Intravascular guide wire aspiration in a patient on extracorporeal membrane oxygenation: A case report

被引:0
作者
Choi, Jae Ho [1 ]
Yoo, Jae Hwa [2 ]
Chang, Won Ho [1 ]
Park, Young Woo [1 ]
Kim, Hyun Jo [1 ]
Oh, Hong Chul [1 ]
机构
[1] Soonchunhyang Univ, Seoul Hosp, Dept Thorac & Cardiovasc Surg, 59 Daesagwan Gil, Seoul 04401, South Korea
[2] Soonchunhyang Univ, Hosp Seoul, Dept Anesthesiol & Pain Med, Seoul, South Korea
关键词
Case report; extracorporeal membrane oxygenation; guide wire;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Rationale:Guide wire aspiration during central venous catheter (CVC) insertion in a patient on extracorporeal membrane oxygenation (ECMO) is a very rare but dangerous complication. A guide wire aspirated inside the ECMO can cause thrombosis, the ECMO to break down or shut off, and unnecessary ECMO replacement.Patient concerns:A 58-year-old man was scheduled for venovenous ECMO for acute respiratory distress syndrome. After his vital signs stabilized, we inserted a CVC. During CVC insertion, the guide wire was aspirated into the ECMO venous line.Intervention:After confirming the guide wire inside the ECMO venous line, we replaced the entire ECMO circuit.Outcomes:ECMO was maintained for 57 days, and weaning was successful but the patient died 5 days afterward.Lessons:Care must be taken when inserting a CVC using a guide wire in ECMO patients: the guide wire should not be inserted deeply, it should be secured during insertion, the ECMO venous cannula tip requires proper positioning, and ECMO flow should be temporarily reduced.
引用
收藏
页数:3
相关论文
共 7 条
  • [1] Alves DR, 2018, REV BRAS ANESTESIOL, V68, P183, DOI [10.1016/j.bjane.2015.03.013, 10.1016/j.bjan.2016.12.004]
  • [2] How much guidewire is too much? Direct measurement of the distance from subclavian and internal jugular vein access sites to the superior vena cava-atrial junction during central venous catheter placement
    Andrews, RT
    Bova, DA
    Venbrux, AC
    [J]. CRITICAL CARE MEDICINE, 2000, 28 (01) : 138 - 142
  • [3] The human error:: delayed diagnosis, of intravascular loss of guidewires for central venous catheterization
    Auweiler, M
    Kampe, S
    Zähringer, M
    Buzello, S
    von Spiegel, T
    Buzello, W
    Hekmat, K
    [J]. JOURNAL OF CLINICAL ANESTHESIA, 2005, 17 (07) : 562 - 564
  • [4] Khasawneh Faisal A, 2011, Case Rep Crit Care, V2011, P287261, DOI 10.1155/2011/287261
  • [5] Central venous catheter placement during extracorporeal membrane oxygenation therapy
    Picard, Lucile
    Cherait, Chamsedine
    Constant, Ophelie
    Boulanger, Boris
    Dorget, Amandine
    Vodovar, Dominique
    Fiore, Antonio
    Haouache, Hakim
    Dhonneur, Gilles
    Mongardon, Nicolas
    [J]. ANAESTHESIA CRITICAL CARE & PAIN MEDICINE, 2018, 37 (03) : 269 - 270
  • [6] Missed Central Venous Guide Wires: A Systematic Analysis of Published Case Reports
    Pokharel, Krishna
    Biswas, Kumar
    Tripathi, Mukesh
    Subedi, Asish
    [J]. CRITICAL CARE MEDICINE, 2015, 43 (08) : 1745 - 1756
  • [7] Aspiration of the guidewire of a central venous jugular catheter by the venous cannula of a veno-arterial extracorporeal membrane oxygenation
    Seguin, Philippe
    Arnouat, Matthieu
    Launey, Yoann
    [J]. EUROPEAN HEART JOURNAL-CASE REPORTS, 2019, 3 (02)