Comparison of two different respiratory monitoring systems with 4D-CT images for target volume definition in patients undergoing para-aortic nodal irradiation

被引:1
作者
Ergen, Sefika Arzu [1 ]
Karacam, Songul [1 ]
Catal, Tuba Kurt [1 ]
Dincbas, Fazilet Oner [1 ]
Oksuz, Didem Colpan [1 ]
Sahinler, Ismet [1 ]
机构
[1] Istanbul Univ Cerrahpasa, Cerrahpasa Fac Med, Dept Radiat Oncol, Istanbul, Turkiye
关键词
Four-dimensional computed tomography; respiratory monitoring system; RPM; smart deviceless 4D; 4D CT ACQUISITION; MOTION; ARTIFACTS; SCANS; RADIOTHERAPY; MODEL;
D O I
10.14744/nci.2023.06856
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE: Today, respiratory movement can be monitored and recorded with different methods during a simulation on a four-dimensional (4D) computed tomography (CT) device to be used in radiotherapy planning. A synchronized respiratory monitoring system (RPM) with an externally equipped device is one of these methods. Another method is to create 4D images of the patient's breathing phases without the need for extra equipment, with an anatomy-based software program integrated into the CT device. Our aim is to compare the RPM system and the software system (Deviceless) which are two different respiratory monitoring methods used in tracking moving targets during 4D-CT imaging and to assess their clinical usability. METHODS: Ten patients who underwent paraaortic nodal irradiation were enrolled. The simulation was performed using intravenous contrast material on a 4D-CT device with both respiratory monitoring methods. The right/left kidneys and renal arteries were chosen as references to evaluate abdominal organ movement. It was then manually contoured one by one on both sets of images. The images were compared volumetrically and geometrically after rigid reconstruction. The similarity between the contours was determined by the Dice index. Wilcoxon test was used for statistical comparisons. RESULTS: The motion of the kidneys in all three directions was found to be 0.0 cm in both methods. The shifts in the right/ left renal arteries were submillimetric. The Dice index showed a high similarity in both kidney and renal artery contours. CONCLUSION: In our study, no difference was found between RPM and Deviceless systems used for tracking and detection of moving targets during simulation in 4D-CT. Both methods can be used safely for radiotherapy planning according to the available possibilities in the clinic.
引用
收藏
页码:120 / 126
页数:7
相关论文
共 27 条
[1]   Effects of motion on the total dose distribution [J].
Bortfeld, T ;
Jiang, SB ;
Rietzel, E .
SEMINARS IN RADIATION ONCOLOGY, 2004, 14 (01) :41-51
[2]   Artifacts in computed tomography scanning of moving objects [J].
Chen, GTY ;
Kung, JH ;
Beaudette, KP .
SEMINARS IN RADIATION ONCOLOGY, 2004, 14 (01) :19-26
[3]   Displacement-based binning of time-dependent computed tomography image data sets [J].
Fitzpatrick, MJ ;
Starkschall, G ;
Antolak, JA ;
Fu, J ;
Shukla, H ;
Keall, PJ ;
Klahr, P ;
Mohan, R .
MEDICAL PHYSICS, 2006, 33 (01) :235-246
[4]   Respiratory gating for radiotherapy: main technical aspects and clinical benefits [J].
Giraud, P. ;
Garcia, R. .
BULLETIN DU CANCER, 2010, 97 (07) :847-856
[5]   A comparison of dosimetric variance for external-beam partial breast irradiation using three-dimensional and four-dimensional computed tomography [J].
Guo, Bing ;
Li, Jian-Bin ;
Wang, Wei ;
Xu, Min ;
Li, Yan-Kang ;
Liu, Tong-Hai .
ONCOTARGETS AND THERAPY, 2016, 9 :1857-1863
[6]   Technical evaluation of different respiratory monitoring systems used for 4D CT acquisition under free breathing [J].
Heinz, Christian ;
Reiner, Michael ;
Belka, Claus ;
Walter, Franziska ;
Soehn, Matthias .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (02) :334-349
[7]   Correlation of lung tumor motion with external surrogate indicators of respiration [J].
Hoisak, JDP ;
Sixel, KE ;
Tirona, R ;
Cheung, PCF ;
Pignol, JP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (04) :1298-1306
[8]   Investigation of Internal Target Volumes Using Device and Deviceless Four-dimensional Respiratory Monitoring Systems for Moving Targets in Four-dimensional Computed Tomography Acquisition [J].
Holla, Raghavendra ;
Khanna, D. ;
Barsing, Shubhangi ;
Pillai, Bhaskaran K. ;
Ganesh, Tharmarnadar .
JOURNAL OF MEDICAL PHYSICS, 2019, 44 (02) :77-83
[9]   Respiratory regularity gated 4D CT acquisition: concepts and proof of principle [J].
Keall, P. J. ;
Vedam, S. S. ;
George, R. ;
Williamson, J. F. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2007, 30 (03) :211-220
[10]   The management of respiratory motion in radiation oncology report of AAPM Task Group 76 [J].
Keall, Paul J. ;
Mageras, Gig S. ;
Balter, James M. ;
Emery, Richard S. ;
Forster, Kenneth M. ;
Jiang, Steve B. ;
Kapatoes, Jeffrey M. ;
Low, Daniel A. ;
Murphy, Martin J. ;
Murray, Brad R. ;
Ramsey, Chester R. ;
Van Herk, Marcel B. ;
Vedam, S. Sastry ;
Wong, John W. ;
Yorke, Ellen .
MEDICAL PHYSICS, 2006, 33 (10) :3874-3900