Facility for spectral radiance calibration at low light level

被引:0
作者
Wu, Zhifeng [1 ]
Wang, Qiming [1 ,2 ]
Dai, Caihong [1 ]
Li, Ling [1 ]
Sun, Ruoduan [1 ]
Wang, Yanfei [1 ]
He, Sijie [1 ]
Guo, Qing [2 ]
机构
[1] Natl Inst Metrol, Div Opt Metrol, Beijing 100029, Peoples R China
[2] Beijing Univ Chem Technol, Coll Informat Sci & Technol, Beijing 100029, Peoples R China
关键词
741.3 Optical Devices and Systems - 944.7 Radiation Measuring Instruments;
D O I
10.1364/OE.520431
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The spectral radiance measurement at daytime level can be realized with high accuracy, while it's difficult when the spectral radiance is at nighttime level. We design a spectral radiance calibration facility which has the characteristics of completely unchanged spectrum over 3 orders of magnitude and approximately unchanged spectrum for about 6 orders of magnitude. It combines a spectral radiance light source, a precision aperture and a white diffuser together, make it easy to reproduce the spectral radiance at 380 nm from 4 x 10-9 W/(m2<middle dot>sr<middle dot>nm) to 4 x 10-3 W/(m2<middle dot>sr<middle dot>nm). The facility can be easily used to calibrate a spectroradiometer at nighttime level. When the spectral radiance from 380 nm to 780 nm is around 1 x 10-7W/(m2<middle dot>sr<middle dot>nm), the calibration uncertainty of the spectroradiometer is 0.87%similar to 1.0% (k= 1).
引用
收藏
页码:14837 / 14846
页数:10
相关论文
共 21 条
  • [11] Improved VIIRS Day/Night Band Imagery With Near-Constant Contrast
    Liang, Calvin K.
    Mills, Stephen
    Hauss, Bruce I.
    Miller, Steven D.
    [J]. IEEE TRANSACTIONS ON GEOSCIENCE AND REMOTE SENSING, 2014, 52 (11): : 6964 - 6971
  • [12] Suomi NPP VIIRS day-night band on-orbit performance
    Liao, L. B.
    Weiss, Stephanie
    Mills, Steve
    Hauss, Bruce
    [J]. JOURNAL OF GEOPHYSICAL RESEARCH-ATMOSPHERES, 2013, 118 (22) : 12705 - 12718
  • [13] Suomi satellite brings to light a unique frontier of nighttime environmental sensing capabilities
    Miller, Steven D.
    Mills, Stephen P.
    Elvidge, Christopher D.
    Lindsey, Daniel T.
    Lee, Thomas F.
    Hawkins, Jeffrey D.
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2012, 109 (39) : 15706 - 15711
  • [14] Sadler F. B., 2008, PROC 33 ANN EUROPEAN, V292, P87
  • [15] Report on the APMP.PR-S6: 2012-2013 supplementary comparison of spectral radiance from 250 nm to 2500 nm
    Shin, Dong-Joo
    Park, Seongchong
    Lee, Dong-Hoon
    Dai Caihong
    Wu Zhifeng
    Wang Yanfei
    Khlevnoy, Boris
    Solodilov, Maxim
    Kolesnikova, Svetlana
    [J]. METROLOGIA, 2020, 57 (1A)
  • [16] Steve M., 2013, Proc. of SPIE, V8866
  • [17] Sun R. D., 2021, Metrology Science and Technology, V65, P36
  • [18] Wu Z. F., 2022, J. Phys.: Conf. Ser., V2149
  • [19] Methods to reduce the size-of-source eitect in radiometers
    Yoon, HW
    Allen, DW
    Saunders, RD
    [J]. METROLOGIA, 2005, 42 (02) : 89 - 96
  • [20] On-Orbit Relative Radiometric Calibration of the Night-Time Sensor of the LuoJia1-01 Satellite
    Zhang, Guo
    Li, Litao
    Jiang, Yonghua
    Shen, Xin
    Li, Deren
    [J]. SENSORS, 2018, 18 (12)