Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices

被引:4
作者
Feng, Yue [1 ]
Nitter, Tore
Bertelsen, Geir [1 ,2 ]
Stojanovic, Aleksandar [2 ,3 ]
机构
[1] Univ Tromso, Inst Community Med, Fac Hlth Sci, Tromso, Norway
[2] Univ Hosp North Norway, Dept Ophthalmol, Sykehusveien 38, N-9019 Tromso, Norway
[3] Univ Tromso, Inst Clin Med, Fac Hlth Sci, Tromso, Norway
关键词
axial length; keratoconus; optical coherence tomography; total corneal astigmatism; PRECISION REPEATABILITY; HEIDELBERG ANTERION; POWER; KERATOMETRY; POSTERIOR; REPRODUCIBILITY; FORMULA;
D O I
10.1111/ceo.14423
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: To evaluate repeatability and agreement in measurements of total corneal astigmatism (TCA) in keratoconic eyes, using four optical coherence tomography (OCT)-based devices: Anterion, Casia SS-1000, IOLMaster 700, and MS-39. Methods: Three consecutive measurements were taken with each device in 136 eyes. TCA values were converted into components J(0) and J(45). The Anterion and the IOLMaster 700 also provided axial length (AL) measurements. The repeatability was calculated using pooled within-subject standard deviation (S-w). The agreement among the four devices was assessed by pairwise comparisons and Bland-Altman plots. Results: For all devices, the repeatability of TCA measurements showed S-w <= 0.23 D for TCA magnitude, <= 0.14 D for J(0), and <= 0.12 D for J(45). There were statistically significant differences in TCA magnitude for each pair, except for IOLMaster 700 with MS-39, and Anterion with MS-39. The repeatability (S-w) of axis measurements had a statistically significant negative correlation with the TCA magnitude (p < 0.001 for all devices). Both Anterion and IOLMaster 700 had high repeatability in AL measurements (S-w: 0.007 mm for Anterion and 0.009 mm for IOLMaster 700). The difference in AL between the two was 0.015 +/- 0.033 mm (p < 0.001). Conclusions: All four devices showed good repeatability in TCA measurements in keratoconic eyes, the agreement for TCA measurements between the tested devices was generally low. Anterion and IOLMaster 700 showed good repeatability and agreement in AL measurements.
引用
收藏
页码:800 / 810
页数:11
相关论文
共 54 条
  • [41] An Analysis of the Factors Influencing the Residual Refractive Astigmatism After Cataract Surgery With Toric Intraocular Lenses
    Savini, Giacomo
    Naeser, Kristian
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (02) : 827 - 835
  • [42] Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices
    Schiano-Lomoriello, Domenico
    Hoffer, Kenneth J.
    Abicca, Irene
    Savini, Giacomo
    [J]. SCIENTIFIC REPORTS, 2021, 11 (01)
  • [43] Repeatability of anterior segment measurements by optical coherence tomography combined with Placido disk corneal topography in eyes with keratoconus
    Schiano-Lomoriello, Domenico
    Bono, Valeria
    Abicca, Irene
    Savini, Giacomo
    [J]. SCIENTIFIC REPORTS, 2020, 10 (01)
  • [44] Evaluation of total corneal power measurements with a new optical biometer
    Shajari, Mehdi
    Sonntag, Ruven
    Ramsauer, Michaela
    Kreutzer, Thomas
    Vounotrypidis, Efstathios
    Kohnen, Thomas
    Priglinger, Siegfried
    Mayer, Wolfgang J.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2020, 46 (05) : 675 - 681
  • [45] Comparison of biometry measurements and intraocular lens power prediction between 2 SS-OCT-based biometers
    Sorkin, Nir
    Zadok, Tal
    Barrett, Graham D.
    Chasid, Otzem
    Abulafia, Adi
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2023, 49 (05) : 460 - 466
  • [46] Artificial stiffening of the cornea by induction of intrastromal cross-links
    Sporl, E
    Huhle, M
    Kasper, M
    Seiler, T
    [J]. OPHTHALMOLOGE, 1997, 94 (12): : 902 - 906
  • [47] An in depth analysis of histopathological characteristics found in keratoconus
    Sykakis, E.
    Carley, F.
    Irion, L.
    Denton, J.
    Hillarby, M. C.
    [J]. PATHOLOGY, 2012, 44 (03) : 234 - 239
  • [48] Repeatability of whole-cornea measurements using a new swept-source optical coherence tomographer
    Tana-Rivero, Pedro
    Aguilar-Corcoles, Salvador
    Ruiz-Mesa, Ramon
    Montes-Mico, Robert
    [J]. EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2021, 31 (04) : 1709 - 1719
  • [49] Power vector analysis of the optical outcome of refractive surgery
    Thibos, LN
    Horner, D
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (01) : 80 - 85
  • [50] Corneal Epithelial Thickness Intrasubject Repeatability and its Relation With Visual Limitation in Keratoconus
    Vega-Estrada, Alfredo
    Mimouni, Michael
    Espla, Elias
    Alto del Barrio, Jorge
    Alio, Jorge L.
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2019, 200 : 255 - 262