Safety and effectiveness of very-high-power, short-duration ablation in patients with atrial fibrillation: Preliminary results

被引:0
作者
Mitrzak, Karolina [1 ]
Peller, Michal [1 ]
Krzowski, Bartosz [1 ]
Maciejewski, Cezary [1 ]
Balsam, Pawel [1 ]
Marchel, Michal [1 ]
Grabowski, Marcin [1 ]
Lodzinski, Piotr [1 ]
机构
[1] Med Univ Warsaw, Dept Cardiol 1, Ul Banacha 1A, PL-02097 Warsaw, Poland
关键词
ablation; pulmonary vein isolation; atrial fibrillation; vHPSD; QDot Micro (TM); PULMONARY VEIN ISOLATION; RADIOFREQUENCY ABLATION; ANTIARRHYTHMIC-DRUGS; CATHETER ABLATION; CONTACT FORCE; RECURRENCE;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Pulmonary vein isolation (PVI) is at the forefront of rhythm control strategies in patients with atrial fibrillation (AF). A very-high-power, short-duration (vHPSD) catheter, QDot Micro (TM) (Bio-- sense Webster) was designed to improve the effectiveness of AF ablation within a shorter procedure time. The aim of this study was to compare the effectiveness and safety of PVI ablation between this vHPSD ablation mode and conventional ablation-index-guided ablation (ThermoCool Smarttouch SF catheter). Methods: This single-center, retrospective, observational study enrolled 108 patients with AF, referred for catheter ablation between December 16, 2019 and December 3, 2021. In 54 procedures (mean age: 58.0 +/- 12.3; 66.67% male), a QDot Micro(TM )catheter was used (vHPSD-group), and 54 patients (mean age: 57.2 +/- 11.8; 70.37% male) were treated with a ThermoCool Smarttouch (TM) SF catheter (AI-group). The primary endpoint was freedom from AF 3 months after ablation. Results: Atrial fibrillation was found to recur in 14.81% of patients in the vHPSD-group and in 31.48% of patients in the AI-group (p = 0.07). There was no difference in treatment-emergent adverse events between the two groups (6.3% vs. 0%; p = 0.10). One severe adverse event (a cerebral vascular accident) was observed in the vHPSD-group. The mean dose of remifentanil was reported to be lower during QDot Micro (TM) catheter-based PVI (p < 0.001). The vHPSD-based PVI was associated with shorter radiofrequency application time (p < 0.001), fluoroscopy time (p < 0.001), and total procedure time (p < 0.001). Conclusions: This study suggests vHPSD ablation is safe, can reduce the dosage of analgesics during significantly shorter procedures and may enhance the success rate of catheter-based PVI.
引用
收藏
页码:603 / 611
页数:145
相关论文
共 43 条
[31]   High-power short duration vs. conventional radiofrequency ablation of atrial fibrillation: a systematic review and meta-analysis [J].
Ravi, Venkatesh ;
Poudyal, Abhushan ;
Abid, Qurrat-Ul-Ain ;
Larsen, Timothy ;
Krishnan, Kousik ;
Sharma, Parikshit S. ;
Trohman, Richard G. ;
Huang, Henry D. .
EUROPACE, 2021, 23 (05) :710-721
[32]   Pulmonary Vein Isolation With Very High Power, Short Duration, Temperature-Controlled Lesions The QDOT-FAST Trial [J].
Reddy, Vivek Y. ;
Grimaldi, Massimo ;
De Potter, Tom ;
Vijgen, Johan M. ;
Bulava, Alan ;
Duytschaever, Mattias Francis ;
Martinek, Martin ;
Natale, Andrea ;
Knecht, Sebastien ;
Neuzil, Petr ;
Puererfellner, Helmut .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2019, 5 (07) :778-786
[33]   Randomized, Controlled Trial of the Safety and Effectiveness of a Contact Force-Sensing Irrigated Catheter for Ablation of Paroxysmal Atrial Fibrillation Results of the TactiCath Contact Force Ablation Catheter Study for Atrial Fibrillation (TOCCASTAR) Study [J].
