Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices

被引:4
作者
Feng, Yue [1 ]
Nitter, Tore
Bertelsen, Geir [1 ,2 ]
Stojanovic, Aleksandar [2 ,3 ]
机构
[1] Univ Tromso, Inst Community Med, Fac Hlth Sci, Tromso, Norway
[2] Univ Hosp North Norway, Dept Ophthalmol, Sykehusveien 38, N-9019 Tromso, Norway
[3] Univ Tromso, Inst Clin Med, Fac Hlth Sci, Tromso, Norway
关键词
axial length; keratoconus; optical coherence tomography; total corneal astigmatism; PRECISION REPEATABILITY; HEIDELBERG ANTERION; POWER; KERATOMETRY; POSTERIOR; REPRODUCIBILITY; FORMULA;
D O I
10.1111/ceo.14423
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: To evaluate repeatability and agreement in measurements of total corneal astigmatism (TCA) in keratoconic eyes, using four optical coherence tomography (OCT)-based devices: Anterion, Casia SS-1000, IOLMaster 700, and MS-39. Methods: Three consecutive measurements were taken with each device in 136 eyes. TCA values were converted into components J(0) and J(45). The Anterion and the IOLMaster 700 also provided axial length (AL) measurements. The repeatability was calculated using pooled within-subject standard deviation (S-w). The agreement among the four devices was assessed by pairwise comparisons and Bland-Altman plots. Results: For all devices, the repeatability of TCA measurements showed S-w <= 0.23 D for TCA magnitude, <= 0.14 D for J(0), and <= 0.12 D for J(45). There were statistically significant differences in TCA magnitude for each pair, except for IOLMaster 700 with MS-39, and Anterion with MS-39. The repeatability (S-w) of axis measurements had a statistically significant negative correlation with the TCA magnitude (p < 0.001 for all devices). Both Anterion and IOLMaster 700 had high repeatability in AL measurements (S-w: 0.007 mm for Anterion and 0.009 mm for IOLMaster 700). The difference in AL between the two was 0.015 +/- 0.033 mm (p < 0.001). Conclusions: All four devices showed good repeatability in TCA measurements in keratoconic eyes, the agreement for TCA measurements between the tested devices was generally low. Anterion and IOLMaster 700 showed good repeatability and agreement in AL measurements.
引用
收藏
页码:800 / 810
页数:11
相关论文
共 54 条
  • [1] New regression formula for tonic intraocular lens calculations
    Abulafia, Adi
    Koch, Douglas D.
    Wang, Li
    Hill, Warren E.
    Assia, Ehud I.
    Franchina, Maria
    Barrett, Graham D.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (05) : 663 - 671
  • [2] New method of quantifying corneal topographic astigmatism that corresponds with manifest refractive cylinder
    Alpins, Noel
    Ong, James K. Y.
    Stamatelatos, George
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (11) : 1978 - 1988
  • [3] *KERATOCONE CLASSIQUE ET KERATOCONE FRUSTE - ARGUMENTS UNITAIRES
    AMSLER, M
    [J]. OPHTHALMOLOGICA, 1946, 111 (2-3) : 96 - 101
  • [4] BIOMECHANICAL PROPERTIES OF KERATOCONUS AND NORMAL CORNEAS
    ANDREASSEN, TT
    SIMONSEN, AH
    OXLUND, H
    [J]. EXPERIMENTAL EYE RESEARCH, 1980, 31 (04) : 435 - 441
  • [5] Corneal and Whole-Eye Higher Order Aberrations Do Not Correlate With Ocular Residual Astigmatism in Prepresbyopic Refractive Surgery Candidates
    Balparda, Kepa
    Maya-Naranjo, Maria Isabel
    Mesa-Mesa, Sara
    Herrera-Chalarca, Tatiana
    [J]. CORNEA, 2023, 42 (07) : 867 - 873
  • [6] Bland JM, 1996, BRIT MED J, V313, P744
  • [7] Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers
    Cheng, Shi-Ming
    Zhang, Jia-Sheng
    Shao, Xu
    Wu, Ze-Tong
    Li, Tian-Tian
    Wang, Po
    Lin, Jun-Hai
    Yu, A-Yong
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2022, 260 (07) : 2271 - 2281
  • [8] Correcting keratoconus with intracorneal rings
    Colin, J
    Cochener, B
    Savary, G
    Malet, F
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (08) : 1117 - 1122
  • [9] Multicentre study: reliability and repeatability of Scheimpflug system measurement in keratoconus
    de Luis Eguileor, Beatriz
    Arriola-Villalobos, Pedro
    Pijoan Zubizarreta, Jose Ignacio
    Feijoo Lera, Raquel
    Santamaria Carro, Alaitz
    Diaz-Valle, David
    Etxebarria, Jaime
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2021, 105 (01) : 22 - 26
  • [10] Evaluation of the Reliability and Repeatability of Scheimpflug System Measurement in Keratoconus
    de Luis Eguileor, Beatriz
    Escudero Argaluza, Julene
    Pijoan Zubizarreta, Jose Ignacio
    Santamaria Carro, Alaitz
    Etxebarria Ecenarro, Jaime
    [J]. CORNEA, 2018, 37 (02) : 177 - 181