Comprehensive dosimetric commissioning of proton minibeam radiotherapy on a single gantry proton system

被引:0
作者
Lin, Yuting [1 ]
Li, Wangyao [1 ]
Wang, Aoxiang [1 ,2 ]
Johnson, Daniel [1 ]
Gan, Gregory N. [1 ]
Gao, Hao [1 ]
机构
[1] Univ Kansas, Med Ctr, Dept Radiat Oncol, Kansas City, KS 66103 USA
[2] Huazhong Univ Sci & Technol, Dept Biomed Engn, Wuhan, Peoples R China
来源
FRONTIERS IN ONCOLOGY | 2024年 / 14卷
关键词
proton therapy; spatially fractionated radiotherapy; proton mini-beam; small field dosimetry; film dosimetry; proton dosimetry; FRACTIONATED RADIATION-THERAPY;
D O I
10.3389/fonc.2024.1421869
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background Proton minibeam radiation therapy (pMBRT) can deliver spatially fractionated dose distributions with submillimeter resolution. These dose distributions exhibit significant heterogeneity in both depth and lateral directions. Accurate characterization of pMBRT doses requires dosimetry devices with high spatial resolution and a wide dynamic range. Furthermore, the dependency of dosimetric measurements on Linear Energy Transfer (LET), as observed in conventional proton therapy, is also present in pMBRT depth dose measurements.Purpose This work demonstrates the process of performing comprehensive dosimetric measurements to characterize the pMBRT collimator on a clinical single-gantry proton machine, utilizing commercially available dosimetry devices.Methods The minibeam collimator is designed to be mounted on the clinical nozzle as a beam-modifying accessory. Three collimators, each with a slit opening of 0.4 mm, are thoroughly evaluated. The center-to-center (c-t-c) distances of the slits for these collimators are 2.8 mm, 3.2 mm, and 4.0 mm, respectively. High spatial resolution dosimetry devices are essential for PMBRT dose characterizations. To meet this requirement, two-dimensional (2D) dose measurement devices, Gafchromic films, are used to measure lateral profiles at various depths. Films are also used for depth dose profile measurements in solid water. Additionally, high-resolution point dose detectors, microDiamond, and Razor diode detectors are employed for lateral profile measurements at various depths. Percent depth dose (PDD) measurements of pMBRT in solid water, with various proton energies, collimators, and air gaps, are performed using Gafchromic films. The film's LET dependency for proton beams is corrected to ensure accurate pMBRT PDD measurements. The Monte Carlo simulation tool TOPAS is utilized to compare and validate all experimental measurements.Results At depths where LET is not a concern, film dose measurements were consistent with microDiamond and Razor diode point measurements. The point detectors need to be orientated with the thin side aligned to the incoming beam. Comparison of the lateral dose profiles extracted from TOPAS simulations, films, microDiamond, and Razor diode detectors shows a passing rate exceeding 98% in 1D gamma analysis at 3% 0.1 mm criteria.However, when the microDiamond detector is orientated to face the pMBRT beam, its spatial resolution may not be sufficient to capture the peak and valley dose accurately. Nevertheless, an accuracy within 2% can still be achieved when comparing the average dose. The PDD measurements show that the peak valley dose ratio (PVDR) of pMBRT can be altered at different depths with different air gaps using the same collimator or different collimators of different c-t-c distances.Conclusion Our study demonstrates that comprehensive dose measurements for pMBRT can be conducted using standard clinical dose measurement devices. These measurements are indispensable for guiding and ensuring accurate dose reporting in pre-clinical studies using the pMBRT technique.
引用
收藏
页数:12
相关论文
共 39 条
  • [1] Evaluation of the Role of the Immune System Response After Minibeam Radiation Therapy
    Bertho, Annaig
    Iturri, Lorea
    Brisebard, Elise
    Juchaux, Marjorie
    Gilbert, Cristele
    Ortiz, Ramon
    Sebrie, Catherine
    Jourdain, Laurene
    Lamirault, Charlotte
    Ramasamy, Gabriel
    Pouzoulet, Frederic
    Prezado, Yolanda
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2023, 115 (02): : 426 - 439
  • [2] First Evaluation of Temporal and Spatial Fractionation in Proton Minibeam Radiation Therapy of Glioma-Bearing Rats
    Bertho, Annaig
    Ortiz, Ramon
    Juchaux, Marjorie
    Gilbert, Cristele
    Lamirault, Charlotte
    Pouzoulet, Frederic
    Polledo, Laura
    Liens, Alethea
    Warfving, Nils
    Sebrie, Catherine
    Jourdain, Laurene
    Patriarca, Annalisa
    de Marzi, Ludovic
    Prezado, Yolanda
    [J]. CANCERS, 2021, 13 (19)
  • [3] A Current Review of Spatial Fractionation: Back to the Future?
    Billena, Cole
    Khan, Atif J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 104 (01): : 177 - 187
  • [4] Chang S, 2023, MED PHYS, V50, P40, DOI 10.1002/mp.16314
  • [5] Optimization of hexagonal-pattern minibeams for spatially fractionated radiotherapy using proton beam scanning
    Charyyev, Serdar
    Artz, Mark
    Szalkowski, Gregory
    Chang Chih-Wei
    Stanforth, Alexander
    Lin Liyong
    Zhang Rongxiao
    Wang, C. -K. Chris
    [J]. MEDICAL PHYSICS, 2020, 47 (08) : 3485 - 3495
  • [6] Spectroscopic analysis of irradiated radiochromic EBT-XD films in proton and photon beams
    Darafsheh, Arash
    Zhao, Tianyu
    Khan, Rao
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (20)
  • [7] Implementation of planar proton minibeam radiation therapy using a pencil beam scanning system: A proof of concept study
    De Marzi, Ludovic
    Patriarca, Annalisa
    Nauraye, Catherine
    Hierso, Eric
    Dendale, Remi
    Guardiola, Consuelo
    Prezado, Yolanda
    [J]. MEDICAL PHYSICS, 2018, 45 (11) : 5305 - 5316
  • [8] Minibeam Therapy With Protons and Light Ions: Physical Feasibility and Potential to Reduce Radiation Side Effects and to Facilitate Hypofractionation
    Dilmanian, F. Avraham
    Eley, John G.
    Krishnan, Sunil
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 92 (02): : 469 - 474
  • [9] Survival Analysis of F98 Glioma Rat Cells Following Minibeam or Broad-Beam Synchrotron Radiation Therapy
    Gil, Silvia
    Sarun, Sukhena
    Biete, Albert
    Prezado, Yolanda
    Sabes, Manel
    [J]. RADIATION ONCOLOGY, 2011, 6
  • [10] Dosimetric response of Gafchromic™ EBT-XD film to therapeutic protons
    Guan, Fada
    Wang, Xiaochun
    Yang, Ming
    Draeger, Emily
    Han, Dae
    Iga, Kiminori
    Guo, Fanqing
    Perles, Luis
    Li, Yuting
    Sahoo, Narayan
    Mohan, Radhe
    Chen, Zhe
    [J]. PRECISION RADIATION ONCOLOGY, 2023, 7 (01): : 15 - 26