Inst Appl Mech, Fukuoka, Japan
[4] Singapore Inst Technol, Engn Cluster, 10 Dover Dr, Singapore 138683, Singapore
关键词
Total hip replacement; Finite element analysis; Cemented fixation; Proximal cementing; Validated FE model; FEMORAL STEM; ARTHROPLASTY; FRACTURE; PREDICTION; PATTERNS; STRENGTH; BEHAVIOR; SURFACE; STRAIN; VARUS;
D O I
10.1007/s11517-024-03152-6
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Total hip replacement (THR) with cemented stem is a common procedure for patients with hip osteoarthritis. When primary THR fails, removal of the cement is problematic and poses challenges during revision surgeries. The possibility of proximal partial cementing of the hip stem was explored to mitigate the problem. 3D finite element analysis was performed to investigate the feasibility of reduced cement length for effective implant fixation and load transmission. Three levels of cement reduction (40 mm, 80 mm, and 100 mm) in the femoral stem were evaluated. All models were assigned loadings of peak forces acting on the femur during walking and stair climbing. The experimental and predicted max/min principal bone strains were fitted into regression models and showed good correlations. FE results indicated stress increment in the femoral bone, stem, and cement due to cement reduction. A notable increase of bone stress was observed with large cement reduction of 80-100 mm, particularly in Gruen zones 3 and 5 during walking and Gruen zones 3 and 6 during stair climbing. The increase of cement stresses could be limited to 11% with a cement reduction of 40 mm. The findings suggested that a 40-mm cement reduction in hip stem fixation was desirable to avoid unwanted complications after cemented THR.
引用
收藏
页码:3531 / 3542
页数:12
相关论文
共 50 条
  • [31] Total hip replacement with a collarless polished cemented anatomic stem: clinical and gait analysis results at ten years follow-up
    Arthur Grzesiak
    Kamiar Aminian
    Estelle Lécureux
    Florence Jobin
    Brigitte M. Jolles
    International Orthopaedics, 2014, 38 : 717 - 724
  • [32] Biomechanical analysis of atlantoaxial dorsal fixation using finite element models
    Bae, Beomju
    Kim, Dongwook
    Oh, Hyejong
    Kim, Gonhyung
    THAI JOURNAL OF VETERINARY MEDICINE, 2022, 52 (01): : 185 - 192
  • [33] Development and Finite Element (FE) analysis of a novel short hip stem concept
    Jahnke, Alexander
    Harz, Torben
    Ulloa, Carlos Alfonso Fonseca
    Ishaque, Bernd Alexander
    Rickert, Markus
    JOURNAL OF ORTHOPAEDICS, 2023, 46 : 117 - 123
  • [34] Minimum 10-Year Follow-Up of Total Hip Arthroplasty With a Collarless Triple-Tapered Polished Cemented Stem With Line-to-Line Implantation Using a Direct Anterior Approach
    Miyamoto, Shuichi
    Iida, Satoshi
    Suzuki, Chiho
    Kawamoto, Taisei
    Shinada, Yoshiyuki
    Ohtori, Seiji
    JOURNAL OF ARTHROPLASTY, 2022, 37 (11) : 2214 - 2224
  • [35] Finite Element Analysis and Biomechanical Testing of the New MiniMIS Short Stem
    Trieb, Klemens
    Huber, Dominik
    Sonntag, Robert
    Kretzer, Jan Philippe
    ZEITSCHRIFT FUR ORTHOPADIE UND UNFALLCHIRURGIE, 2019, 157 (02): : 188 - 193
  • [36] Finite Element Analysis of Porous Titanium Alloy Hip Stem to Evaluate the Biomechanical Performance During Walking and Stair Climbing
    Sami Emad Alkhatib
    Hassan Mehboob
    Faris Tarlochan
    Journal of Bionic Engineering, 2019, 16 : 1103 - 1115
  • [37] The Impact of Coronal Configuration of the Proximal Femur on its Mechanical Properties and the Validation of a New Theoretical Model: Finite Element Analysis and Biomechanical Examination
    Zhang, Lijia
    Zhu, Baozhang
    Chen, Liwan
    Wang, Wenqing
    Zhang, Xiaoyong
    Zhang, Jianguo
    ORTHOPAEDIC SURGERY, 2023, 15 (01) : 62 - 69
  • [38] Validated finite element models of the proximal femur using two-dimensional projected geometry and bone density
    Den Buijs, Jorn Op
    Dragomir-Daescu, Dan
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2011, 104 (02) : 168 - 174
  • [39] PREDICTION OF PROXIMAL FEMORAL FRACTURE IN SIDEWAYS FALLS USING NONLINEAR DYNAMIC FINITE ELEMENT ANALYSIS
    Zhang, Rui
    Gong, He
    Fang, Juan
    Gao, Zhenhai
    Zhu, Dong
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2014, 14 (02)
  • [40] Effects of Impactor Size on Biomechanical Characteristics of Spinal Cord in Hemicontusion Injury Model Using Finite Element Analysis
    Khuyagbaatar, Batbayar
    Kim, Kyungsoo
    Batbayar, Temuujin
    Kim, Yoon Hyuk
    APPLIED SCIENCES-BASEL, 2020, 10 (12):