Robotic-Assisted Navigation for Stereotactic Neurosurgery: A Cadaveric Investigation of Accuracy, Time, and Radiation

被引:1
作者
Anderson, William [1 ]
Ponce, Francisco A. [2 ]
Kinsman, Michael J. [3 ]
Sani, Sepehr [4 ]
Hwang, Brian [1 ,8 ]
Ghinda, Diana [1 ]
Kogan, Michael [5 ]
Mahoney, Jonathan M. [6 ]
Amin, Dhara B. [6 ]
Van Horn, Margaret [6 ]
McGuckin, Joshua P. [6 ,9 ]
Razo-Castaneda, Dominic [7 ]
Bucklen, Brandon S. [6 ]
机构
[1] Johns Hopkins Univ Hosp, Dept Neurosurg, Baltimore, MD USA
[2] St Josephs Hosp, Barrow Neurol Inst, Dept Neurosurg, Phoenix, AZ USA
[3] Univ Kansas, Med Ctr, Neurosurg, Kansas City, KS USA
[4] Rush Univ, Med Ctr, Dept Neurosurg, Chicago, IL USA
[5] Thomas Jefferson Univ Hosp, Dept Neurol Surg, Philadelphia, PA USA
[6] Musculoskeletal Educ & Res Ctr, Audubon, PA USA
[7] Drexel Univ, Sch Biomed Engn Sci & Hlth Syst, Philadelphia, PA USA
[8] Orange Cty Neurosurg Associates, Laguna Hills, CA USA
[9] Musculoskeletal Educ & Res Ctr MERC, 2560 Gen Armistead Ave, Audubon, PA 19403 USA
关键词
Cranial; DBS; Navigation; Radiation; Robot; SEEG; Stereotactic; DEEP BRAIN-STIMULATION; PLACEMENT; RADIOSURGERY; EXPERIENCE;
D O I
10.1227/ons.0000000000001024
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND OBJECTIVES: Despite frequent use, stereotactic head frames require manual coordinate calculations and manual frame settings that are associated with human error. This study examines freestanding robot-assisted navigation (RAN) as a means to reduce the drawbacks of traditional cranial stereotaxy and improve targeting accuracy. METHODS: Seven cadaveric human torsos with heads were tested with 8 anatomic coordinates selected for lead placement mirrored in each hemisphere. Right and left hemispheres of the brain were randomly assigned to either the traditional stereotactic arc-based (ARC) group or the RAN group. Both target accuracy and trajectory accuracy were measured. Procedural time and the radiation required for registration were also measured. RESULTS: The accuracy of the RAN group was significantly greater than that of the ARC group in both target (1.2 +/- 0.5 mm vs 1.7 +/- 1.2 mm, P = .005) and trajectory (0.9 +/- 0.6 mm vs 1.3 +/- 0.9 mm, P = .004) measurements. Total procedural time was also significantly faster for the RAN group than for the ARC group (44.6 +/- 7.7 minutes vs 86.0 +/- 12.5 minutes, P < .001). The RAN group had significantly reduced time per electrode placement (2.9 +/- 0.9 minutes vs 5.8 +/- 2.0 minutes, P < .001) and significantly reduced radiation during registration (1.9 +/- 1.1 mGy vs 76.2 +/- 5.0 mGy, P < .001) compared with the ARC group. CONCLUSION: In this cadaveric study, cranial leads were placed faster and with greater accuracy using RAN than those placed with conventional stereotactic arc-based technique. RAN also required significantly less radiation to register the specimen's coordinate system to the planned trajectories. Clinical testing should be performed to further investigate RAN for stereotactic cranial surgery.
引用
收藏
页码:568 / 575
页数:8
相关论文
共 32 条
  • [1] Frameless robot-assisted stereoelectroencephalography in children: technical aspects and comparison with Talairach frame technique
    Abel, Taylor J.
    Osorio, Rene Varela
    Amorim-Leite, Ricardo
    Mathieu, Francois
    Kahane, Philippe
    Minotti, Lorella
    Hoffmann, Dominique
    Chabardes, Stephan
    [J]. JOURNAL OF NEUROSURGERY-PEDIATRICS, 2018, 22 (01) : 37 - 46
  • [2] Double-blind clinical trial of thalamic stimulation in patients with Tourette syndrome
    Ackermans, Linda
    Duits, Annelien
    van der Linden, Chris
    Tijssen, Marina
    Schruers, Koen
    Temel, Yasin
    Kleijer, Mariska
    Nederveen, Pieter
    Bruggeman, Richard
    Tromp, Selma
    van Kranen-Mastenbroek, Vivianne
    Kingma, Herman
    Cath, Danielle
    Visser-Vandewalle, Veerle
    [J]. BRAIN, 2011, 134 : 832 - 844
  • [3] [Anonymous], 2012, StealthStationTMApplication. Version 2.0.1-7
  • [4] Strategies to Alleviate Anxiety Before the Placement of a Stereotactic Radiosurgery Frame
    Avbovbo, Ufuoma E.
    Appel, Susan J.
    [J]. JOURNAL OF NEUROSCIENCE NURSING, 2016, 48 (04) : 224 - 228
  • [5] Surgery and ergonomics
    Berguer, R
    [J]. ARCHIVES OF SURGERY, 1999, 134 (09) : 1011 - 1016
  • [6] Preclinical Evaluation of the Stealth Autoguide Robotic Guidance Device for Stereotactic Cranial Surgery: A Human Cadaveric Study
    Brandman, David
    Hong, Murray
    Clarke, David B.
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2021, 99 (04) : 343 - 350
  • [7] Accuracy of deep brain stimulation electrode placement using intraoperative computed tomography without microelectrode recording Clinical article
    Burchiel, Kim J.
    McCartney, Shirley
    Lee, Albert
    Raslan, Ahmed M.
    [J]. JOURNAL OF NEUROSURGERY, 2013, 119 (02) : 301 - 306
  • [8] Intraoperative 3D imaging control during subthalamic Deep Brain Stimulation procedures using O-arm® technology: Experience in 15 patients
    Caire, F.
    Guehl, D.
    Burbaud, P.
    Benazzouz, A.
    Cuny, E.
    [J]. NEUROCHIRURGIE, 2014, 60 (06) : 276 - 282
  • [9] Frameless robot-assisted stereoelectroencephalography for refractory epilepsy in pediatric patients: accuracy, usefulness, and technical issues
    Candela-Canto, Santiago
    Aparicio, Javier
    Muchart Lopez, Jordi
    Banos-Carrasco, Pilar
    Ramirez-Camacho, Alia
    Climent, Alejandra
    Alamar, Mariana
    Jou, Cristina
    Rumia, Jordi
    San Antonio-Arce, Victoria
    Arzimanoglou, Alexis
    Ferrer, Enrique
    [J]. ACTA NEUROCHIRURGICA, 2018, 160 (12) : 2489 - 2500
  • [10] A novel miniature robotic device for frameless implantation of depth electrodes in refractory epilepsy
    Dorfer, Christian
    Minchev, Georgi
    Czech, Thomas
    Stefanits, Harald
    Feucht, Martha
    Pataraia, Ekaterina
    Baumgartner, Christoph
    Kronreif, Gernot
    Wolfsberger, Stefan
    [J]. JOURNAL OF NEUROSURGERY, 2017, 126 (05) : 1622 - 1628