Canine Cerebrospinal Fluid Analysis Using Two New Automated Techniques: The Sysmex XN-V Body Fluid Mode and an Artificial-Intelligence-Based Algorithm

被引:0
作者
Lapsina, Sandra [1 ,2 ]
Riond, Barbara [1 ]
Hofmann-Lehmann, Regina [1 ]
Stirn, Martina [1 ,3 ]
机构
[1] Univ Zurich, Vetsuisse Fac, Dept Clin Diagnost & Serv, Clin Lab, Winterthurerstr 260, CH-8057 Zurich, Switzerland
[2] SIA Laboklin, Fridriha Candera Iela 4, LV-1046 Riga, Latvia
[3] Roche Innovat Ctr, Roche Pharm Res & Early Dev Pharmaceut Sci, CH-4070 Basel, Switzerland
来源
ANIMALS | 2024年 / 14卷 / 11期
关键词
canine; cerebrospinal fluid; Sysmex; body fluid mode; regating; total nucleated cell count; differential count; deep learning; artificial intelligence; FLOW-CYTOMETRIC ANALYSIS; CELL COUNTS; BLOOD-CELLS; ADVIA-120; XE-5000;
D O I
10.3390/ani14111655
中图分类号
S8 [畜牧、 动物医学、狩猎、蚕、蜂];
学科分类号
0905 ;
摘要
Simple Summary Cerebrospinal fluid analysis is an important diagnostic test for a neurological canine patient. For this analysis, the total nucleated cell count and differential cell counts are routinely taken, but both involve time-consuming manual methods. The Sysmex XN-V body fluid mode is a specific setting used for analyzing different body fluids, including cerebrospinal fluid samples, while the deep-learning-based algorithm generated by the Olympus VS200 slide scanner allows recognition and classification of various cell types on scanned slides. In this study, the faster automated methods mentioned above were assessed by comparing them with the manual methods for both the total nucleated and differential cell counts. Manually set gates were used for the Sysmex body fluid mode following incorrect cell classifications recorded when using the predefined settings. This study demonstrates that automated methods may be used for total nucleated cell count assessments in canine cerebrospinal fluid samples, but optimization is still needed for both differential counts.Abstract Cerebrospinal fluid analysis is an important diagnostic test when assessing a neurological canine patient. For this analysis, the total nucleated cell count and differential cell counts are routinely taken, but both involve time-consuming manual methods. To investigate faster automated methods, in this study, the Sysmex XN-V body fluid mode and the deep-learning-based algorithm generated by the Olympus VS200 slide scanner were compared with the manual methods in 161 canine cerebrospinal fluid samples for the total nucleated cell count and in 65 samples with pleocytosis for the differential counts. Following incorrect gating by the Sysmex body fluid mode, all samples were reanalyzed with manually set gates. The Sysmex body fluid mode then showed a mean bias of 15.19 cells/mu L for the total nucleated cell count and mean biases of 4.95% and -4.95% for the two-part differential cell count, while the deep-learning-based algorithm showed mean biases of -7.25%, -0.03% and 7.27% for the lymphocytes, neutrophils and monocytoid cells, respectively. Based on our findings, we propose that the automated Sysmex body fluid mode be used to measure the total nucleated cell count in canine cerebrospinal fluid samples after making adjustments to the predefined settings from the manufacturer. However, the two-part differential count of the Sysmex body fluid mode and the deep-learning-based algorithm require some optimization.
引用
收藏
页数:21
相关论文
共 44 条
  • [1] Reporting of method comparison studies: a review of advice, an assessment of current practice, and specific suggestions for future reports
    Abu-Arafeh, A.
    Jordan, H.
    Drummond, G.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2016, 117 (05) : 569 - 575
  • [2] Evaluation of biological fluid analysis using the sysmex XN automatic hematology analyzer
    Aguadero, Vicente
    Cano-Corres, Ruth
    Berlanga, Eugenio
    Torra, Montserrat
    [J]. CYTOMETRY PART B-CLINICAL CYTOMETRY, 2018, 94 (05) : 680 - 688
  • [3] [Anonymous], Corporation Sysmex Meeting Various Diagnostic Demands: Total Nucleated and White Blood Cell Counts, with WBC Differentiation into Monoand Polymorphonuclear Cells
  • [4] Automated flow cytometric analysis of blood cells in cerebrospinal fluid - Analytic performance
    Aune, MW
    Becker, JL
    Brugnara, C
    Canfield, W
    Dorfman, DM
    Fiehn, W
    Fischer, G
    Fitzpatrick, P
    Flaming, TH
    Henriksen, HK
    Kunicka, JE
    Lackner, KJ
    Minchello, E
    Mullenix, PA
    Myers, M
    Petersen, A
    Ternstrom, W
    Wilson, S
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2004, 121 (05) : 690 - 700
  • [5] Automated flow cytometric cell count and differentiation of canine cerebrospinal fluid cells using the ADVIA 2120
    Becker, Martina
    Bauer, Natali
    Moritz, Andreas
    [J]. VETERINARY CLINICAL PATHOLOGY, 2008, 37 (03) : 344 - 352
  • [6] Evaluation of the XE-5000 for the automated analysis of blood cells in cerebrospinal fluid
    Boer, Klas
    Deufel, Thomas
    Reinhoefer, Mike
    [J]. CLINICAL BIOCHEMISTRY, 2009, 42 (7-8) : 684 - 691
  • [7] Lack of harmonization in high fluorescent cell automated counts with body fluids mode in ascitic, pleural, synovial, and cerebrospinal fluids
    Buoro, Sabrina
    Seghezzi, Michela
    Dominoni, Paola
    Moioli, Valentina
    Manenti, Barbara
    Previtali, Giulia
    Ottomano, Cosimo
    Lippi, Giuseppe
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2019, 41 (02) : 277 - 286
  • [8] Evaluation of Mindray BC-6800 body fluid mode for automated cerebrospinal fluid cell counting
    Buoro, Sabrina
    Seghezzi, Michela
    Mecca, Tommaso
    Vavassori, Mauro
    Crippa, Alberto
    La Gioia, Antonio
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2016, 54 (11) : 1799 - 1810
  • [9] CEREBROSPINAL-FLUID ANALYSIS
    CHRISMAN, CL
    [J]. VETERINARY CLINICS OF NORTH AMERICA-SMALL ANIMAL PRACTICE, 1992, 22 (04) : 781 - 810
  • [10] CEREBROSPINAL-FLUID
    COOK, JR
    DENICOLA, DB
    [J]. VETERINARY CLINICS OF NORTH AMERICA-SMALL ANIMAL PRACTICE, 1988, 18 (03) : 475 - 499