Comparison between Air-Q Self Pressurized Airway Device with Blocher and Proseal Laryngeal Mash Airway in anesthetized paralyzed adult female patients undergoing elective gynecological operations

被引:0
作者
Youssef, Maha Mohammed Ismail [1 ]
Dobal, Naser Mohammed [1 ]
Hammad, Yahya Mohamed [1 ]
El-Refai, Nesrine Abdel Rahman [1 ]
Abdelrahman, Reham Ali Abdelhaleem [1 ]
机构
[1] Cairo Univ, Fac Med, Dept Anesthesia Surg ICU & Pain Management, 71th Kasr Al Ainy St, Cairo 11562, Egypt
关键词
oropharyngeal leak pressure; fiber-optic view score of glottis; laryngo- pharyngeal morbidity; leak fraction; supraglottic airway devices; OROPHARYNGEAL LEAK PRESSURE; FIBEROPTIC SCORING SYSTEM; TRACHEAL INTUBATION; RANDOMIZED-TRIAL; NECK POSITION; CUFF POSITION; LMA; AIR-Q(R); VENTILATION; SUPREME;
D O I
暂无
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background:The The Air-Q Self Pressurized Airway Device with Blocker (SP Blocker) was compared to the Proseal Laryngeal Mask Airway (PLMA) during positive pressure ventilation regarding the primary outcome (oropharyngeal leak pressure [OLP]), secondary outcomes (peak inspiratory pressure [PIP], inspired tidal volume [ITV], expired tidal volume [ETV], leak volume [LV] and leak fraction [LF]), insertion time, ventilation score, fiber-optic glottis view score, and postoperative laryngopharyngeal parameters (LPM). Methods: Adult healthy female patients scheduled for elective gynecological laparotomies under general anesthesia using controlled mechanical ventilation were recruited to a prospective randomized comparative clinical trial. Exclusion criteria were body mass index (BMI) >= 35 kg m(-2), , El-Ganzouri score >= 5, upper airway problems, hiatus hernia or pregnancy. Patients were classified into an SP Blocker group (n n = 75) and a PLMA group (n n = 75). Primary and secondary outcomes were assessed initially and at fixed time points after successful insertion of devices. Results: Initially after successful device insertion: the SP Blocker group showed statistically significant higher mean OLP (cmH(2)O) ) (29.46 +/- 2.11 vs. 28.06 +/- 1.83 respectively; 95% CI: -2.037 to -0.76, P < 0.0001), lower mean PIP (cmH(2)O)) (15.49 +/- 0.61 vs. 17.78 +/- 1.04 respectively; 95% CI: 2.02 to 2.56, P < 0.0001), higher mean ITV (mL) (411 +/- 30 vs. 403 +/- 15 respectively; 95% CI: -15.65 to -0.347, P = 0.041), higher mean ETV (mL) (389 +/- 12 vs. 354 +/- 11 respectively; 95% CI: -38.72 to -31.29, P < 0.0001), lower mean LV (mL) (22 +/- 18 vs. 49 +/- 10 respectively; 95% CI: 22.3 to 31.7, P < 0.0001) and lower mean LF (%) (5 +/- 2.04 vs. 12 +/- 6.8 respectively; 95% CI: 5.38 to 8.62, P < 0.0001) than the PLMA group. Mean insertion time (seconds) was shorter in the SP Blocker group than the PLMA group (16.39 +/- 2.81 vs. 18.63 +/- 3.44 respectively; 95% CI: 1.23 to 3.25, P < 0.0001). The SP Blocker group offered a better fiber-optic glottis view score than the PLMA group without differences concerning ventilation score and LPM. Conclusions: SP Blocker provided as safe anesthesia during controlled mechanical ventilation as PLMA.
引用
收藏
页码:108 / 120
页数:13
相关论文
共 56 条
[41]  
Moorthy PVC, 2019, Indian J Clin Anaesth, V6, P349, DOI [10.18231/j.ijca.2019.067, DOI 10.18231/J.IJCA.2019.067]
[42]  
Moser B, 2017, CAN J ANESTH, V64, P1119, DOI 10.1007/s12630-017-0936-7
[43]  
Nishimoto Kenta, 2014, Masui, V63, P1125
[44]   Assessing ProSeal laryngeal mask positioning: The suprasternal notch test [J].
O'Connor, CJ ;
Borromeo, CJ ;
Stix, MS .
ANESTHESIA AND ANALGESIA, 2002, 94 (05) :1374-1375
[45]  
Ovasapian A, 2002, Anesthesiology, V96, pA, DOI [10.1097/00000542-200209002-01321, DOI 10.1097/00000542-200209002-01321]
[46]   The Influence of Head and Neck Position on the Oropharyngeal Leak Pressure and Cuff Position of Three Supraglottic Airway Devices [J].
Park, Sang-Hyun ;
Han, Sung-Hee ;
Do, Sang-Hwan ;
Kim, Jung-Won ;
Kim, Jin-Hee .
ANESTHESIA AND ANALGESIA, 2009, 108 (01) :112-117
[47]   Comparative evaluation of self-pressurized Air-Q® and Proseal™ LMA® in patients undergoing elective surgery under general anaesthesia: A randomized clinical trial [J].
Rana, Savita ;
Anand, Lakesh Kumar ;
Singh, Manpreet ;
Kapoor, Dheeraj ;
Gupta, Deepika ;
Kaur, Harpreet .
JOURNAL OF ANAESTHESIOLOGY CLINICAL PHARMACOLOGY, 2024, 40 (02) :336-343
[48]   The influence of head and neck position on ventilation with the i-gel airway in paralysed, anaesthetised patients [J].
Sanuki, Takuro ;
Uda, Rumiko ;
Sugioka, Shingo ;
Daigo, Erina ;
Son, Hiroki ;
Akatsuka, Masafumi ;
Kotani, Junichiro .
EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2011, 28 (08) :597-599
[49]   Distance from the glottis to the grille: the LMA Unique, Air-Q and CobraPLA as intubation conduits [J].
Schebesta, Karl ;
Karanovic, Gordana ;
Krafft, Peter ;
Roessler, Bernhard ;
Kimberger, Oliver .
EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2014, 31 (03) :159-165
[50]   A comparison of metal introducer and bougie-guided techniques of insertion PLMA (TM) with respect to cuff position and air leak [J].
Sharma, Deepak ;
Ahmad, Bilal ;
Tiwary, Vaibhav ;
Malhotra, M. M. K. ;
Agarwal, Divya ;
Kaur, Sukhreet .
INDIAN ANAESTHETISTS FORUM, 2018, 19 (01) :15-21