Effective and organ doses in patient undergoing interventional neuroradiology procedures: A multicentre study

被引:0
|
作者
D'Alessio, Andrea [1 ]
Strocchi, Sabina [2 ]
Dalmasso, Federico [3 ]
Cannillo, Barbara [1 ]
Matheoud, Roberta [1 ]
Ponzetti, Alex [3 ,4 ]
Aimonetto, Stefania [3 ]
Cernigliaro, Massimiliano [5 ]
Azzalin, Giulia [5 ]
Giorgianni, Andrea [6 ]
Natrella, Massimiliano [7 ]
Carriero, Alessandro [5 ]
Guzzardi, Giuseppe [5 ]
Brambilla, Marco [1 ]
机构
[1] Univ Hosp Maggiore Carita, Med Phys Dept, Novara, Italy
[2] ASST Sette Laghi Osped Circolo & Fdn Macchi, Med Phys Dept, Varese, Italy
[3] AUSL Valle Aosta U Parini Reg Hosp, Med Phys Dept, Aosta, Italy
[4] Univ Turin, Med Phys Postgrad Sch, Turin, Italy
[5] Univ Hosp Maggiore Carita, Radiol Dept, Novara, Italy
[6] ASST Sette Laghi Osped Circolo & Fdn Macchi, Neuroradiol Dept, Varese, Italy
[7] AUSL Valle Aosta U PariniReg Hosp, Radiol Dept, Aosta, Italy
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2024年 / 122卷
关键词
Interventional neuroradiology; Conversion factors; Dosimetry; Effective/organ dose; KERMA AREA PRODUCT; CONVERSION FACTORS; ANGIOGRAPHY; TECHNOLOGY; RADIOLOGY; DSA;
D O I
10.1016/j.ejmp.2024.103383
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Radiation doses to adult patients submitted to cerebral angiography and intracranial aneurysms treatments were assessed by using DICOM Radiation Dose Structured Reports (RDSR) and Monte Carlo simulations. Conversion factors to estimate effective and organ doses from Kerma-Area Product (PKA) values were determined. Methods: 77 cerebral procedures performed with five angiographic equipment installed in three Italian centres were analyzed. Local settings and acquisition protocols were considered. The geometrical, technical and dosimetric data of 16,244 irradiation events (13305 fluoroscopy, 2811 digital subtraction angiography, 128 conebeam CT) were extracted from RDSRs by local dose monitoring systems and were input in MonteCarlo PCXMC software to calculate effective and organ doses. Finally, conversion factors to determine effective and organ doses from PKA were determined. Differences between centres were assessed through statistical analysis and accuracy of dose calculation method based on conversion factors was assessed through Bland-Altman analysi. Results: Large variations in PKA (14-561 Gycm2) and effective dose (1.2-73.5 mSv) were observed due to different degrees of complexity in the procedures and angiographic system technology. The most exposed organs were brain, salivary glands, oral mucosa, thyroid and skeleton. The study highlights the importance of recent technology in reducing patient exposure (about fourfold, even more in DSA). No statistically significant difference was observed in conversion factors between centres, except for some organs. A conversion factor of 0.09 +/- 0.02 mSv/Gycm2 was obtained for effective dose. Conclusions: Organ and effective doses were assessed for neuro-interventional procedures. Conversion factors for calculating effective and organ doses from PKA were provided.
引用
收藏
页数:8
相关论文
共 42 条
  • [31] A study of estimated glomerular filtration rate in patients undergoing diagnostic or interventional coronary contrast procedures
    Khan, Mohammed Shoeb A.
    Kumar, Vinod K.
    Kumar, Anil R.
    Prasannan, Bipi
    Urs, Vishnudev
    Unni, V. N.
    INDIAN JOURNAL OF NEPHROLOGY, 2022, 32 (02) : 132 - 137
  • [32] Patient radiation dose during diagnostic and interventional cardiology procedures A study in a tertiary hospital
    Anim-Sampong, Samuel
    Antwi, William Kwadwo
    Adomako, John Bright
    Botwe, Benard Ohene
    Sarkodie, Benjamin Dabo
    Brakohiapa, Edmund K.
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2023, 54 (02) : 298 - 305
  • [33] VISUAL AND NUMERICAL METHODS TO MEASURE PATIENT SKIN DOSES IN INTERVENTIONAL PROCEDURES USING RADIOCHROMIC XR-RV2 FILMS
    Sanchez, R.
    Vano, E.
    Fernandez, J. M.
    Machado, A.
    Roas, N.
    RADIATION PROTECTION DOSIMETRY, 2011, 147 (1-2) : 94 - 98
  • [34] A COMPARATIVE STUDY OF ORGAN DOSES ASSESSMENT FOR PATIENTS UNDERGOING CONVENTIONAL X-RAY EXAMINATIONS: PHANTOM EXPERIMENTS VS. CALCULATIONS
    Golikov, V.
    Barkovsky, A.
    Wallstrom, E.
    Cederblad, A.
    RADIATION PROTECTION DOSIMETRY, 2018, 178 (02) : 223 - 234
  • [35] Radiation doses from fluoroscopically guided cardiac catheterization procedures in children and young adults in the United Kingdom: a multicentre study
    Harbron, R. W.
    Pearce, M. S.
    Salotti, J. A.
    McHugh, K.
    McLaren, C.
    Abernethy, L.
    Reed, S.
    O'Sullivan, J.
    Chapple, C-L
    BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1048)
  • [36] Conversion factors of effective and equivalent organ doses with the air kerma area product in patients undergoing coronary angiography and percutaneous coronary interventions
    Brambilla, M.
    Cannillo, B.
    Matheoud, R.
    Compagnone, G.
    Rognoni, A.
    Bongo, A. S.
    Carriero, A.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 42 : 189 - 196
  • [37] A Monte Carlo study of organ and effective doses of cone beam computed tomography (CBCT) scans in radiotherapy
    Abuhaimed, Abdullah
    Martin, Colin J.
    Sankaralingam, Marimuthu
    JOURNAL OF RADIOLOGICAL PROTECTION, 2018, 38 (01) : 61 - 80
  • [38] A multicenter study of radiation doses to the eye lenses of medical staff performing non-vascular imaging and interventional radiology procedures in Japan
    Matsubara, Kosuke
    Takei, Yasutaka
    Mori, Hiroshige
    Kobayashi, Ikuo
    Noto, Kimiya
    Igarashi, Takayuki
    Suzuki, Shoichi
    Akahane, Keiichi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 74 : 83 - 91
  • [39] Monte Carlo study of organ doses and related secondary cancer risk estimations for patients undergoing prostate radiotherapy: Algerian population-based study
    Alem-Bezoubiri, Asma
    Bezoubiri, Fethi
    Speiser, Michael
    Suleiman, Suleiman Ameir
    Donya, Hossam
    Chami, Ahmed Chafik
    APPLIED RADIATION AND ISOTOPES, 2025, 216
  • [40] Evaluation of coefficients to derive organ and effective doses from cone-beam CT (CBCT) scans: a Monte Carlo study
    Abuhaimed, Abdullah
    Martin, Colin J.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2018, 38 (01) : 189 - 206