Reddy, Vivek Y. ;
Dukkipati, Srinivas R. ;
Neuzil, Petr ;
Natale, Andrea ;
Albenque, Jean-Paul ;
Kautzner, Josef ;
Shah, Dipen ;
Michaud, Gregory ;
Wharton, Marcus ;
Harari, David ;
Mahapatra, Srijoy ;
Lambert, Hendrik ;
Mansour, Moussa .
CIRCULATION, 2015, 132 (10) :907-915
[34]   Safety and efficacy of delivering high-power short-duration radiofrequency ablation lesions utilizing a novel temperature sensing technology [J].
Rozen, Guy ;
Ptaszek, Leon M. ;
Zilberman, Israel ;
Douglas, Victoria ;
Heist, E. Kevin ;
Beeckler, Christopher ;
Altmann, Andres ;
Ruskin, Jeremy N. ;
Govari, Assaf ;
Mansour, Moussa .
EUROPACE, 2018, 20 :F444-F450
[35]   Temperature- and flow-controlled ablation/very-high-power short-duration ablation vs conventional power-controlled ablation: Comparison of focal and linear lesion characteristics [J].
Takigawa, Masateru ;
Kitamura, Takeshi ;
Martin, Claire A. ;
Fuimaono, Kristine ;
Datta, Keshava ;
Joshi, Helee ;
Constantin, Marion ;
Bourier, Felix ;
Cheniti, Ghassen ;
Duchateau, Josselin ;
Pambrun, Thomas ;
Denis, Arnaud ;
Derval, Nicolas ;
Sacher, Frederic ;
Cochet, Hubert ;
Hocini, Meleze ;
Haissaguerre, Michel ;
Jais, Pierre .
HEART RHYTHM, 2021, 18 (04) :553-561
[36]   Quality of life in patients with atrial fibrillation: A systematic review [J].
Thrall, G ;
Lane, D ;
Carroll, D ;
Lip, GYH .
AMERICAN JOURNAL OF MEDICINE, 2006, 119 (05)
[37]   Very high-power short-duration temperature-controlled ablation versus conventional power-controlled ablation for pulmonary vein isolation: The fast and furious - AF study [J].
Tilz, Roland Richard ;
Sano, Makoto ;
Vogler, Julia ;
Fink, Thomas ;
Saraei, Roza ;
Sciacca, Vanessa ;
Kirstein, Bettina ;
Phan, Huong-Lan ;
Hatahet, Sascha ;
Lopez, Lisbeth Delgado ;
Traub, Anna ;
Eitel, Charlotte ;
Schluter, Michael ;
Kuck, Karl-Heinz ;
Heeger, Christian-Hendrik .
IJC HEART & VASCULATURE, 2021, 35
[38]   Recurrence after pulmonary vein isolation is associated with low contact force [J].
Tofig, Bawer Jalal ;
Lukac, Peter ;
Borregaard, Rune ;
Nielsen, Jan Moller ;
Jensen, Henrik Kjaerulf ;
Gerdes, Christian ;
Kristensen, Jens ;
Nielsen, Jens Cosedis ;
Kristiansen, Steen Buus .
SCANDINAVIAN CARDIOVASCULAR JOURNAL, 2018, 52 (01) :28-33
[39]   Assessment of Catheter Ablation or Antiarrhythmic Drugs for First-line Therapy of Atrial Fibrillation A Meta-analysis of Randomized Clinical Trials [J].
Turagam, Mohit K. ;
Musikantow, Daniel ;
Whang, William ;
Koruth, Jacob S. ;
Miller, Marc A. ;
Langan, Marie-Noelle ;
Sofi, Aamir ;
Choudry, Subbarao ;
Dukkipati, Srinivas R. ;
Reddy, Vivek Y. .
JAMA CARDIOLOGY, 2021, 6 (06) :697-705
[40]   Biosense Webster's QDOT Micro™ radiofrequency ablation catheter [J].
Wielandts, Jean-Yves ;
Almorad, Alexandre ;
Hilfiker, Gabriela ;
Gillis, Kris ;
El Haddad, Milad ;
Vijgen, Johan ;
Berte, Benjamin ;
de Waroux, Jean-Benoit Le Polain ;
Tavernier, Rene ;
Duytschaever, Mattias ;
Knecht, Sebastien .