Validity and reliability of inertial measurement units used to measure motion of the lumbar spine: A systematic review of individuals with and without low back pain

被引:9
作者
Mcclintock, Frederick A. [1 ]
Callaway, Andrew J. [1 ]
Clark, Carol J. [1 ]
Williams, Jonathan M. [1 ]
机构
[1] Bournemouth Univ, Fac Hlth & Social Sci, Poole BH12 5BB, England
关键词
Low back; Pain; Sensors; Review; Biomechanics; CONCURRENT VALIDITY; GLOBAL BURDEN; KINEMATICS; ORIENTATION; PREVALENCE; DISEASE; QUALITY; STAND; TORSO; GAIT;
D O I
10.1016/j.medengphy.2024.104146
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Low back pain (LBP) is a leading cause of disability, resulting in aberrant movement. This movement is difficult to measure accurately in clinical practice and gold standard methods, such as optoelectronic systems involve the use of expensive laboratory equipment. Inertial measurement units (IMU) offer an alternative method of quantifying movement that is accessible in most environments. However, there is no consensus around the validity and reliability of IMUs for quantifying lumbar spine movements compared with gold standard measures. The aim of this systematic review was to establish concurrent validity and repeated measures reliability of using IMUs for the measurement of lumbar spine movements in individuals with and without LBP. A systematic search of electronic databases, incorporating PRISMA guidelines was completed, limited to the English language. 503 studies were identified where 15 studies met the inclusion criteria. Overall, 305 individuals were included, and 109 of these individuals had LBP. Weighted synthesis of the results demonstrated root mean squared differences of <2.4 degrees compared to the gold standard and intraclass correlations >0.84 for lumbar spine movements. IMUs offer clinicians and researchers valid and reliable measurement of motion in the lumbar spine, comparable to laboratory methods, such as optoelectronic motion capture for individuals with and without LBP.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Validity of inertial measurement units for tracking human motion: a systematic review
    Ghattas, John
    Jarvis, Danielle N.
    SPORTS BIOMECHANICS, 2021, : 1853 - 1866
  • [2] VALIDITY AND RELIABILITY OF INERTIAL MEASUREMENT UNITS FOR JUMP HEIGHT ESTIMATIONS: A SYSTEMATIC REVIEW
    Clemente, Filipe Manuel
    Badicu, Georgian
    Hasan, Uday Ch.
    Akyildiz, Zeki
    Pino-Ortega, Jose
    Silva, Rui
    Rico-Gonzalez, Markel
    HUMAN MOVEMENT, 2022, 23 (04) : 1 - 20
  • [3] Measurement of lumbar spine range of movement and coupled motion using inertial sensors - A protocol validity study
    Ha, Tshui-Hung
    Saber-Sheikh, Kambiz
    Moore, Ann P.
    Jones, Mark P.
    MANUAL THERAPY, 2013, 18 (01) : 87 - 91
  • [4] Validity, reliability and accuracy of inertial measurement units (IMUs) to measure angles: application in swimming
    Guignard, Brice
    Ayad, Omar
    Baillet, Heloise
    Mell, Florian
    Simbana Escobar, David
    Boulanger, Jeremie
    Seifert, Ludovic
    SPORTS BIOMECHANICS, 2021, : 1471 - 1503
  • [5] Assessing the Reliability and Validity of Inertial Measurement Units to Measure Three-Dimensional Spine and Hip Kinematics During Clinical Movement Tasks
    Bailes, Anna H.
    Johnson, Marit
    Roos, Rachel
    Clark, William
    Cook, Harold
    McKernan, Gina
    Sowa, Gwendolyn A.
    Cham, Rakie
    Bell, Kevin M.
    SENSORS, 2024, 24 (20)
  • [6] The Kinematics and Spondylosis of the Lumbar Spine Vary Depending on the Levels of Motion Segments in Individuals With Low Back Pain
    Basques, Bryce A.
    Orias, Alejandro A. Espinoza
    Shifflett, Grant D.
    Fice, Michael P.
    Andersson, Gunnar B.
    An, Howard S.
    Inoue, Nozomu
    SPINE, 2017, 42 (13) : E767 - E774
  • [7] Multi-segment spine range of motion in dancers with and without recent low back pain
    Swain, Christopher T. V.
    Bradshaw, Elizabeth J.
    Ekegren, Christina L.
    Orishimo, Karl F.
    Kremenic, Ian J.
    Liederbach, Marijeanne
    Hagins, Marshall
    GAIT & POSTURE, 2019, 70 : 53 - 58
  • [8] Lumbar spine kinematics during walking in people with and people without low back pain
    Gombatto, Sara P.
    Brock, Tricia
    DeLork, Anthony
    Jones, Glynis
    Madden, Erin
    Rinere, Chelsea
    GAIT & POSTURE, 2015, 42 (04) : 539 - 544
  • [9] Validity and Reliability of Inertial Measurement Units in Active Range of Motion Assessment in the Hip Joint
    Stolowski, Lukasz
    Niedziela, Maciej
    Lubiatowski, Bartlomiej
    Lubiatowski, Przemyslaw
    Piontek, Tomasz
    SENSORS, 2023, 23 (21)
  • [10] A systematic review and meta-analysis of the reliability and validity of sensorimotor measurement instruments in people with chronic low back pain
    Ehrenbrusthoff, Katja
    Ryan, Cormac G.
    Grueneberg, Christian
    Martin, Denis J.
    MUSCULOSKELETAL SCIENCE AND PRACTICE, 2018, 35 : 73 - 83