Assessing Concordance of Results: A Comparative Study of the Manual and Automated Urinalysis Methods

被引:0
作者
Gyamfi, Nicholas Kwame Afriyie [1 ]
Osei, George Nkrumah [1 ]
Brenyah, Ruth C. [2 ]
Agyemang, Lawrence Duah [3 ]
Ampomah, Paulina [4 ]
Darkwah, Kwame Osei [5 ]
Toboh, Emmanuel [6 ]
Ephraim, Richard K. D. [1 ]
机构
[1] Univ Cape Coast, Sch Allied Hlth Sci, Dept Med Lab Sci, Cape Coast, Ghana
[2] Kwame Nkrumah Univ Sci Technol, Sch Med Sci, Dept Clin Microbiol, Kumasi, Ghana
[3] Komfo Anokye Teaching Hosp, Publ Hlth Unit, Kumasi, Ghana
[4] Univ Cape Coast, Sch Allied Hlth Sci, Dept Biomed Sci, Cape Coast, Ghana
[5] Jeonbuk Natl Univ, Inst Mol Biol & Genet, Coll Nat Sci, Jeonju, South Korea
[6] Dansoman Polyclin, Lab Unit, Accra, Ghana
关键词
URINE SEDIMENT EXAMINATION; PERFORMANCE; DIPSTICK; SYSTEMS;
D O I
10.1155/2024/6963423
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Introduction. An accurate urine analysis is a good indicator of the status of the renal and genitourinary system. However, limited studies have been done on comparing the diagnostic performance of the fully automated analyser and manual urinalysis especially in Ghana. This study evaluated the concordance of results of the fully automated urine analyser (Sysmex UN series) and the manual method urinalysis at the Komfo Anokye Teaching Hospital in Kumasi, Ghana. Methodology. Sixty-seven (67) freshly voided urine samples were analysed by the automated urine analyser Sysmex UN series and by manual examination at Komfo Anokye Teaching Hospital, Ghana. Kappa and Bland-Altman plot analyses were used to evaluate the degree of concordance and correlation of both methods, respectively. Results. Substantial (kappa=0.711, p<0.01), slight (kappa=0.193, p=0.004), and slight (kappa=0.109, p<0.001) agreements were found for urine colour, appearance, and pH, respectively, between the manual and automated methods. A strong and significant correlation (r=0.593, p<0.001) was found between both methods for specific gravity with a strong positive linear correlation observed for red blood cell count (r=0.951, R-2=0.904, p<0.001), white blood cell count (r=0.907, R-2=0.822, p<0.001), and epithelial cell count (r=0.729, R-2=0.532, p<0.001). A perfect agreement of urine chemistry results in both methods was observed for nitrite 67 (100%) (kappa=1.000, p<0.001) with a fair agreement for protein 46 (68.7%) (kappa=0.395, p<0.001). A strong agreement was found in both methods for the presence of cast 65 (97.0%) (kappa=0.734, p<0.001) with no concordance observed for the presence of crystals (kappa=0.115, p=0.326) and yeast-like cells (YLC) (kappa=0.171, p=0.116). Conclusion. The automated and manual methods showed similar performances and good correlation, especially for physical and chemical examination. However, manual microscopy remains necessary to classify urine sediments, particularly for bacteria and yeast-like cells. Future research with larger samples could help validate automated urinalysis for wider clinical use and identify areas requiring improved automated detection capabilities.
引用
收藏
页数:8
相关论文
共 25 条
  • [1] Urinalysis: The Automated Versus Manual Techniques; Is It Time To Change?
    Ahmed, Asmaa Ismail
    Baz, Heba
    Lotfy, Sarah
    [J]. CLINICAL LABORATORY, 2016, 62 (1-2) : 49 - 56
  • [2] Preliminary evaluation of the IRIS IQ™ 200 automated urine analyser
    Alves, L
    Ballester, F
    Camps, J
    Joven, J
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2005, 43 (09) : 967 - 970
  • [3] [Anonymous], 2000, SCAND J CLIN LAB I S, V231, P1, DOI [10.1080/00365513.2000.12056993, DOI 10.1080/00365513.2000.12056993]
  • [4] Evaluation of the analytical performances of Cobas 6500 and Sysmex UN series automated urinalysis systems with manual microscopic particle counting
    Bakan, Ebubekir
    Bayraktutan, Zafer
    Baygutalp, Nurcan Kilic
    Gul, Mehmet Ali
    Umudum, Fatma Zuhal
    Bakan, Nuri
    [J]. BIOCHEMIA MEDICA, 2018, 28 (02)
  • [5] Budak YU, 2011, CLIN LAB, V57, P47
  • [6] Urine sediment examination: A comparison of automated urinalysis systems and manual microscopy
    Chien, Tzu-I
    Kao, Jau-Tsuen
    Liu, Hui-Lan
    Lin, Po-Chang
    Hong, Jhih-Sian
    Hsieh, Han-Peng
    Chien, Miao-Ju
    [J]. CLINICA CHIMICA ACTA, 2007, 384 (1-2) : 28 - 34
  • [7] The efficient workflow to decrease the manual microscopic examination of urine sediment using on-screen review of images
    Cho, Eun-Jung
    Ko, Dae-Hyun
    Lee, Woochang
    Chun, Sail
    Lee, Hae Kyung
    Min, Won-Ki
    [J]. CLINICAL BIOCHEMISTRY, 2018, 56 : 70 - 74
  • [8] Urinalysis: Core curriculum 2008
    Fogazzi, Giovanni B.
    Verdesca, Simona
    Garigali, Giuseppe
    [J]. AMERICAN JOURNAL OF KIDNEY DISEASES, 2008, 51 (06) : 1052 - 1067
  • [9] BA-plotteR-A web tool for generating Bland-Altman plots and constructing limits of agreement
    Goedhart, Joachim
    Rishniw, Mark
    [J]. RESEARCH IN VETERINARY SCIENCE, 2021, 137 : 281 - 286
  • [10] Ince Fatma Demet, 2016, Pract Lab Med, V5, P14, DOI 10.1016/j.plabm.2016.03.002