Effect of limb and cuff positioning on measurement of arterial blood pressure with an oscillometry device (PetMAP) in anaesthetized cats

被引:0
作者
Franchino, G. [1 ,2 ]
Fordyce, P. [1 ]
Adami, C. [1 ]
机构
[1] Univ Cambridge, Dept Vet Med, Cambridge CB3 0ES, England
[2] A4 Fawcett Court,Charles Babbage Rd, Cambridge CB3 0FZ, England
关键词
Arterial blood pressure underestimation; Feline; Oscillometry; PetMAP; Pitfalls of measurement; DOPPLER ULTRASONIC DEVICES; HEALTHY CATS; ARM POSITION; HYPOTENSION; DEFINITION; MONITOR; DOGS;
D O I
10.1016/j.tvjl.2024.106137
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Arterial blood pressure (ABP) is often measured with oscillometry during anaesthesia. Changing the height of the measuring cuff with respect to the level of the heart is known to affect oscillometry accuracy in some species; however, this effect has not been investigated in cats. The objective of this study was to determine the effects of raising and lowering the measuring cuff from standard position (level of the heart) on ABP, measured with PetMAP, in anaesthetised cats. ABP readings were obtained from 29 cats with the cuff at standard position (baseline), and 5 cm above and below the heart. The end-tidal isoflurane concentrations were maintained constant during data acquisition. There were no differences between baseline values and those measured below the heart, while ABP measured above the heart was consistently lower than baseline for both the thoracic and pelvic limbs (P < 0.001), with absolute differences of 8.2 (2.5 - 14) mmHg and 6.5 (3.0 - 15.0) mmHg, respectively. Systolic ABP readings at the pelvic limb were consistently higher than those at the thoracic limb at standard position (112 +/- 26 versus 103 +/- 21 mmHg, p = 0.010), above (106 +/- 22 versus 95 +/- 20 mmHg, p = 0.003), and below the heart (116 +/- 26 versus 107 +/- 22 mmHg, p = 0.011). This study shows that raising the cuff by 5 cm above the heart, which may become necessary during procedural positioning, results in clinically significant underestimation of ABP measured with PetMAP.
引用
收藏
页数:5
相关论文
共 29 条
[1]   ACVIM consensus statement: Guidelines for the identification, evaluation, and management of systemic hypertension in dogs and cats [J].
Acierno, Mark J. ;
Brown, Scott ;
Coleman, Amanda E. ;
Jepson, Rosanne E. ;
Papich, Mark ;
Stepien, Rebecca L. ;
Syme, Harriet M. .
JOURNAL OF VETERINARY INTERNAL MEDICINE, 2018, 32 (06) :1803-1822
[2]   Comparison of directly measured arterial blood pressure at various anatomic locations in anesthetized dogs [J].
Acierno, Mark J. ;
Domingues, Michelle E. ;
Ramos, Sara J. ;
Shelby, Amanda M. ;
da Cunha, Anderson F. .
AMERICAN JOURNAL OF VETERINARY RESEARCH, 2015, 76 (03) :266-271
[3]   Changes in Systolic Blood Pressure over Time in Healthy Cats and Cats with Chronic Kidney Disease [J].
Bijsmans, E. S. ;
Jepson, R. E. ;
Chang, Y. M. ;
Syme, H. M. ;
Elliott, J. .
JOURNAL OF VETERINARY INTERNAL MEDICINE, 2015, 29 (03) :855-861
[4]   DOPPLER ULTRASONOGRAPHIC, OSCILLOMETRIC SPHYGMOMANOMETRIC, AND PHOTOPLETHYSMOGRAPHIC TECHNIQUES FOR NONINVASIVE BLOOD-PRESSURE MEASUREMENT IN ANESTHETIZED CATS [J].
BINNS, SH ;
SISSON, DD ;
BUOSCIO, DA ;
SCHAEFFER, DJ .
JOURNAL OF VETERINARY INTERNAL MEDICINE, 1995, 9 (06) :405-414
[5]   Evaluation of an oscillometric blood pressure monitor on anesthetized cats and the effect of cuff placement and fur on accuracy [J].
Branson, KR ;
WagnerMann, CC ;
Mann, FA .
VETERINARY SURGERY, 1997, 26 (04) :347-353
[6]  
Brown S, 2007, J VET INTERN MED, V21, P542, DOI 10.1892/0891-6640(2007)21[542:GFTIEA]2.0.CO
[7]  
2
[8]   Arm Position During Ambulatory Blood Pressure Monitoring: A Review of the Evidence and Clinical Guidelines [J].
Byrd, James B. ;
Brook, Robert D. .
JOURNAL OF CLINICAL HYPERTENSION, 2014, 16 (03) :225-230
[9]   Effects of cuff size and position on the agreement between arterial blood pressure measured by Doppler ultrasound and through a dorsal pedal artery catheter in anesthetized cats [J].
Cerejo, Sofia A. ;
Teixeira-Neto, Francisco J. ;
Garofalo, Natache A. ;
Pimenta, Eutalio L. M. ;
Zanuzzo, Felipe S. ;
Klein, Adriana, V .
VETERINARY ANAESTHESIA AND ANALGESIA, 2020, 47 (02) :191-199
[10]   Comparison of two species-specific oscillometric blood pressure monitors with direct blood pressure measurement in anesthetized cats [J].
Cerejo, Sofia A. ;
Teixeira-Neto, Francisco J. ;
Garofalo, Natache A. ;
Rodrigues, Jessica C. ;
Celeita-Rodriguez, Nathalia ;
Lagos-Carvajal, Angie P. .
JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2017, 27 (04) :409-418