An Interlaboratory Comparison of Sizing and Counting of Subvisible Particles Mimicking Protein Aggregates

被引:33
作者
Ripple, Dean C. [1 ]
Montgomery, Christopher B. [1 ]
Hu, Zhishang [2 ]
机构
[1] NIST, Biomol Measurement Div, Gaithersburg, MD 20899 USA
[2] Chinese Acad Sci, Inst Biophys, Ctr Computat & Syst Biol, Beijing 100080, Peoples R China
关键词
image analysis; imaging methods; light scattering; microparticles; particle size; physical characterization; FILLING PUMP OPERATION; REFRACTIVE-INDEX; NANOPARTICLES;
D O I
10.1002/jps.24287
中图分类号
R914 [药物化学];
学科分类号
100701 ;
摘要
Accurate counting and sizing of protein particles has been limited by discrepancies of counts obtained by different methods. To understand the bias and repeatability of techniques in common use in the biopharmaceutical community, the National Institute of Standards and Technology has conducted an interlaboratory comparison for sizing and counting subvisible particles from 1 to 25 m. Twenty-three laboratories from industry, government, and academic institutions participated. The circulated samples consisted of a polydisperse suspension of abraded ethylene tetrafluoroethylene particles, which closely mimic the optical contrast and morphology of protein particles. For restricted data sets, agreement between data sets was reasonably good: relative standard deviations (RSDs) of approximately 25% for light obscuration counts with lower diameter limits from 1 to 5 m, and approximately 30% for flow imaging with specified manufacturer and instrument setting. RSDs of the reported counts for unrestricted data sets were approximately 50% for both light obscuration and flow imaging. Differences between instrument manufacturers were not statistically significant for light obscuration but were significant for flow imaging. We also report a method for accounting for differences in the reported diameter for flow imaging and electrical sensing zone techniques; the method worked well for diameters greater than 15 m. (c) 2014 Wiley Periodicals, Inc. and the American Pharmacists Association J Pharm Sci 104:666-677, 2015
引用
收藏
页码:666 / 677
页数:12
相关论文
共 30 条
[1]   Critical evaluation and guidance for using the coulter method for counting subvisible particles in protein solutions [J].
Barnard, James G. ;
Rhyner, Matthew N. ;
Carpenter, John F. .
JOURNAL OF PHARMACEUTICAL SCIENCES, 2012, 101 (01) :140-153
[2]   Effects of Surfaces and Leachables on the Stability of Biopharmaceuticals [J].
Bee, Jared S. ;
Randolph, Theodore W. ;
Carpenter, John F. ;
Bishop, Steven M. ;
Dimitrova, Mariana N. .
JOURNAL OF PHARMACEUTICAL SCIENCES, 2011, 100 (10) :4158-4170
[3]   Monoclonal Antibody Interactions With Micro- and Nanoparticles: Adsorption, Aggregation, and Accelerated Stress Studies [J].
Bee, Jared S. ;
Chiu, David ;
Sawicki, Suzanne ;
Stevenson, Jennifer L. ;
Chatterjee, Koustuv ;
Freund, Erwin ;
Carpenter, John F. ;
Randolph, Theodore W. .
JOURNAL OF PHARMACEUTICAL SCIENCES, 2009, 98 (09) :3218-3238
[4]   Quantitative phase amplitude microscopy IV: imaging thick specimens [J].
Bellair, CJ ;
Curl, CL ;
Allman, BE ;
Harris, PJ ;
Roberts, A ;
Delbridge, LMD ;
Nugent, KA .
JOURNAL OF MICROSCOPY, 2004, 214 (01) :62-69
[5]   Weighing of biomolecules, single cells and single nanoparticles in fluid [J].
Burg, Thomas P. ;
Godin, Michel ;
Knudsen, Scott M. ;
Shen, Wenjiang ;
Carlson, Greg ;
Foster, John S. ;
Babcock, Ken ;
Manalis, Scott R. .
NATURE, 2007, 446 (7139) :1066-1069
[6]   Overlooking Subvisible Particles in Therapeutic Protein Products: Gaps That May Compromise Product Quality [J].
Carpenter, John F. ;
Randolph, Theodore W. ;
Jiskoot, Wim ;
Crommelin, Daan J. A. ;
Middaugh, C. Russell ;
Winter, Gerhard ;
Fan, Ying-Xin ;
Kirshner, Susan ;
Verthelyi, Daniela ;
Kozlowski, Steven ;
Clouse, Kathleen A. ;
Swann, Patrick G. ;
Rosenberg, Amy ;
Cherney, Barry .
JOURNAL OF PHARMACEUTICAL SCIENCES, 2009, 98 (04) :1201-1205
[7]  
Chaturvedi V., 2010, Journal of Cell and Molecular Biology, V8, P103
[8]   Characterization of Particles in Protein Solutions: Reaching the Limits of Current Technologies [J].
Demeule, Barthelemy ;
Messick, Steven ;
Shire, Steven J. ;
Liu, Jun .
AAPS JOURNAL, 2010, 12 (04) :708-715
[9]  
DuPontTMTefzel® fluoropolymer resin, PROPERTIES HDB
[10]   Refractive index determination of transparent samples by noniterative phase retrieval [J].
Frank, Johannes ;
Matrisch, Jan ;
Horstmann, Jens ;
Altmeyer, Stefan ;
Wernicke, Guenther .
APPLIED OPTICS, 2011, 50 (04) :427-433