Agreement on Anterior Chamber Depth Measurement between Three Commercially Available Devices

被引:1
作者
Noya-Padin, Veronica [1 ,2 ]
Pena-Verdeal, Hugo [1 ,2 ]
Garcia-Queiruga, Jacobo [1 ,2 ]
Sampedro-Marino, Irene [1 ]
Giraldez, Maria Jesus [1 ,2 ]
Yebra-Pimentel, Eva [1 ,2 ]
机构
[1] Univ Santiago de Compostela, Appl Phys Dept, Fac Opt & Optometria, Optometry Area, Santiago De Compostela 15705, Spain
[2] Inst Invest Sanitaria Santiago de Compostela IDIS, Optometry Grp, Santiago De Compostela 15706, Spain
关键词
anterior chamber depth; ocular biometric agreement; VX-120multi-diagnostic platform; Lenstar LS900; EchoScan US-800; LOW-COHERENCE REFLECTOMETRY; OPTICAL BIOMETRY; STATISTICAL GUIDELINES; REPEATABILITY; DESIGNS;
D O I
10.3390/photonics11040377
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The present study aimed to assess the agreement of three commercially available devices on the measurement of anterior chamber depth (ACD) with and without compensation by central corneal thickness measurement (CCT). Fifty eyes were included in an observational cross-sectional study. Participants underwent a single visit during which devices were used to obtain the inclusion/exclusion (ARK510A, Canon TX-10) and studied (VX-120, Lenstar LS900 and EchoScan US-800) parameters. Based on invasiveness, tests were always performed in the same order by one researcher (to avoid inter-observer variability) and only in the right eye (to avoid overstating the precision of estimates) in each participant. The keratometry, autorefraction, intraocular pressure and anterior chamber angle values were used as inclusion criteria, while the CCT and ACD values were used in the agreement analysis between devices. There was a general and a paired difference in ACD measurements between devices (Greenhouse-Geisser: p <= 0.001; Sidak: all p <= 0.001). No significant difference was found in ACD measurements compensated by CCT values between the devices (Greenhouse-Geisser: p = 0.200). Pairwise analysis showed a significant difference in VX-120 vs. Lenstar (Sidak: p = 0.021). The differences in ACD measurements compensated by CCT values between the devices were clinically acceptable. Consequently, using these instruments interchangeably in daily routines based on this correction is justified.
引用
收藏
页数:12
相关论文
共 49 条
[1]   Validation of the multi-metric D-index change in the assessment of keratoconus progression [J].
Achiron, Asaf ;
Yavnieli, Roy ;
Olshaker, Hagar ;
Levinger, Eliya ;
Tuuminen, Raimo ;
Livny, Eitan ;
Elbaz, Uri ;
Bahar, Irit ;
Nahum, Yoav .
INTERNATIONAL OPHTHALMOLOGY, 2022, 42 (09) :2665-2671
[2]   Refractive Results Using a New Optical Biometry Device Comparison With Ultrasound Biometry Data [J].
Aktas, Serdar ;
Aktas, Hatice ;
Tetikoglu, Mehmet ;
Sagdik, Haci Murat ;
Ozcura, Fatih .
MEDICINE, 2015, 94 (48)
[3]   The application of analysis of variance (ANOVA) to different experimental designs in optometry [J].
Armstrong, RA ;
Eperjesi, F ;
Gilmartin, B .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2002, 22 (03) :248-256
[4]   Recommendations for analysis of repeated-measures designs: testing and correcting for sphericity and use of manova and mixed model analysis [J].
Armstrong, Richard A. .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2017, 37 (05) :585-593
[5]   When to use the Bonferroni correction [J].
Armstrong, Richard A. .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2014, 34 (05) :502-508
[6]   Statistical guidelines for the analysis of data obtained from one or both eyes [J].
Armstrong, Richard A. .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2013, 33 (01) :7-14
[7]   Statistical guidelines for clinical studies of human vision [J].
Armstrong, Richard A. ;
Davies, Leon N. ;
Dunne, Mark C. M. ;
Gilmartin, Bernard .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2011, 31 (02) :123-136
[8]   Agreement and clinical comparison between a new swept-source optical coherence tomography-based optical biometer and an optical low-coherence reflectometry biometer [J].
Arriola-Villalobos, P. ;
Almendral-Gomez, J. ;
Garzon, N. ;
Ruiz-Medrano, J. ;
Fernandez-Perez, C. ;
Martinez-de-la-Casa, J. M. ;
Diaz-Valle, D. .
EYE, 2017, 31 (03) :437-442
[9]   Evaluation of the effect of fasting on intraocular pressure, anterior segment parameters and density of crystalline lens and cornea [J].
Beyoglu, Abdullah ;
Karakucuk, Yalcin ;
comez, Aysegul .
INTERNATIONAL OPHTHALMOLOGY, 2020, 40 (11) :2987-2994
[10]   Ocular Biometry Features and Their Relationship with Anterior and Posterior Segment Lengths among a Myopia Population in Northern China [J].
Bian, Linbo ;
Li, Wenlong ;
Qin, Rui ;
Sun, Zhengze ;
Zhao, Lu ;
Zhou, Yifan ;
Liu, Dehai ;
Liu, Yiyun ;
Sun, Tong ;
Qi, Hong .
JOURNAL OF CLINICAL MEDICINE, 2024, 13 (04)