Key comparison BIPM.RI(I)-K8 of high dose-rate 192Ir brachytherapy standards for reference air kerma rate of the PTB and the BIPM

被引:5
作者
Kessler, C.
Behrens, R.
Kasper, A.
Grote, F.
机构
[1] Bureau International des Poids et Mesures, Sèvres,Cedex F-92312, France
[2] Physikalisch-Technische Bundesanstalt, Braunschweig,D-38116, Germany
关键词
D O I
10.1088/0026-1394/61/1A/06002
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An indirect comparison of the standards for reference air kerma rate for 192Ir high dose rate (HDR) brachytherapy sources of the Physikalisch-Technische Bundesanstalt (PTB), Germany, and of the Bureau International des Poids et Mesures (BIPM) was carried out at the PTB in May 2023. The comparison result, based on the calibration coefficients for a transfer standard and expressed as a ratio of the PTB and the BIPM standards for reference air kerma rate, is 1.0022 with a combined standard uncertainty of 0.0101. © 2024 Institute of Physics Publishing. All rights reserved.